Maintenance, function check and temperature citations

Reviewed November 26, 2026 by the Sound Lab Consultants laboratory compliance team
Short answer

These citations mean the lab could not show that instrument maintenance and function checks were done at least as often as the manufacturer requires, or that temperature, humidity and other conditions were monitored and corrected when out of range. Missing logs and out-of-range readings with no action are the most common causes.

Regulations behind this citation

RegulationWhat it requires
42 CFR § 493.1254Standard: Maintenance and function checksMaintenance and function checks
42 CFR § 493.1252Standard: Test systems, equipment, instruments, reagents, materials, and suppliesTest systems, equipment, instruments, reagents, materials and supplies

Full text of the regulations

Tap a regulation to read the complete official text. Each one links to its section on eCFR.

§ 493.1254 Standard: Maintenance and function checks.

(a) Unmodified manufacturer's equipment, instruments, or test systems. The laboratory must perform and document the following:

(1) Maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer.

(2) Function checks as defined by the manufacturer and with at least the frequency specified by the manufacturer. Function checks must be within the manufacturer's established limits before patient testing is conducted.

(b) Equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer. The laboratory must do the following:

(1)(i) Establish a maintenance protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting.

(ii) Perform and document the maintenance activities specified in paragraph (b)(1)(i) of this section.

(2)(i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting.

(ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted.

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

§ 493.1252 Standard: Test systems, equipment, instruments, reagents, materials, and supplies.

(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under § 493.1253.

(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following:

(1) Water quality.

(2) Temperature.

(3) Humidity.

(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports.

(c) Reagents, solutions, culture media, control materials, calibration materials, and other supplies, as appropriate, must be labeled to indicate the following:

(1) Identity and when significant, titer, strength or concentration.

(2) Storage requirements.

(3) Preparation and expiration dates.

(4) Other pertinent information required for proper use.

(d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality.

(e) Components of reagent kits of different lot numbers must not be interchanged unless otherwise specified by the manufacturer.

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

What surveyors review

Example plan of correction

A model response to a typical citation in this area. Adapt it to your lab's facts and records.

Deficiency cited
The reagent refrigerator was recorded out of range on three days with no corrective action.
Root cause & correction
Root cause: staff did not know the acceptable range or what to do. Correction: reagents were evaluated, the unit was serviced, and the range and steps are now printed on the log.
Evidence attached
Annotated temperature logs, service record and training sign-off.
Monitoring
Continuous monitoring with alerts, with weekly log review by the supervisor.

Frequently asked questions

Do I need temperature logs on weekends?

Record conditions on every day testing is performed, and make sure reagents stayed within the required conditions on other days too. Continuous monitors with alerts make this easier.

Related citation guides

Cited on your last survey?

Our lab consultants write plans of correction for COLA and CMS-2567 citations and fixes the root cause so the citation doesn't come back. MyLabCompliance.io keeps competency, QC, PT, calibration and procedures survey-ready.

Book a free consultationSee MyLabCompliance.io

Or call (602) 874-1311

These guides are independent educational content from Sound Lab Consultants, based on 42 CFR Part 493 and published accreditation guidance, reviewed November 26, 2026. Not affiliated with or endorsed by COLA Inc., CMS or CDC; COLA is a registered trademark of COLA Inc. Not legal advice. Always follow your current COLA Accreditation Manual and the instructions in your survey report.