CLIA Competency Assessment for Supervisors and Technical Consultants Course
How to run CLIA-compliant competency assessments: who may assess, the six required methods, timing, test systems, documentation and remediation.
Free40 min5 lessons10-question quiz
Lessons
Lesson 1: Training versus competency
Training teaches a person to perform a test. Competency assessment confirms they can perform it correctly in your lab with your test systems. CLIA requires both for non-waived testing, and inspectors frequently cite missing or incomplete competency assessments.
Lesson 2: Who may assess
- Moderate complexity: the technical consultant (42 CFR 493.1413), or the laboratory director.
- High complexity: the technical supervisor (42 CFR 493.1451); the general supervisor may assess testing personnel when delegated in writing.
- Assessors must meet the qualifications for their role, and delegation must be documented.
Lesson 3: When
Assess each person semiannually during the first year they test patient specimens (at about 6 and 12 months), then at least annually. Reassess when methods or instruments change and before the person resumes testing after problems are identified.
Lesson 4: The six required methods
- Direct observation of routine patient test performance, including patient preparation, specimen handling, processing and testing.
- Monitoring the recording and reporting of test results, including critical results.
- Review of intermediate test results or worksheets, QC records, PT results and preventive maintenance records.
- Direct observation of performance of instrument maintenance and function checks.
- Assessment of test performance through testing previously analyzed specimens, internal blind samples or external PT samples.
- Evaluation of problem-solving skills.
All six apply to each test system the person uses, unless a method does not apply to that test (document why). Test systems that share the same instrument and methodology can sometimes be grouped; follow CMS guidance.
Lesson 5: Documentation and remediation
- Record the date, test system, each method used, the assessor and the outcome, with the employee's and assessor's signatures.
- If performance is not acceptable, retrain, reassess and document before the person returns to independent testing, and review patient results they reported if quality may have been affected.
- Keep records for the period required by CLIA, your state and accreditor.
Quiz and certificate
Answer all 10 questions. Score 80% or higher to get your free certificate from Sound Lab Consultants - Technical Consultants.
Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy. This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.