Method verification (performance specification) citations
A method verification citation means the lab began reporting patient results on a new test or instrument without documenting that it verified the manufacturer's performance claims. For unmodified FDA-cleared or approved tests, that means accuracy, precision, reportable range and reference intervals. Modified or lab-developed tests need full validation.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1253(b)(1)Standard: Establishment and verification of performance specifications | Verification for unmodified FDA-cleared or approved tests |
| 42 CFR § 493.1253(b)(2)Standard: Establishment and verification of performance specifications | Establishment for modified tests and LDTs |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1253 Standard: Establishment and verification of performance specifications.
(a) Applicability. Laboratories are not required to verify or establish performance specifications for any test system used by the laboratory before April 24, 2003.
(b)(1) Verification of performance specifications. Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results:
(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics:
(A) Accuracy.
(B) Precision.
(C) Reportable range of test results for the test system.
(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population.
(2) Establishment of performance specifications. Each laboratory that modifies an FDA-cleared or approved test system, or introduces a test system not subject to FDA clearance or approval (including methods developed in-house and standardized methods such as text book procedures), or uses a test system in which performance specifications are not provided by the manufacturer must, before reporting patient test results, establish for each test system the performance specifications for the following performance characteristics, as applicable:
(i) Accuracy.
(ii) Precision.
(iii) Analytical sensitivity.
(iv) Analytical specificity to include interfering substances.
(v) Reportable range of test results for the test system.
(vi) Reference intervals (normal values).
(vii) Any other performance characteristic required for test performance.
(3) Determination of calibration and control procedures. The laboratory must determine the test system's calibration procedures and control procedures based upon the performance specifications verified or established under paragraph (b)(1) or (b)(2) of this section.
(c) Documentation. The laboratory must document all activities specified in this section.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
Verification checklist
| Characteristic | Typical evidence |
|---|---|
| Accuracy | Comparison with a reference method, PT samples or prior patient samples |
| Precision | Repeat runs over several days and at more than one level |
| Reportable range | Linearity or calibration verification material across the claimed range |
| Reference intervals | Verified as suitable for your patient population |
| Director approval | Signed and dated before patient reporting starts |
Common gaps
- Data collected but the summary was never approved by the director.
- Replacement instruments put into use without verification.
- No comparison between the old and new instrument before the switch.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
Is the vendor's installation report enough?
Usually not. The lab must document its own verification and have the director approve it before reporting patient results.
Do waived tests need method verification?
CLIA's performance specification rules apply to nonwaived tests. Follow the manufacturer's instructions for waived tests.