Maintenance, function check and temperature citations
These citations mean the lab could not show that instrument maintenance and function checks were done at least as often as the manufacturer requires, or that temperature, humidity and other conditions were monitored and corrected when out of range. Missing logs and out-of-range readings with no action are the most common causes.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1254Standard: Maintenance and function checks | Maintenance and function checks |
| 42 CFR § 493.1252Standard: Test systems, equipment, instruments, reagents, materials, and supplies | Test systems, equipment, instruments, reagents, materials and supplies |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1254 Standard: Maintenance and function checks.
(a) Unmodified manufacturer's equipment, instruments, or test systems. The laboratory must perform and document the following:
(1) Maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer.
(2) Function checks as defined by the manufacturer and with at least the frequency specified by the manufacturer. Function checks must be within the manufacturer's established limits before patient testing is conducted.
(b) Equipment, instruments, or test systems developed in-house, commercially available and modified by the laboratory, or maintenance and function check protocols are not provided by the manufacturer. The laboratory must do the following:
(1)(i) Establish a maintenance protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting.
(ii) Perform and document the maintenance activities specified in paragraph (b)(1)(i) of this section.
(2)(i) Define a function check protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting.
(ii) Perform and document the function checks, including background or baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks must be within the laboratory's established limits before patient testing is conducted.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1252 Standard: Test systems, equipment, instruments, reagents, materials, and supplies.
(a) Test systems must be selected by the laboratory. The testing must be performed following the manufacturer's instructions and in a manner that provides test results within the laboratory's stated performance specifications for each test system as determined under § 493.1253.
(b) The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented and, if applicable, include the following:
(1) Water quality.
(2) Temperature.
(3) Humidity.
(4) Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports.
(c) Reagents, solutions, culture media, control materials, calibration materials, and other supplies, as appropriate, must be labeled to indicate the following:
(1) Identity and when significant, titer, strength or concentration.
(2) Storage requirements.
(3) Preparation and expiration dates.
(4) Other pertinent information required for proper use.
(d) Reagents, solutions, culture media, control materials, calibration materials, and other supplies must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality.
(e) Components of reagent kits of different lot numbers must not be interchanged unless otherwise specified by the manufacturer.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
What surveyors review
- Daily, weekly and monthly maintenance completed as the manufacturer specifies.
- Refrigerator, freezer, incubator and room temperatures recorded every day of testing, with acceptable ranges.
- Corrective action when readings are out of range, including what happened to the reagents.
- Reagent expiration dates and open dates.
- Pipette, centrifuge and thermometer checks.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
Do I need temperature logs on weekends?
Record conditions on every day testing is performed, and make sure reagents stayed within the required conditions on other days too. Continuous monitors with alerts make this easier.