Comparison of test results citations

Reviewed November 26, 2026 by the Sound Lab Consultants laboratory compliance team
Short answer

This citation means the lab performs the same test on more than one instrument, method or site but did not compare results at least twice a year to show they agree. A related citation covers tests with no approved PT program, where the lab must verify accuracy at least twice a year by another method.

Regulations behind this citation

RegulationWhat it requires
42 CFR § 493.1281Standard: Comparison of test resultsComparison of test results
42 CFR § 493.1236(c)Standard: Evaluation of proficiency testing performanceTwice-annual accuracy verification for tests without regulated PT

Full text of the regulations

Tap a regulation to read the complete official text. Each one links to its section on eCFR.

§ 493.1281 Standard: Comparison of test results.

(a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites.

(b) The laboratory must have a system to identify and assess patient test results that appear inconsistent with the following relevant criteria, when available:

(1) Patient age.

(2) Sex.

(3) Diagnosis or pertinent clinical data.

(4) Distribution of patient test results.

(5) Relationship with other test parameters.

(c) The laboratory must document all test result comparison activities.

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

§ 493.1236 Standard: Evaluation of proficiency testing performance.

(a) The laboratory must review and evaluate the results obtained on proficiency testing performed as specified in subpart H of this part.

(b) The laboratory must verify the accuracy of the following:

(1) Any analyte or subspecialty without analytes listed in subpart I of this part that is not evaluated or scored by a CMS-approved proficiency testing program.

(2) Any analyte, specialty or subspecialty assigned a proficiency testing score that does not reflect laboratory test performance (that is, when the proficiency testing program does not obtain the agreement required for scoring as specified in subpart I of this part, or the laboratory receives a zero score for nonparticipation, or late return of results).

(c) At least twice annually, the laboratory must verify the accuracy of the following:

(1) Any test or procedure it performs that is not included in subpart I of this part.

(2) Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program.

(d) All proficiency testing evaluation and verification activities must be documented.

Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.

How to document it

Example plan of correction

A model response to a typical citation in this area. Adapt it to your lab's facts and records.

Deficiency cited
Glucose results from the main analyzer and the backup analyzer were not compared twice a year.
Root cause & correction
Root cause: the backup was seen as an emergency-only unit. Correction: a comparison study was completed, and all dual-method analytes were added to the schedule.
Evidence attached
Comparison data with criteria and signature.
Monitoring
Semiannual comparisons are tracked in the compliance calendar.

Frequently asked questions

Does a backup instrument count?

Yes. If it is used for patient testing, its results must be compared with the primary method.

Related citation guides

Cited on your last survey?

Our lab consultants write plans of correction for COLA and CMS-2567 citations and fixes the root cause so the citation doesn't come back. MyLabCompliance.io keeps competency, QC, PT, calibration and procedures survey-ready.

Book a free consultationSee MyLabCompliance.io

Or call (602) 874-1311

These guides are independent educational content from Sound Lab Consultants, based on 42 CFR Part 493 and published accreditation guidance, reviewed November 26, 2026. Not affiliated with or endorsed by COLA Inc., CMS or CDC; COLA is a registered trademark of COLA Inc. Not legal advice. Always follow your current COLA Accreditation Manual and the instructions in your survey report.