IQCP: the Individualized Quality Control Plan, explained

When you can run less external QC, and what CMS expects in the plan.

Short answer: IQCP is an optional CLIA quality control approach for non-waived tests. Instead of the default of two levels of external QC each day of testing (42 CFR 493.1256), a lab can document a risk-based plan and run QC at the frequency the plan supports. It does not apply to waived tests, which follow the manufacturer's instructions.

The three parts of an IQCP

  1. Risk assessment covering the specimen, test system, reagents, environment and testing personnel, across the pre-analytic, analytic and post-analytic phases.
  2. Quality control plan: the controls you will use (external, internal or built-in), how often, and what you do when they fail. It must meet at least the manufacturer's instructions.
  3. Quality assessment: ongoing monitoring (QC failures, complaints, corrected reports, PT results) to confirm the plan works, with the director's review.

Worked example: cartridge-based blood gas analyzer

Risks identified: clotted samples, cartridge storage temperature, operator technique. Controls: built-in electronic checks every run; liquid QC on each new lot and shipment and at the interval the risk assessment supports; temperature logs; operator competency. Quality assessment: monthly review of QC failures and error codes, quarterly director sign-off.