How to write a COLA plan of correction

Reviewed November 26, 2026 by the Sound Lab Consultants laboratory compliance team
Short answer

To answer a COLA citation, submit a plan of correction for each cited criterion by the deadline in your post-survey report. Each response should state the root cause, the correction already made, how you fixed every affected record or patient result, the evidence attached, who monitors it, and the date compliance was reached.

Regulations behind this citation

RegulationWhat it requires
42 CFR § 493.1832Directed plan of correction and directed portion of a plan of correctionDirected plan of correction and directed portion of a plan of correction
42 CFR § 493.1282Standard: Corrective actionsCorrective actions
42 CFR § 493.1200IntroductionQuality systems for nonwaived testing

Full text of the regulations

Tap a regulation to read the complete official text. Each one links to its section on eCFR.

§ 493.1832 Directed plan of correction and directed portion of a plan of correction.

(a) Application. CMS may impose a directed plan of correction as an alternative sanction for any laboratory that has condition level deficiencies. If CMS does not impose a directed plan of correction as an alternative sanction for a laboratory that has condition level deficiencies, it at least imposes a directed portion of a plan of correction when it imposes any of the following alternative sanctions:

(1) State onsite monitoring.

(2) Civil money penalty.

(3) Suspension of all or part of Medicare payments.

(b) Procedures—(1) Directed plan of correction. When imposing this sanction, CMS—

(i) Gives the laboratory prior notice of the sanction and opportunity to respond in accordance with § 493.1810;

(ii) Directs the laboratory to take specific corrective action within specific time frames in order to achieve compliance; and

(iii) May direct the laboratory to submit the names of laboratory clients for notification purposes, as specified in paragraph (b)(3) of this section.

(2) Directed portion of a plan of correction. CMS may decide to notify clients of a sanctioned laboratory, because of the seriousness of the noncompliance (e.g., the existence of immediate jeopardy) or for other reasons. When imposing this sanction, CMS takes the following steps—

(i) Directs the laboratory to submit to CMS, the State survey agency, or other CMS agent, within 10 calendar days after the notice of the alternative sanction, a list of names and addresses of all physicians, providers, suppliers, and other clients who have used some or all of the services of the laboratory since the last certification inspection or within any other timeframe specified by CMS.

(ii) Within 30 calendar days of receipt of the information, may send to each laboratory client, via the State survey agency, a notice containing the name and address of the laboratory, the nature of the laboratory's noncompliance, and the kind and effective date of the alternative sanction.

(iii) Sends to each laboratory client, via the State survey agency, notice of the recission of an adverse action within 30 days of the rescission.

(3) Notice of imposition of a principal sanction following the imposition of an alternative sanction. If CMS imposes a principal sanction following the imposition of an alternative sanction, and for which CMS has already obtained a list of laboratory clients, CMS may use that list to notify the clients of the imposition of the principal sanction.

(c) Duration of a directed plan of correction. If CMS imposes a directed plan of correction, and on revisit it is found that the laboratory has not corrected the deficiencies within 12 months from the last day of inspection, the following rules apply:

(1) CMS cancels the laboratory's approval for Medicare payment of its services, and notifies the laboratory of CMS's intent to suspend, limit, or revoke the laboratory's CLIA certificate.

(2) The directed plan of correction continues in effect until the day suspension, limitation, or revocation of the laboratory's CLIA certificate.

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

§ 493.1282 Standard: Corrective actions.

(a) Corrective action policies and procedures must be available and followed as necessary to maintain the laboratory's operation for testing patient specimens in a manner that ensures accurate and reliable patient test results and reports.

(b) The laboratory must document all corrective actions taken, including actions taken when any of the following occur:

(1) Test systems do not meet the laboratory's verified or established performance specifications, as determined in § 493.1253(b), which include but are not limited to—

(i) Equipment or methodologies that perform outside of established operating parameters or performance specifications;

(ii) Patient test values that are outside of the laboratory's reportable range of test results for the test system; and

(iii) When the laboratory determines that the reference intervals (normal values) for a test procedure are inappropriate for the laboratory's patient population.

