IQCP: the Individualized Quality Control Plan, explained
When you can run less external QC, and what CMS expects in the plan.
The three parts of an IQCP
- Risk assessment covering the specimen, test system, reagents, environment and testing personnel, across the pre-analytic, analytic and post-analytic phases.
- Quality control plan: the controls you will use (external, internal or built-in), how often, and what you do when they fail. It must meet at least the manufacturer's instructions.
- Quality assessment: ongoing monitoring (QC failures, complaints, corrected reports, PT results) to confirm the plan works, with the director's review.
Worked example: cartridge-based blood gas analyzer
Risks identified: clotted samples, cartridge storage temperature, operator technique. Controls: built-in electronic checks every run; liquid QC on each new lot and shipment and at the interval the risk assessment supports; temperature logs; operator competency. Quality assessment: monthly review of QC failures and error codes, quarterly director sign-off.