COLA quality control (QC) citations
A quality control citation means the lab could not show that controls were run at the required frequency, met acceptance criteria before patient results were released, and were reviewed and corrected when they failed. CLIA's default rule is two levels of control for quantitative tests, or positive and negative controls for qualitative tests, each day of patient testing, unless an approved IQCP applies.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1256Standard: Control procedures | Control procedures |
| 42 CFR § 493.1282Standard: Corrective actions | Corrective actions |
| 42 CFR § 493.1281Standard: Comparison of test results | Comparison of test results |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1256 Standard: Control procedures.
(a) For each test system, the laboratory is responsible for having control procedures that monitor the accuracy and precision of the complete analytic process.
(b) The laboratory must establish the number, type, and frequency of testing control materials using, if applicable, the performance specifications verified or established by the laboratory as specified in § 493.1253(b)(3).
(c) The control procedures must—
(1) Detect immediate errors that occur due to test system failure, adverse environmental conditions, and operator performance.
(2) Monitor over time the accuracy and precision of test performance that may be influenced by changes in test system performance and environmental conditions, and variance in operator performance.
(d) Unless CMS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must—
(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at §§ 493.1261 through 493.1278.
(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section.
(3) At least once each day patient specimens are assayed or examined perform the following for—
(i) Each quantitative procedure, include two control materials of different concentrations;
(ii) Each qualitative procedure, include a negative and positive control material;
(iii) Test procedures producing graded or titered results, include a negative control material and a control material with graded or titered reactivity, respectively;
(iv) Each test system that has an extraction phase, include two control materials, including one that is capable of detecting errors in the extraction process; and
(v) Each molecular amplification procedure, include two control materials and, if reaction inhibition is a significant source of false negative results, a control material capable of detecting the inhibition.
(4) For thin layer chromatography—
(i) Spot each plate or card, as applicable, with a calibrator containing all known substances or drug groups, as appropriate, which are identified by thin layer chromatography and reported by the laboratory; and
(ii) Include at least one control material on each plate or card, as applicable, which must be processed through each step of patient testing, including extraction processes.
(5) For each electrophoretic procedure include, concurrent with patient specimens, at least one control material containing the substances being identified or measured.
(6) Perform control material testing as specified in this paragraph before resuming patient testing when a complete change of reagents is introduced; major preventive maintenance is performed; or any critical part that may influence test performance is replaced.
(7) Over time, rotate control material testing among all operators who perform the test.
(8) Test control materials in the same manner as patient specimens.
(9) When using calibration material as a control material, use calibration material from a different lot number than that used to establish a cut-off value or to calibrate the test system.
(10) Establish or verify the criteria for acceptability of all control materials.
(i) When control materials providing quantitative results are used, statistical parameters (for example, mean and standard deviation) for each batch and lot number of control materials must be defined and available.
(ii) The laboratory may use the stated value of a commercially assayed control material provided the stated value is for the methodology and instrumentation employed by the laboratory and is verified by the laboratory.
(iii) Statistical parameters for unassayed control materials must be established over time by the laboratory through concurrent testing of control materials having previously determined statistical parameters.
(e) For reagent, media, and supply checks, the laboratory must do the following:
(1) Check each batch (prepared in-house), lot number (commercially prepared) and shipment of reagents, disks, stains, antisera, (except those specifically referenced in § 493.1261(a)(3)) and identification systems (systems using two or more substrates or two or more reagents, or a combination) when prepared or opened for positive and negative reactivity, as well as graded reactivity, if applicable.
(2) Each day of use (unless otherwise specified in this subpart), test staining materials for intended reactivity to ensure predictable staining characteristics. Control materials for both positive and negative reactivity must be included, as appropriate.
(3) Check fluorescent and immunohistochemical stains for positive and negative reactivity each time of use.
(4) Before, or concurrent with the initial use—
(i) Check each batch of media for sterility if sterility is required for testing;
(ii) Check each batch of media for its ability to support growth and, as appropriate, select or inhibit specific organisms or produce a biochemical response; and
(iii) Document the physical characteristics of the media when compromised and report any deterioration in the media to the manufacturer.
(5) Follow the manufacturer's specifications for using reagents, media, and supplies and be responsible for results.
(f) Results of control materials must meet the laboratory's and, as applicable, the manufacturer's test system criteria for acceptability before reporting patient test results.
(g) The laboratory must document all control procedures performed.
(h) If control materials are not available, the laboratory must have an alternative mechanism to detect immediate errors and monitor test system performance over time. The performance of alternative control procedures must be documented.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1282 Standard: Corrective actions.
(a) Corrective action policies and procedures must be available and followed as necessary to maintain the laboratory's operation for testing patient specimens in a manner that ensures accurate and reliable patient test results and reports.
(b) The laboratory must document all corrective actions taken, including actions taken when any of the following occur:
(1) Test systems do not meet the laboratory's verified or established performance specifications, as determined in § 493.1253(b), which include but are not limited to—
(i) Equipment or methodologies that perform outside of established operating parameters or performance specifications;
(ii) Patient test values that are outside of the laboratory's reportable range of test results for the test system; and
(iii) When the laboratory determines that the reference intervals (normal values) for a test procedure are inappropriate for the laboratory's patient population.
(2) Results of control or calibration materials, or both, fail to meet the laboratory's established criteria for acceptability. All patient test results obtained in the unacceptable test run and since the last acceptable test run must be evaluated to determine if patient test results have been adversely affected. The laboratory must take the corrective action necessary to ensure the reporting of accurate and reliable patient test results.
(3) The criteria for proper storage of reagents and specimens, as specified under § 493.1252(b), are not met.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1281 Standard: Comparison of test results.
(a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites.
(b) The laboratory must have a system to identify and assess patient test results that appear inconsistent with the following relevant criteria, when available:
(1) Patient age.
(2) Sex.
(3) Diagnosis or pertinent clinical data.
(4) Distribution of patient test results.
(5) Relationship with other test parameters.
(c) The laboratory must document all test result comparison activities.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
What surveyors check
- Control frequency matches the CLIA default, the manufacturer's instructions, or your approved IQCP, whichever is stricter.
- Acceptable ranges are established or verified for each new lot.
- Controls were in range before patient results were reported.
- Corrective action is documented for every out-of-range control, including a patient lookback.
- QC records show periodic review by a qualified supervisor or consultant.
- New reagent and control lots are checked before use.
Most common QC citations
- Patient results released after a failed control.
- QC performed but never reviewed or signed.
- Ranges copied from the package insert without verification.
- Reduced QC frequency without an IQCP.
- Trends and shifts ignored on Levey-Jennings charts.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
How many levels of QC does CLIA require?
For quantitative tests, two control materials of different concentrations each day of patient testing. For qualitative tests, a positive and a negative control each day. Stricter manufacturer instructions apply, and an approved IQCP can change frequency.
Can I report patient results if QC fails?
No. Control results must meet acceptance criteria before patient results are reported. Investigate, correct, and rerun before releasing results.