(2) Results of control or calibration materials, or both, fail to meet the laboratory's established criteria for acceptability. All patient test results obtained in the unacceptable test run and since the last acceptable test run must be evaluated to determine if patient test results have been adversely affected. The laboratory must take the corrective action necessary to ensure the reporting of accurate and reliable patient test results.

(3) The criteria for proper storage of reagents and specimens, as specified under § 493.1252(b), are not met.

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

§ 493.1200 Introduction.

(a) Each laboratory that performs nonwaived testing must establish and maintain written policies and procedures that implement and monitor a quality system for all phases of the total testing process (that is, preanalytic, analytic, and postanalytic) as well as general laboratory systems.

(b) The laboratory's quality systems must include a quality assessment component that ensures continuous improvement of the laboratory's performance and services through ongoing monitoring that identifies, evaluates and resolves problems.

(c) The various components of the laboratory's quality system are used to meet the requirements in this part and must be appropriate for the specialties and subspecialties of testing the laboratory performs, services it offers, and clients it serves.

[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

The six parts of an acceptable response

PartWhat to write
1. FindingRestate the cited criterion in one sentence, in your own words.
2. Root causeWhy it happened: no procedure, no owner, no reminder, a staffing gap, a misunderstood rule.
3. CorrectionWhat you fixed, and the date. Include a lookback at affected records or patient results.
4. Systemic fixThe process change that stops a repeat: a revised procedure, a calendar, a checklist, software.
5. EvidenceCopies of the completed records, signed policies, training sign-offs.
6. MonitoringWho checks, how often, for how long, and where it is reported (usually QA review).

Deadlines

Use the due date printed in your COLA post-survey report. That date controls, so put it on the calendar the day you receive it. If the citation came from a CMS or state agency survey on Form CMS-2567, the plan of correction is generally due within 10 days of receipt. Ask for an extension in writing before the deadline if you need more time to collect evidence. Do not wait until after it has passed.

Mistakes that get a plan sent back

Example plan of correction

A model response to a typical citation in this area. Adapt it to your lab's facts and records.

Deficiency cited
Competency assessment for two testing personnel was not documented in the past year.
Root cause & correction
Root cause: competency was tracked on paper with no due-date reminder. Correction: the technical consultant completed all six competency elements for both employees on the date shown; a lookback found no other overdue staff.
Evidence attached
Completed competency forms for both employees, the revised competency procedure signed by the laboratory director, and the new tracking calendar.
Monitoring
The technical consultant reviews the competency calendar each month and reports status at the quarterly QA meeting for 12 months.

Frequently asked questions

How long do I have to respond to a COLA citation?

Follow the due date in your COLA post-survey report, since that date controls. For CMS or state surveys documented on Form CMS-2567, a plan of correction is generally due within 10 days of receipt.

Who signs the plan of correction?

The laboratory director should review and sign it. The director is ultimately responsible for the lab's compliance even when duties are delegated.

What is a directed plan of correction?

It is a correction plan the accrediting body or CMS writes for the lab, rather than one the lab chooses. Examples include hiring a qualified consultant or completing specific training.

Do I need evidence, or just a written statement?

Send evidence. Completed records, signed procedures and training documentation show the problem is fixed, and a statement alone is often not accepted.

Related citation guides

Cited on your last survey?

Our lab consultants write plans of correction for COLA and CMS-2567 citations and fixes the root cause so the citation doesn't come back. MyLabCompliance.io keeps competency, QC, PT, calibration and procedures survey-ready.

Book a free consultationSee MyLabCompliance.io

Or call (602) 874-1311

These guides are independent educational content from Sound Lab Consultants, based on 42 CFR Part 493 and published accreditation guidance, reviewed November 26, 2026. Not affiliated with or endorsed by COLA Inc., CMS or CDC; COLA is a registered trademark of COLA Inc. Not legal advice. Always follow your current COLA Accreditation Manual and the instructions in your survey report.