Laboratory director responsibility citations
A director responsibility citation is issued when a duty CLIA assigns to the laboratory director was not performed or documented, including when the director delegated it without written records. Director duties appear in the LDR section of the COLA Accreditation Manual. The director can delegate tasks but remains responsible for all of them.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1407Standard; Laboratory director responsibilities | Laboratory director responsibilities (moderate complexity) |
| 42 CFR § 493.1445Standard; Laboratory director responsibilities | Laboratory director responsibilities (high complexity) |
| 42 CFR § 493.1251(d)Standard: Procedure manual | Director approves, signs and dates procedures |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1407 Standard; Laboratory director responsibilities.
The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, and record and report test results promptly, accurate, and proficiently and for assuring compliance with the applicable regulations.
(a) The laboratory director, if qualified, may perform the duties of the technical consultant, clinical consultant, and testing personnel, or delegate these responsibilities to personnel meeting the qualifications of §§ 493.1409, 493.1415, and 493.1421, respectively.
(b) If the laboratory director reapportions performance of his or her responsibilities, he or she remains responsible for ensuring that all duties are properly performed.
(c) The laboratory director must:
(1) Be onsite at least once every 6 months, with at least 4 months between the minimum two on-site visits. Laboratory directors may elect to be on-site more frequently and must continue to be accessible to the laboratory to provide telephone or electronic consultation as needed; and
(2) Provide documentation of these visits, including evidence of performing activities that are part of the laboratory director responsibilities.
(d) Each individual may direct no more than five laboratories.
(e) The laboratory director must—
(1) Ensure that testing systems developed and used for each of the tests performed in the laboratory provide quality laboratory services for all aspects of test performance, which includes the preanalytic, analytic, and postanalytic phases of testing;
(2) Ensure that the physical plant and environmental conditions of the laboratory are appropriate for the testing performed and provide a safe environment in which employees are protected from physical, chemical, and biological hazards;
(3) Ensure that—
(i) The test methodologies selected have the capability of providing the quality of results required for patient care;
(ii) Verification procedures used are adequate to determine the accuracy, precision, and other pertinent performance characteristics of the method; and
(iii) Laboratory personnel are performing the test methods as required for accurate and reliable results;
(4) Ensure that the laboratory is enrolled in an HHS approved proficiency testing program for the testing performed and that—
(i) The proficiency testing samples are tested as required under subpart H of this part;
(ii) The results are returned within the timeframes established by the proficiency testing program;
(iii) All proficiency testing reports received are reviewed by the appropriate staff to evaluate the laboratory's performance and to identify any problems that require corrective action; and
(iv) An approved corrective action plan is followed when any proficiency testing results are found to be unacceptable or unsatisfactory;
(5) Ensure that the quality control and quality assessment programs are established and maintained to assure the quality of laboratory services provided and to identify failures in quality as they occur;
(6) Ensure the establishment and maintenance of acceptable levels of analytical performance for each test system;
(7) Ensure that all necessary remedial actions are taken and documented whenever significant deviations from the laboratory's established performance specifications are identified, and that patient test results are reported only when the system is functioning properly;
(8) Ensure that reports of test results include pertinent information required for interpretation;
(9) Ensure that consultation is available to the laboratory's clients on matters relating to the quality of the test results reported and their interpretation concerning specific patient conditions;
(10) Employ a sufficient number of laboratory personnel with the appropriate education and either experience or training to provide appropriate consultation, properly supervise and accurately perform tests and report test results in accordance with the personnel responsibilities described in this subpart;
(11) Ensure that prior to testing patients' specimens, all personnel have the appropriate education and experience, receive the appropriate training for the type and complexity of the services offered, and have demonstrated that they can perform all testing operations reliably to provide and report accurate results;
(12) Ensure that policies and procedures are established for monitoring individuals who conduct preanalytical, analytical, and postanalytical phases of testing to assure that they are competent and maintain their competency to process specimens, perform test procedures and report test results promptly and proficiently, and whenever necessary, identify needs for remedial training or continuing education to improve skills;
(13) Ensure that an approved procedure manual is available to all personnel responsible for any aspect of the testing process; and
(14) Specify, in writing, the responsibilities and duties of each consultant and each person, engaged in the performance of the preanalytic, analytic, and postanalytic phases of testing, that identifies which examinations and procedures each individual is authorized to perform, whether supervision is required for specimen processing, test performance or results reporting, and whether consultant or director review is required prior to reporting patient test results.
[57 FR 7172, Feb. 28, 1992, as amended at 68 FR 3713, Jan. 24, 2003; 88 FR 90039, Dec. 28, 2023]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1445 Standard; Laboratory director responsibilities.
The laboratory director is responsible for the overall operation and administration of the laboratory, including the employment of personnel who are competent to perform test procedures, record and report test results promptly, accurately and proficiently, and for assuring compliance with the applicable regulations.
(a) The laboratory director, if qualified, may perform the duties of the technical supervisor, clinical consultant, general supervisor, and testing personnel, or delegate these responsibilities to personnel meeting the qualifications under §§ 493.1447, 493.1453, 493.1459, and 493.1487, respectively.
(b) If the laboratory director reapportions performance of his or her responsibilities, he or she remains responsible for ensuring that all duties are properly performed.
(c) The laboratory director must:
(1) Be onsite at least once every 6 months, with at least 4 months between the minimum two on-site visits. Laboratory directors may elect to be on-site more frequently and must continue to be accessible to the laboratory to provide telephone or electronic consultation as needed; and
(2) Provide documentation of these visits, including evidence of performing activities that are part of the laboratory director responsibilities.
(d) Each individual may direct no more than five laboratories.
