Comparison of test results citations
This citation means the lab performs the same test on more than one instrument, method or site but did not compare results at least twice a year to show they agree. A related citation covers tests with no approved PT program, where the lab must verify accuracy at least twice a year by another method.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1281Standard: Comparison of test results | Comparison of test results |
| 42 CFR § 493.1236(c)Standard: Evaluation of proficiency testing performance | Twice-annual accuracy verification for tests without regulated PT |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1281 Standard: Comparison of test results.
(a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites.
(b) The laboratory must have a system to identify and assess patient test results that appear inconsistent with the following relevant criteria, when available:
(1) Patient age.
(2) Sex.
(3) Diagnosis or pertinent clinical data.
(4) Distribution of patient test results.
(5) Relationship with other test parameters.
(c) The laboratory must document all test result comparison activities.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1236 Standard: Evaluation of proficiency testing performance.
(a) The laboratory must review and evaluate the results obtained on proficiency testing performed as specified in subpart H of this part.
(b) The laboratory must verify the accuracy of the following:
(1) Any analyte or subspecialty without analytes listed in subpart I of this part that is not evaluated or scored by a CMS-approved proficiency testing program.
(2) Any analyte, specialty or subspecialty assigned a proficiency testing score that does not reflect laboratory test performance (that is, when the proficiency testing program does not obtain the agreement required for scoring as specified in subpart I of this part, or the laboratory receives a zero score for nonparticipation, or late return of results).
(c) At least twice annually, the laboratory must verify the accuracy of the following:
(1) Any test or procedure it performs that is not included in subpart I of this part.
(2) Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program.
(d) All proficiency testing evaluation and verification activities must be documented.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
How to document it
- List every analyte run on more than one system.
- Choose patient samples that cover the reportable range.
- Set acceptance criteria before testing.
- Record the results, the decision, the reviewer and any action taken.
- Repeat at least every six months.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
Does a backup instrument count?
Yes. If it is used for patient testing, its results must be compared with the primary method.