COLA citation FAQ
These are direct answers to the questions laboratory directors, managers and testing staff ask most often about COLA citations and CLIA compliance. Each answer links to the full guide for that citation area.
Is a COLA citation the same as a CLIA deficiency?
They are closely linked. COLA criteria are written to meet or exceed CLIA, so a COLA citation usually corresponds to a CLIA regulation in 42 CFR Part 493. A CMS or state survey uses Form CMS-2567 and D-tags instead of COLA criteria. Read the full guide
Can a lab lose accreditation over one citation?
Usually not for a single standard-level finding that is corrected promptly. Serious, systemic or repeated findings, failure to submit acceptable corrections, or proficiency testing referral can lead to directed corrections, limits on testing or loss of accreditation. Read the full guide
How often does COLA survey a laboratory?
COLA surveys accredited laboratories on a two-year cycle, generally scheduled before the two-year anniversary of the last survey. Complaint or follow-up surveys can happen in between. Read the full guide
Can I dispute a COLA citation?
Yes. If you believe a criterion was cited in error, the laboratory director can submit records that show compliance and ask COLA to review it. Send clear evidence, such as dated records, invoices or signed documents. Read the full guide
How long do I have to respond to a COLA citation?
Follow the due date in your COLA post-survey report, since that date controls. For CMS or state surveys documented on Form CMS-2567, a plan of correction is generally due within 10 days of receipt. Read the full guide
Who signs the plan of correction?
The laboratory director should review and sign it. The director is ultimately responsible for the lab's compliance even when duties are delegated. Read the full guide
What is a directed plan of correction?
It is a correction plan the accrediting body or CMS writes for the lab, rather than one the lab chooses. Examples include hiring a qualified consultant or completing specific training. Read the full guide
Do I need evidence, or just a written statement?
Send evidence. Completed records, signed procedures and training documentation show the problem is fixed, and a statement alone is often not accepted. Read the full guide
What is an immediate jeopardy finding in a lab?
It is a situation where noncompliance has caused, or is likely to cause, serious harm. It calls for urgent action, and sanctions can follow quickly. Read the full guide
What are D-tags?
D-tags are the codes CMS surveyors use on Form CMS-2567 to identify each CLIA regulation that was cited. Read the full guide
How often is competency assessment required under CLIA?
Twice during the first year an individual tests patient specimens, then at least once a year. Retraining and reassessment are needed when problems are identified. Read the full guide
Does the laboratory director need a competency assessment?
Not as testing personnel. The director's regulatory duties are reviewed during survey. If the director also serves in another role, such as technical consultant, competency for that role should be documented. Read the full guide
Is competency required for waived tests?
CLIA's six-element competency requirement applies to nonwaived testing. Training and good practice are still expected for waived testing, and some states and accreditors have their own rules. Read the full guide
Can proficiency testing count toward competency?
Yes. Testing external proficiency testing samples can satisfy the element that assesses test performance, as long as the person actually tested the samples and it is documented. Read the full guide
How many levels of QC does CLIA require?
For quantitative tests, two control materials of different concentrations each day of patient testing. For qualitative tests, a positive and a negative control each day. Stricter manufacturer instructions apply, and an approved IQCP can change frequency. Read the full guide
Can I report patient results if QC fails?
No. Control results must meet acceptance criteria before patient results are reported. Investigate, correct, and rerun before releasing results. Read the full guide
What counts as unsuccessful PT performance?
Failing the same analyte, specialty or subspecialty in two consecutive testing events, or in two of three consecutive events. Read the full guide
What happens if a lab refers PT samples to another lab?
PT referral is a serious violation. It can lead to revocation of the CLIA certificate, and the owner and director can be barred from operating a lab. Read the full guide
Do I need to investigate a passing score?
Review every PT result. Investigate passing results that show bias or a trend, and evaluate results that were ungraded. Read the full guide
What laboratory director duties cannot be delegated?
Some duties, such as approving, signing and dating procedures, cannot be delegated. For any duty that can be delegated, the director stays responsible and must make sure it is done. Read the full guide
How many labs can one director oversee?
CLIA limits an individual to directing no more than five laboratories. Read the full guide
Does the director have to visit the lab in person?
Under the rules effective December 28, 2024, yes: at least once every six months, with at least four months between visits, and documented. A 2026 proposed rule would remove this, but it is not final. Read the full guide
Is a diploma enough, or do I need a transcript?
A diploma that clearly states the degree can work. A transcript is often needed to prove specific coursework, for example the science credits used in an educational equivalency pathway. Read the full guide
Are nurses qualified to perform moderate complexity testing?
Nursing degrees can qualify under the CLIA testing personnel rules. Confirm the specific pathway under the rules in effect and keep the transcript on file. Read the full guide
Is calibration verification the same as calibration?
No. Calibration adjusts the instrument. Calibration verification tests materials of known value to confirm the existing calibration still gives accurate results across the reportable range. Read the full guide
How many levels are needed?
At least three: low, mid and high in the reportable range. Read the full guide
Is the vendor's installation report enough?
Usually not. The lab must document its own verification and have the director approve it before reporting patient results. Read the full guide
Do waived tests need method verification?
CLIA's performance specification rules apply to nonwaived tests. Follow the manufacturer's instructions for waived tests. Read the full guide
Can I use the package insert as my procedure?
A manufacturer's insert can serve as part of the procedure if it covers the required content and the director approves it, along with anything the lab does differently. Read the full guide
How long are retired procedures kept?
At least two years after they are discontinued, with the dates they were in use. Read the full guide
Do I need temperature logs on weekends?
Record conditions on every day testing is performed, and make sure reagents stayed within the required conditions on other days too. Continuous monitors with alerts make this easier. Read the full guide
How often should QA be reviewed?
CLIA says ongoing and does not set a single interval. Most labs document a monthly or quarterly review and adjust based on test volume and risk. Read the full guide
Does a backup instrument count?
Yes. If it is used for patient testing, its results must be compared with the primary method. Read the full guide
How long must lab records be kept under CLIA?
Generally at least two years. Immunohematology records are kept for five years, and pathology reports and slides for longer periods. State law may require more. Read the full guide
Are electronic orders acceptable?
Yes. Electronic test requests are acceptable when they come from an authorized person and contain the required information. Read the full guide
Can IQCP lower QC below the manufacturer's instructions?
No. The manufacturer's minimum QC frequency always applies. Read the full guide
What documents does COLA review during a survey?
Expect review of personnel and competency records, procedures, QC, proficiency testing, calibration, maintenance, quality assessment and patient reports, usually covering the period since the last survey. Read the full guide
How long does a COLA survey take?
It depends on the test menu and volume. Small physician office labs often take part of a day, and large or hospital labs can take several days. Read the full guide
Is the director onsite visit requirement gone?
Not yet. The 2026 rule only proposes removing it. Until a final rule takes effect, directors must still visit at least once every six months, with at least four months between visits, and document each visit. Read the full guide
When could a final rule take effect?
CMS has not announced a date. A final rule would come after the comment period and would state its own effective date. Read the full guide