(e) The laboratory director must—
(1) Ensure that testing systems developed and used for each of the tests performed in the laboratory provide quality laboratory services for all aspects of test performance, which includes the preanalytic, analytic, and postanalytic phases of testing;
(2) Ensure that the physical plant and environmental conditions of the laboratory are appropriate for the testing performed and provide a safe environment in which employees are protected from physical, chemical, and biological hazards;
(3) Ensure that—
(i) The test methodologies selected have the capability of providing the quality of results required for patient care;
(ii) Verification procedures used are adequate to determine the accuracy, precision, and other pertinent performance characteristics of the method; and
(iii) Laboratory personnel are performing the test methods as required for accurate and reliable results;
(4) Ensure that the laboratory is enrolled in an HHS-approved proficiency testing program for the testing performed and that—
(i) The proficiency testing samples are tested as required under subpart H of this part;
(ii) The results are returned within the timeframes established by the proficiency testing program;
(iii) All proficiency testing reports received are reviewed by the appropriate staff to evaluate the laboratory's performance and to identify any problems that require corrective action; and
(iv) An approved corrective action plan is followed when any proficiency testing result is found to be unacceptable or unsatisfactory;
(5) Ensure that the quality control and quality assessment programs are established and maintained to assure the quality of laboratory services provided and to identify failures in quality as they occur;
(6) Ensure the establishment and maintenance of acceptable levels of analytical performance for each test system;
(7) Ensure that all necessary remedial actions are taken and documented whenever significant deviations from the laboratory's established performance characteristics are identified, and that patient test results are reported only when the system is functioning properly;
(8) Ensure that reports of test results include pertinent information required for interpretation;
(9) Ensure that consultation is available to the laboratory's clients on matters relating to the quality of the test results reported and their interpretation concerning specific patient conditions;
(10) Ensure that a general supervisor provides on-site supervision of high complexity test performance by testing personnel qualified under § 493.1489(b)(5);
(11) Employ a sufficient number of laboratory personnel with the appropriate education and either experience or training to provide appropriate consultation, properly supervise and accurately perform tests and report test results in accordance with the personnel responsibilities described in this subpart;
(12) Ensure that prior to testing patients' specimens, all personnel have the appropriate education and experience, receive the appropriate training for the type and complexity of the services offered, and have demonstrated that they can perform all testing operations reliably to provide and report accurate results;
(13) Ensure that policies and procedures are established for monitoring individuals who conduct preanalytical, analytical, and postanalytical phases of testing to assure that they are competent and maintain their competency to process specimens, perform test procedures and report test results promptly and proficiently, and whenever necessary, identify needs for remedial training or continuing education to improve skills;
(14) Ensure that an approved procedure manual is available to all personnel responsible for any aspect of the testing process; and
(15) Specify, in writing, the responsibilities and duties of each consultant and each supervisor, as well as each person engaged in the performance of the preanalytic, analytic, and postanalytic phases of testing, that identifies which examinations and procedures each individual is authorized to perform, whether supervision is required for specimen processing, test performance or result reporting and whether supervisory or director review is required prior to reporting patient test results.
[57 FR 7172, Feb. 28, 1992, as amended at 68 FR 3714, Jan. 24, 2003; 88 FR 90041, Dec. 28, 2023]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1251 Standard: Procedure manual.
(a) A written procedure manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens.
(b) The procedure manual must include the following when applicable to the test procedure:
(1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in § 493.1242.
(2) Microscopic examination, including the detection of inadequately prepared slides.
(3) Step-by-step performance of the procedure, including test calculations and interpretation of results.
(4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing.
(5) Calibration and calibration verification procedures.
(6) The reportable range for test results for the test system as established or verified in § 493.1253.
(7) Control procedures.
(8) Corrective action to take when calibration or control results fail to meet the laboratory's criteria for acceptability.
(9) Limitations in the test methodology, including interfering substances.
(10) Reference intervals (normal values).
(11) Imminently life-threatening test results, or panic or alert values.
(12) Pertinent literature references.
(13) The laboratory's system for entering results in the patient record and reporting patient results including, when appropriate, the protocol for reporting imminently life-threatening results, or panic, or alert values.
(14) Description of the course of action to take if a test system becomes inoperable.
(c) Manufacturer's test system instructions or operator manuals may be used, when applicable, to meet the requirements of paragraphs (b)(1) through (b)(12) of this section. Any of the items under paragraphs (b)(1) through (b)(12) of this section not provided by the manufacturer must be provided by the laboratory.
(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use.
(e) The laboratory must maintain a copy of each procedure with the dates of initial use and discontinuance as described in § 493.1105(a)(2).
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
Duties that most often lead to citations
- Procedures not approved, signed and dated before use, or not re-approved by a new director.
- Delegation of duties not documented in writing.
- No evidence the director reviewed QC, PT, QA or competency.
- Onsite visits not documented.
- Directing more than five laboratories.
- Staff roles and responsibilities not defined in writing.
Onsite visit rule (current)
Under personnel rules effective December 28, 2024, a director who is not onsite regularly must visit at least once every six months, with at least four months between visits, and document what was done at each visit. A proposed rule published in October 2026 would remove this visit requirement. It is not final, so keep documenting visits. See the 2026 proposed CLIA rule.
Build a delegation file
Keep one signed delegation document that lists each duty, the qualified person it is assigned to, and how the director checks on it. Add a quarterly director review log. Surveyors look for those two documents first.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
What laboratory director duties cannot be delegated?
Some duties, such as approving, signing and dating procedures, cannot be delegated. For any duty that can be delegated, the director stays responsible and must make sure it is done.
How many labs can one director oversee?
CLIA limits an individual to directing no more than five laboratories.
Does the director have to visit the lab in person?
Under the rules effective December 28, 2024, yes: at least once every six months, with at least four months between visits, and documented. A 2026 proposed rule would remove this, but it is not final.