Procedure manual citations

Reviewed November 26, 2026 by the Sound Lab Consultants laboratory compliance team
Short answer

A procedure manual citation means a written procedure was missing, incomplete, or not approved, signed and dated by the current laboratory director before it was used. A new director must review and approve existing procedures, retired procedures must be kept for at least two years, and staff must follow the procedure as written.

Regulations behind this citation

RegulationWhat it requires
42 CFR § 493.1251Standard: Procedure manualProcedure manual
42 CFR § 493.1105Standard: Retention requirementsRetention requirements

Full text of the regulations

Tap a regulation to read the complete official text. Each one links to its section on eCFR.

§ 493.1251 Standard: Procedure manual.

(a) A written procedure manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel. Textbooks may supplement but not replace the laboratory's written procedures for testing or examining specimens.

(b) The procedure manual must include the following when applicable to the test procedure:

(1) Requirements for patient preparation; specimen collection, labeling, storage, preservation, transportation, processing, and referral; and criteria for specimen acceptability and rejection as described in § 493.1242.

(2) Microscopic examination, including the detection of inadequately prepared slides.

(3) Step-by-step performance of the procedure, including test calculations and interpretation of results.

(4) Preparation of slides, solutions, calibrators, controls, reagents, stains, and other materials used in testing.

(5) Calibration and calibration verification procedures.

(6) The reportable range for test results for the test system as established or verified in § 493.1253.

(7) Control procedures.

(8) Corrective action to take when calibration or control results fail to meet the laboratory's criteria for acceptability.

(9) Limitations in the test methodology, including interfering substances.

(10) Reference intervals (normal values).

(11) Imminently life-threatening test results, or panic or alert values.

(12) Pertinent literature references.

(13) The laboratory's system for entering results in the patient record and reporting patient results including, when appropriate, the protocol for reporting imminently life-threatening results, or panic, or alert values.

(14) Description of the course of action to take if a test system becomes inoperable.

(c) Manufacturer's test system instructions or operator manuals may be used, when applicable, to meet the requirements of paragraphs (b)(1) through (b)(12) of this section. Any of the items under paragraphs (b)(1) through (b)(12) of this section not provided by the manufacturer must be provided by the laboratory.

(d) Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use.

(e) The laboratory must maintain a copy of each procedure with the dates of initial use and discontinuance as described in § 493.1105(a)(2).

[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

§ 493.1105 Standard: Retention requirements.

(a) The laboratory must retain its records and, as applicable, slides, blocks, and tissues as follows:

(1) Test requisitions and authorizations. Retain records of test requisitions and test authorizations, including the patient's chart or medical record if used as the test requisition or authorization, for at least 2 years.

(2) Test procedures. Retain a copy of each test procedure for at least 2 years after a procedure has been discontinued. Each test procedure must include the dates of initial use and discontinuance.

(3) Analytic systems records. Retain quality control and patient test records (including instrument printouts, if applicable) and records documenting all analytic systems activities specified in §§ 493.1252 through 493.1289 for at least 2 years. In addition, retain the following:

(i) Records of test system performance specifications that the laboratory establishes or verifies under § 493.1253 for the period of time the laboratory uses the test system but no less than 2 years.

(ii) Immunohematology records, blood and blood product records, and transfusion records as specified in 21 CFR 606.160(b)(3)(ii), (b)(3)(iv), (b)(3)(v) and (d).

(4) Proficiency testing records. Retain all proficiency testing records for at least 2 years.

(5) Quality system assessment records. Retain all laboratory quality systems assessment records for at least 2 years.

(6) Test reports. Retain or be able to retrieve a copy of the original report (including final, preliminary, and corrected reports) at least 2 years after the date of reporting. In addition, retain the following:

(i) Immunohematology reports as specified in 21 CFR 606.160(d).

(ii) Pathology test reports for at least 10 years after the date of reporting.

(7) Slide, block, and tissue retention—(i) Slides. (A) Retain cytology slide preparations for at least 5 years from the date of examination (see § 493.1274(f) for proficiency testing exception).

(B) Retain histopathology slides for at least 10 years from the date of examination.

(ii) Blocks. Retain pathology specimen blocks for at least 2 years from the date of examination.

(iii) Tissue. Preserve remnants of tissue for pathology examination until a diagnosis is made on the specimen.

(b) If the laboratory ceases operation, the laboratory must make provisions to ensure that all records and, as applicable, slides, blocks, and tissue are retained and available for the time frames specified in this section.

[68 FR 3703, Jan. 24, 2003; 68 FR 50723, Aug. 22, 2003]

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

What each procedure should cover

Common citations

Example plan of correction

A model response to a typical citation in this area. Adapt it to your lab's facts and records.

Deficiency cited
Procedures were not approved by the current laboratory director after the director change.
Root cause & correction
Root cause: the director-change checklist did not include procedure review. Correction: the director reviewed, signed and dated every procedure.
Evidence attached
Procedure index with signature dates, and the revised director-change checklist.
Monitoring
Electronic document control alerts the director when a review is due.

Frequently asked questions

Can I use the package insert as my procedure?

A manufacturer's insert can serve as part of the procedure if it covers the required content and the director approves it, along with anything the lab does differently.

How long are retired procedures kept?

At least two years after they are discontinued, with the dates they were in use.

Related citation guides

Cited on your last survey?

Our lab consultants write plans of correction for COLA and CMS-2567 citations and fixes the root cause so the citation doesn't come back. MyLabCompliance.io keeps competency, QC, PT, calibration and procedures survey-ready.

Book a free consultationSee MyLabCompliance.io

Or call (602) 874-1311

These guides are independent educational content from Sound Lab Consultants, based on 42 CFR Part 493 and published accreditation guidance, reviewed November 26, 2026. Not affiliated with or endorsed by COLA Inc., CMS or CDC; COLA is a registered trademark of COLA Inc. Not legal advice. Always follow your current COLA Accreditation Manual and the instructions in your survey report.