Quality assessment (QA) citations
A quality assessment citation means the lab could not show an ongoing, documented process to monitor, review and correct problems across its preanalytic, analytic and postanalytic systems. A QA plan that exists on paper but has no dated reviews, no findings and no corrective actions is the most common reason.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1200Introduction | Introduction to quality systems |
| 42 CFR § 493.1239Standard: General laboratory systems quality assessment42 CFR § 493.1249Standard: Preanalytic systems quality assessment42 CFR § 493.1289Standard: Analytic systems quality assessment42 CFR § 493.1299Standard: Postanalytic systems quality assessment | Quality assessment for general, preanalytic, analytic and postanalytic systems |
| 42 CFR § 493.1233Standard: Complaint investigations | Complaint investigations |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1200 Introduction.
(a) Each laboratory that performs nonwaived testing must establish and maintain written policies and procedures that implement and monitor a quality system for all phases of the total testing process (that is, preanalytic, analytic, and postanalytic) as well as general laboratory systems.
(b) The laboratory's quality systems must include a quality assessment component that ensures continuous improvement of the laboratory's performance and services through ongoing monitoring that identifies, evaluates and resolves problems.
(c) The various components of the laboratory's quality system are used to meet the requirements in this part and must be appropriate for the specialties and subspecialties of testing the laboratory performs, services it offers, and clients it serves.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1239 Standard: General laboratory systems quality assessment.
(a) The laboratory must establish and follow written policies and procedures for an ongoing mechanism to monitor, assess, and, when indicated, correct problems identified in the general laboratory systems requirements specified at §§ 493.1231 through 493.1236.
(b) The general laboratory systems quality assessment must include a review of the effectiveness of corrective actions taken to resolve problems, revision of policies and procedures necessary to prevent recurrence of problems, and discussion of general laboratory systems quality assessment reviews with appropriate staff.
(c) The laboratory must document all general laboratory systems quality assessment activities.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1249 Standard: Preanalytic systems quality assessment.
(a) The laboratory must establish and follow written policies and procedures for an ongoing mechanism to monitor, assess, and when indicated, correct problems identified in the preanalytic systems specified at §§ 493.1241 through 493.1242.
(b) The preanalytic systems quality assessment must include a review of the effectiveness of corrective actions taken to resolve problems, revision of policies and procedures necessary to prevent recurrence of problems, and discussion of preanalytic systems quality assessment reviews with appropriate staff.
(c) The laboratory must document all preanalytic systems quality assessment activities.
[68 FR 3703, Jan. 24, 2003; 68 FR 3703, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1289 Standard: Analytic systems quality assessment.
(a) The laboratory must establish and follow written policies and procedures for an ongoing mechanism to monitor, assess, and when indicated, correct problems identified in the analytic systems specified in §§ 493.1251 through 493.1283.
(b) The analytic systems quality assessment must include a review of the effectiveness of corrective actions taken to resolve problems, revision of policies and procedures necessary to prevent recurrence of problems, and discussion of analytic systems quality assessment reviews with appropriate staff.
(c) The laboratory must document all analytic systems quality assessment activities.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1299 Standard: Postanalytic systems quality assessment.
(a) The laboratory must establish and follow written policies and procedures for an ongoing mechanism to monitor, assess and, when indicated, correct problems identified in the postanalytic systems specified in § 493.1291.
(b) The postanalytic systems quality assessment must include a review of the effectiveness of corrective actions taken to resolve problems, revision of policies and procedures necessary to prevent recurrence of problems, and discussion of postanalytic systems quality assessment reviews with appropriate staff.
(c) The laboratory must document all postanalytic systems quality assessment activities.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1233 Standard: Complaint investigations.
The laboratory must have a system in place to ensure that it documents all complaints and problems reported to the laboratory. The laboratory must conduct investigations of complaints, when appropriate.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
What an effective QA program shows
- Defined indicators, such as rejected specimens, turnaround time, corrected reports and QC failures.
- Dated reviews, monthly or quarterly, signed by the director or a delegate.
- Problems found, causes identified, actions taken and follow-up checked.
- Complaint and incident log with investigations.
- Review of PT, QC, competency and maintenance trends.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
How often should QA be reviewed?
CLIA says ongoing and does not set a single interval. Most labs document a monthly or quarterly review and adjust based on test volume and risk.