Condition-level vs standard-level deficiencies
A standard-level deficiency is a single requirement that was not met and is not serious on its own. A condition-level deficiency means the problem is serious or systemic enough to threaten the accuracy of patient results, and it can lead to sanctions such as a directed plan of correction, civil money penalties, or limits on the CLIA certificate.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1806Available sanctions: All laboratories | Available sanctions: all laboratories |
| 42 CFR § 493.1832Directed plan of correction and directed portion of a plan of correction | Directed plan of correction |
| 42 CFR § 493.1840Suspension, limitation, or revocation of any type of CLIA certificate | Suspension, limitation or revocation of a CLIA certificate |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1806 Available sanctions: All laboratories.
(a) Applicability. CMS may impose one or more of the sanctions specified in this section on a laboratory that is out of compliance with one or more CLIA conditions.
(b) Principal sanction. CMS may impose any of the three principal CLIA sanctions, which are suspension, limitation, or revocation of any type of CLIA certificate.
(c) Alternative sanctions. CMS may impose one or more of the following alternative sanctions in lieu of or in addition to imposing a principal sanction, except on a laboratory that has a certificate of waiver.
(1) Directed plan of correction, as set forth at § 493.1832.
(2) State onsite monitoring as set forth at § 493.1836.
(3) Civil money penalty, as set forth at § 493.1834.
(d) Civil suit. CMS may bring suit in the appropriate U.S. District Court to enjoin continuation of any activity of any laboratory (including a CLIA-exempt laboratory that has been found with deficiencies during a validation survey), if CMS has reason to believe that continuation of the activity would constitute a significant hazard to the public health.
(e) Criminal sanctions. Under section 353(1) of the PHS Act, an individual who is convicted of intentionally violating any CLIA requirement may be imprisoned or fined.
[57 FR 7237, Feb. 28, 1992, as amended at 58 FR 5237, Jan. 19, 1993]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1832 Directed plan of correction and directed portion of a plan of correction.
(a) Application. CMS may impose a directed plan of correction as an alternative sanction for any laboratory that has condition level deficiencies. If CMS does not impose a directed plan of correction as an alternative sanction for a laboratory that has condition level deficiencies, it at least imposes a directed portion of a plan of correction when it imposes any of the following alternative sanctions:
(1) State onsite monitoring.
(2) Civil money penalty.
(3) Suspension of all or part of Medicare payments.
(b) Procedures—(1) Directed plan of correction. When imposing this sanction, CMS—
(i) Gives the laboratory prior notice of the sanction and opportunity to respond in accordance with § 493.1810;
(ii) Directs the laboratory to take specific corrective action within specific time frames in order to achieve compliance; and
(iii) May direct the laboratory to submit the names of laboratory clients for notification purposes, as specified in paragraph (b)(3) of this section.
(2) Directed portion of a plan of correction. CMS may decide to notify clients of a sanctioned laboratory, because of the seriousness of the noncompliance (e.g., the existence of immediate jeopardy) or for other reasons. When imposing this sanction, CMS takes the following steps—
(i) Directs the laboratory to submit to CMS, the State survey agency, or other CMS agent, within 10 calendar days after the notice of the alternative sanction, a list of names and addresses of all physicians, providers, suppliers, and other clients who have used some or all of the services of the laboratory since the last certification inspection or within any other timeframe specified by CMS.
(ii) Within 30 calendar days of receipt of the information, may send to each laboratory client, via the State survey agency, a notice containing the name and address of the laboratory, the nature of the laboratory's noncompliance, and the kind and effective date of the alternative sanction.
(iii) Sends to each laboratory client, via the State survey agency, notice of the recission of an adverse action within 30 days of the rescission.
(3) Notice of imposition of a principal sanction following the imposition of an alternative sanction. If CMS imposes a principal sanction following the imposition of an alternative sanction, and for which CMS has already obtained a list of laboratory clients, CMS may use that list to notify the clients of the imposition of the principal sanction.
(c) Duration of a directed plan of correction. If CMS imposes a directed plan of correction, and on revisit it is found that the laboratory has not corrected the deficiencies within 12 months from the last day of inspection, the following rules apply:
(1) CMS cancels the laboratory's approval for Medicare payment of its services, and notifies the laboratory of CMS's intent to suspend, limit, or revoke the laboratory's CLIA certificate.
(2) The directed plan of correction continues in effect until the day suspension, limitation, or revocation of the laboratory's CLIA certificate.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1840 Suspension, limitation, or revocation of any type of CLIA certificate.
(a) Adverse action based on actions of the laboratory's owner, operator or employees. CMS may initiate adverse action to suspend, limit or revoke any CLIA certificate if CMS finds that a laboratory's owner or operator or one of its employees has—
(1) Been guilty of misrepresentation in obtaining a CLIA certificate;
(2) Performed, or represented the laboratory as entitled to perform, a laboratory examination or other procedure that is not within a category of laboratory examinations or other procedures authorized by its CLIA certificate;
(3) Failed to comply with the certificate requirements and performance standards;
(4) Failed to comply with reasonable requests by CMS for any information or work on materials that CMS concludes is necessary to determine the laboratory's continued eligibility for its CLIA certificate or continued compliance with performance standards set by CMS;
(5) Refused a reasonable request by CMS or its agent for permission to inspect the laboratory and its operation and pertinent records during the hours that the laboratory is in operation;
(6) Violated or aided and abetted in the violation of any provisions of CLIA and its implementing regulations;
(7) Failed to comply with an alternative sanction imposed under this subpart; or
(8) Within the preceding two-year period, owned or operated a laboratory that had its CLIA certificate revoked. (This provision applies only to the owner or operator, not to all of the laboratory's employees.)
