Reviewed November 26, 2026 by the Sound Lab Consultants laboratory compliance team
Short answer

A method verification citation means the lab began reporting patient results on a new test or instrument without documenting that it verified the manufacturer's performance claims. For unmodified FDA-cleared or approved tests, that means accuracy, precision, reportable range and reference intervals. Modified or lab-developed tests need full validation.

Regulations behind this citation

RegulationWhat it requires
42 CFR § 493.1253(b)(1)Standard: Establishment and verification of performance specificationsVerification for unmodified FDA-cleared or approved tests
42 CFR § 493.1253(b)(2)Standard: Establishment and verification of performance specificationsEstablishment for modified tests and LDTs

Full text of the regulations

Tap a regulation to read the complete official text. Each one links to its section on eCFR.

§ 493.1253 Standard: Establishment and verification of performance specifications.

(a) Applicability. Laboratories are not required to verify or establish performance specifications for any test system used by the laboratory before April 24, 2003.

(b)(1) Verification of performance specifications. Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results:

(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics:

(A) Accuracy.

(B) Precision.

(C) Reportable range of test results for the test system.

(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population.

(2) Establishment of performance specifications. Each laboratory that modifies an FDA-cleared or approved test system, or introduces a test system not subject to FDA clearance or approval (including methods developed in-house and standardized methods such as text book procedures), or uses a test system in which performance specifications are not provided by the manufacturer must, before reporting patient test results, establish for each test system the performance specifications for the following performance characteristics, as applicable:

(i) Accuracy.

(ii) Precision.

(iii) Analytical sensitivity.

(iv) Analytical specificity to include interfering substances.

(v) Reportable range of test results for the test system.

(vi) Reference intervals (normal values).

(vii) Any other performance characteristic required for test performance.

(3) Determination of calibration and control procedures. The laboratory must determine the test system's calibration procedures and control procedures based upon the performance specifications verified or established under paragraph (b)(1) or (b)(2) of this section.

(c) Documentation. The laboratory must document all activities specified in this section.

[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

Verification checklist

CharacteristicTypical evidence
AccuracyComparison with a reference method, PT samples or prior patient samples
PrecisionRepeat runs over several days and at more than one level
Reportable rangeLinearity or calibration verification material across the claimed range
Reference intervalsVerified as suitable for your patient population
Director approvalSigned and dated before patient reporting starts

Common gaps

Example plan of correction

A model response to a typical citation in this area. Adapt it to your lab's facts and records.

Deficiency cited
A new hematology analyzer was used for patient testing without documented verification of performance specifications.
Root cause & correction
Root cause: the vendor installation report was mistaken for a lab verification. Correction: a verification study was completed and approved by the director, and patient results reported during the gap were reviewed.
Evidence attached
Verification study, director approval page and patient lookback summary.
Monitoring
A new-instrument checklist requires director sign-off before go-live.

Frequently asked questions

Is the vendor's installation report enough?

Usually not. The lab must document its own verification and have the director approve it before reporting patient results.

Do waived tests need method verification?

CLIA's performance specification rules apply to nonwaived tests. Follow the manufacturer's instructions for waived tests.

Related citation guides

Cited on your last survey?

Our lab consultants write plans of correction for COLA and CMS-2567 citations and fixes the root cause so the citation doesn't come back. MyLabCompliance.io keeps competency, QC, PT, calibration and procedures survey-ready.

Book a free consultationSee MyLabCompliance.io

Or call (602) 874-1311

These guides are independent educational content from Sound Lab Consultants, based on 42 CFR Part 493 and published accreditation guidance, reviewed November 26, 2026. Not affiliated with or endorsed by COLA Inc., CMS or CDC; COLA is a registered trademark of COLA Inc. Not legal advice. Always follow your current COLA Accreditation Manual and the instructions in your survey report.