COLA competency assessment citations
A competency assessment citation means the lab could not show that each person performing nonwaived testing was assessed on all six CLIA elements, for each test system, at the required frequency, by a qualified assessor. Assessments are due twice in the first year of patient testing and once a year after that.
Regulations behind this citation
| Regulation | What it requires |
|---|---|
| 42 CFR § 493.1235Standard: Personnel competency assessment policies | Personnel competency assessment policies |
| 42 CFR § 493.1413(b)(8)Standard; Technical consultant responsibilities | Technical consultant evaluates competency (moderate complexity) |
| 42 CFR § 493.1451(b)(8)Standard: Technical supervisor responsibilities | Technical supervisor evaluates competency (high complexity) |
Full text of the regulations
Tap a regulation to read the complete official text. Each one links to its section on eCFR.
§ 493.1235 Standard: Personnel competency assessment policies.
As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1413 Standard; Technical consultant responsibilities.
The technical consultant is responsible for the technical and scientific oversight of the laboratory. The technical consultant is not required to be onsite at all times testing is performed; however, he or she must be available to the laboratory on an as needed basis to provide consultation, as specified in paragraph (a) of this section.
(a) The technical consultant must be accessible to the laboratory to provide on-site, telephone, or electronic consultation; and
(b) The technical consultant is responsible for—
(1) Selection of test methodology appropriate for the clinical use of the test results;
(2) Verification of the test procedures performed and the establishment of the laboratory's test performance characteristics, including the precision and accuracy of each test and test system;
(3) Enrollment and participation in an HHS approved proficiency testing program commensurate with the services offered;
(4) Establishing a quality control program appropriate for the testing performed and establishing the parameters for acceptable levels of analytic performance and ensuring that these levels are maintained throughout the entire testing process from the initial receipt of the specimen, through sample analysis and reporting of test results;
(5) Resolving technical problems and ensuring that remedial actions are taken whenever test systems deviate from the laboratory's established performance specifications;
(6) Ensuring that patient test results are not reported until all corrective actions have been taken and the test system is functioning properly;
(7) Identifying training needs and assuring that each individual performing tests receives regular in-service training and education appropriate for the type and complexity of the laboratory services performed;
(8) Evaluating the competency of all testing personnel and assuring that the staff maintain their competency to perform test procedures and report test results promptly, accurately and proficiently. The procedures for evaluation of the competency of the staff must include, but are not limited to—
(i) Direct observations of routine patient test performance, including patient preparation, if applicable, specimen handling, processing and testing;
(ii) Monitoring the recording and reporting of test results;
(iii) Review of intermediate test results or worksheets, quality control records, proficiency testing results, and preventive maintenance records;
(iv) Direct observation of performance of instrument maintenance and function checks;
(v) Assessment of test performance through testing previously analyzed specimens, internal blind testing samples or external proficiency testing samples; and
(vi) Assessment of problem solving skills; and
(9) Evaluating and documenting the performance of individuals responsible for moderate complexity testing at least semiannually during the first year the individual tests patient specimens. Thereafter, evaluations must be performed at least annually unless test methodology or instrumentation changes, in which case, prior to reporting patient test results, the individual's performance must be reevaluated to include the use of the new test methodology or instrumentation.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1451 Standard: Technical supervisor responsibilities.
The technical supervisor is responsible for the technical and scientific oversight of the laboratory. The technical supervisor is not required to be on site at all times testing is performed; however, he or she must be available to the laboratory on an as needed basis to provide supervision as specified in (a) of this section.