(b) Adverse action based on improper referrals in proficiency testing. If CMS determines that a laboratory has intentionally referred its proficiency testing samples to another laboratory for analysis, CMS does one of the following:
(1)(i) Revokes the laboratory's CLIA certificate for at least 1 year, prohibits the owner and operator from owning or operating a CLIA-certified laboratory for at least 1 year, and may impose a civil money penalty in accordance with § 493.1834(d), if CMS determines that—
(A) A proficiency testing referral is a repeat proficiency testing referral as defined at § 493.2; or
(B) On or before the proficiency testing event close date, a laboratory reported proficiency testing results obtained from another laboratory to the proficiency testing program.
(ii) Following the revocation of a CLIA certificate in accordance with paragraph (b)(1)(i) of this section, CMS may exempt a laboratory owner from the generally applicable prohibition on owning or operating a CLIA-certified laboratory under paragraph (a)(8) of this section on a laboratory-by-laboratory basis if CMS finds, after review of the relevant facts and circumstances, that there is no evidence that—
(A) Patients would be put at risk as a result of the owner being exempted from the ban on a laboratory-by-laboratory basis;
(B) The laboratory for which the owner is to be exempted from the general ownership ban participated in or was otherwise complicit in the PT referral of the laboratory that resulted in the revocation; and
(C) The laboratory for which the owner is to be exempted from the general ownership ban received a PT sample from another laboratory in the prior two survey cycles, and failed to immediately report such receipt to CMS or to the appropriate CMS-approved accrediting organization.
(2) Suspends or limits the CLIA certificate for less than 1 year based on the criteria in § 493.1804(d) and imposes alternative sanctions as appropriate, in accordance with §§ 493.1804(c) and (d), 493.1806(c), 493.1807(b), 493.1809 and, in the case of civil money penalties, § 493.1834(d), when CMS determines that paragraph (b)(1)(i)(A) or (B) of this section does not apply but that the laboratory obtained test results for the proficiency testing samples from another laboratory on or before the proficiency testing event close date. Among other possibilities, alternative sanctions will always include a civil money penalty and a directed plan of correction that includes required training of staff.
(3) Imposes alternative sanctions in accordance with §§ 493.1804(c) and (d), 493.1806(c), 493.1807(b), 493.1809 and, in the case of civil money penalties, § 493.1834(d), when CMS determines that paragraph (b)(1)(i) or (2) of this section do not apply, and a PT referral has occurred, but no test results are received prior to the event close date by the referring laboratory from the laboratory that received the referral. Among other possibilities, alternative sanctions will always include a civil money penalty and a directed plan of correction that includes required training of staff.
(c) Adverse action based on exclusion from Medicare. If the OIG excludes a laboratory from participation in Medicare, CMS suspends the laboratory's CLIA certificate for the period during which the laboratory is excluded.
(d) Procedures for suspension or limitation—(1) Basic rule. Except as provided in paragraph (d)(2) of this section, CMS does not suspend or limit a CLIA certificate until after an ALJ hearing decision (as provided in § 493.1844) that upholds suspension or limitation.
(2) Exceptions. CMS may suspend or limit a CLIA certificate before the ALJ hearing in any of the following circumstances:
(i) The laboratory's deficiencies pose immediate jeopardy.
(ii) The laboratory has refused a reasonable request for information or work on materials.
(iii) The laboratory has refused permission for CMS or a CMS agent to inspect the laboratory or its operation.
(e) Procedures for revocation. (1) CMS does not revoke any type of CLIA certificate until after an ALJ hearing that upholds revocation.
(2) CMS may revoke a CLIA certificate after the hearing decision even if it had not previously suspended or limited that certificate.
(f) Notice to the OIG. CMS notifies the OIG of any violations under paragraphs (a)(1), (a)(2), (a)(6), and (b) of this section within 30 days of the determination of the violation.
[57 FR 7237, Feb. 28, 1992, as amended at 79 FR 25480, May 2, 2014]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
How they relate
CLIA is organized into conditions (broad requirements, such as proficiency testing or laboratory director) and standards (specific requirements underneath each condition). Several standard-level problems in the same area, or one problem with patient impact, can add up to a condition-level finding.
| Standard-level | Condition-level | |
|---|---|---|
| Meaning | One requirement not met | Serious or systemic failure |
| Patient risk | Low or indirect | Actual or potential harm to results |
| Typical outcome | Plan of correction | Plan of correction plus possible sanctions |
| Example | One late competency | No competency program at all |
Where COLA fits
COLA writes citations against its own accreditation criteria, which correspond to CLIA. A lab with serious COLA findings can face directed corrections, restrictions on testing, or loss of accreditation. CMS also runs validation and complaint surveys of accredited labs, and those findings are written on Form CMS-2567 using D-tags.
Frequently asked questions
What is an immediate jeopardy finding in a lab?
It is a situation where noncompliance has caused, or is likely to cause, serious harm. It calls for urgent action, and sanctions can follow quickly.
What are D-tags?
D-tags are the codes CMS surveyors use on Form CMS-2567 to identify each CLIA regulation that was cited.