(a) The technical supervisor must be accessible to the laboratory to provide on-site, telephone, or electronic consultation; and
(b) The technical supervisor is responsible for—
(1) Selection of the test methodology that is appropriate for the clinical use of the test results;
(2) Verification of the test procedures performed and establishment of the laboratory's test performance characteristics, including the precision and accuracy of each test and test system;
(3) Enrollment and participation in an HHS approved proficiency testing program commensurate with the services offered;
(4) Establishing a quality control program appropriate for the testing performed and establishing the parameters for acceptable levels of analytic performance and ensuring that these levels are maintained throughout the entire testing process from the initial receipt of the specimen, through sample analysis and reporting of test results;
(5) Resolving technical problems and ensuring that remedial actions are taken whenever test systems deviate from the laboratory's established performance specifications;
(6) Ensuring that patient test results are not reported until all corrective actions have been taken and the test system is functioning properly;
(7) Identifying training needs and assuring that each individual performing tests receives regular in-service training and education appropriate for the type and complexity of the laboratory services performed;
(8) Evaluating the competency of all testing personnel and assuring that the staff maintain their competency to perform test procedures and report test results promptly, accurately and proficiently. The procedures for evaluation of the competency of the staff must include, but are not limited to—
(i) Direct observations of routine patient test performance, including patient preparation, if applicable, specimen handling, processing and testing;
(ii) Monitoring the recording and reporting of test results;
(iii) Review of intermediate test results or worksheets, quality control records, proficiency testing results, and preventive maintenance records;
(iv) Direct observation of performance of instrument maintenance and function checks;
(v) Assessment of test performance through testing previously analyzed specimens, internal blind testing samples or external proficiency testing samples; and
(vi) Assessment of problem solving skills; and
(9) Evaluating and documenting the performance of individuals responsible for high complexity testing at least semiannually during the first year the individual tests patient specimens. Thereafter, evaluations must be performed at least annually unless test methodology or instrumentation changes, in which case, prior to reporting patient test results, the individual's performance must be reevaluated to include the use of the new test methodology or instrumentation.
(c) In cytology, the technical supervisor or the individual qualified under § 493.1449(e)(2)—
(1) May perform the duties of the cytology general supervisor and the cytotechnologist, as specified in §§ 493.1471 and 493.1485, respectively;
(2) Must establish the workload limit for each individual examining slides;
(3) Must reassess the workload limit for each individual examining slides at least every 6 months and adjust as necessary;
(4) Must perform the functions specified in § 493.1274(d) and (e);
(5) Must ensure that each individual examining gynecologic preparations participates in an HHS approved cytology proficiency testing program, as specified in § 493.945 and achieves a passing score, as specified in § 493.855; and
(6) If responsible for screening cytology slide preparations, must document the number of cytology slides screened in 24 hours and the number of hours devoted during each 24-hour period to screening cytology slides.
[57 FR 7172, Feb. 28, 1992, as amended at 58 FR 5235, Jan. 19, 1993; 68 FR 3714, Jan. 24, 2003; 88 FR 90043, Dec. 28, 2023]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
The six required elements
- Direct observation of routine patient testing, including patient preparation, specimen handling, processing and testing.
- Monitoring how test results are recorded and reported, including critical results.
- Review of intermediate results or worksheets, QC records, proficiency testing results and preventive maintenance records.
- Direct observation of instrument maintenance and function checks.
- Assessment of test performance using previously tested specimens, internal blind samples or external proficiency testing samples.
- Assessment of problem-solving skills.
All six elements apply to each test system the person uses. A single generic checklist for the whole lab is a frequent reason for citation.
Frequency and timing
- Before reporting patient results on a new test or instrument: training documented.
- First year of patient testing: at least semiannually.
- After the first year: at least annually.
- When problems are found: reassess and retrain.
Who can assess
In moderate complexity testing, the technical consultant evaluates competency. In high complexity testing, the technical supervisor does, and some assessments may be delegated to a qualified general supervisor. An office manager or nurse supervisor who does not meet CLIA qualifications cannot sign as the assessor. Supervisors and consultants also need their own competency documented against their regulatory duties.
Typical citation wording
- Competency not documented for one or more testing personnel.
- Not all six elements assessed for each test system.
- Semiannual assessment missing for a new employee.
- Assessor does not meet CLIA qualifications for the role.
Example plan of correction
A model response to a typical citation in this area. Adapt it to your lab's facts and records.
Frequently asked questions
How often is competency assessment required under CLIA?
Twice during the first year an individual tests patient specimens, then at least once a year. Retraining and reassessment are needed when problems are identified.
Does the laboratory director need a competency assessment?
Not as testing personnel. The director's regulatory duties are reviewed during survey. If the director also serves in another role, such as technical consultant, competency for that role should be documented.
Is competency required for waived tests?
CLIA's six-element competency requirement applies to nonwaived testing. Training and good practice are still expected for waived testing, and some states and accreditors have their own rules.
Can proficiency testing count toward competency?
Yes. Testing external proficiency testing samples can satisfy the element that assesses test performance, as long as the person actually tested the samples and it is documented.