Clinical Lab SOP Template: What CLIA Requires in Your Procedure Manual

A CLIA procedure manual must contain a written, director-approved procedure for every test, covering specimen requirements, step-by-step method, calibration, QC, reportable range, reference intervals, critical values, result reporting and corrective action (42 CFR 493.1251). Package inserts can be used only for the parts that match your practice. Keep discontinued procedures two years.

Last reviewed Oct 2026 · By Sound Lab Consultants CLIA team · Sources: 42 CFR 493.1251, 493.1105(a)(2). General information, not legal advice; confirm against the current eCFR and CMS guidance.

Key facts and official sources

A written procedure manual is required for every nonwaived test, and surveyors read it closely. This page turns 42 CFR 493.1251 into a practical SOP outline you can use to build or audit your procedure manual.

Need help with your procedure manual? Sound Lab Consultants helps labs with lab director and technical consultant services, inspection readiness and deficiency response.

Book a free consultation

What does CLIA require in a procedure manual (§493.1251(a))?

A written procedure manual for all tests the lab performs must be available to, and followed by, laboratory personnel. Textbooks may supplement the written procedures but cannot replace them.

What must each SOP contain (§493.1251(b))?

Use these headings for each procedure. Include each item when it applies to the test:

  1. Specimen requirements: patient preparation, collection, labeling, storage, preservation, transportation, processing and referral, plus criteria for specimen acceptability and rejection

  2. Microscopic examination, including detection of inadequately prepared slides

  3. Step-by-step performance of the procedure, including test calculations and interpretation of results

  4. Preparation of slides, solutions, calibrators, controls, reagents, stains and other materials

  5. Calibration and calibration verification procedures

  6. Reportable range for test results

  7. Control procedures

  8. Corrective action when calibration or control results fail to meet acceptability criteria

  9. Limitations in methodologies, including interfering substances

  10. Reference intervals (normal values)

  11. Imminently life-threatening test results (panic or alert values)

  12. Pertinent literature references

  13. Reporting system, including protocols for reporting imminently life-threatening results

  14. Course of action when a test system becomes inoperable, including how patient specimens are handled

Can you use package inserts as SOPs (§493.1251(c))?

Manufacturer inserts or operator manuals may be used for any of items 1 through 12 when they apply. The lab must still write everything the manufacturer doesn't provide, typically your specimen acceptance and rejection rules, critical value list, reporting protocol, and back-up plan when the instrument is down. Reference the insert by version so staff use the current one.

Who must approve SOPs and changes (§493.1251(d))?

Procedures and changes in procedures must be approved, signed and dated by the current laboratory director before use. Because the rule names the current director, a lab that changes directors should have the new director review, sign and date the procedures in use.

How long must SOPs be kept (§493.1105(a)(2))?

Keep a copy of each procedure with the dates of initial use and discontinuance. Retain discontinued procedures for at least two years after they are no longer in use.

What are the basics of SOP document control?

CLIA doesn't prescribe a document control system, but a simple one prevents most findings:

  • A master list of procedures with version numbers and effective dates

  • A review schedule set by lab policy (accreditors such as CAP set their own review frequency)

  • A change log showing what changed and who approved it

  • Removal of superseded versions from the bench

  • Staff read-and-acknowledge records for new or changed SOPs

What SOP problems do surveyors cite most?

  • Package insert used as the full SOP, with no reporting or critical value section

  • Procedure not signed by the current director

  • Bench copy that doesn't match the approved version

  • Reportable range or reference intervals that don't match the LIS or patient report

  • No written plan for when the analyzer is down

Frequently asked questions

Can we just use the package insert as our SOP?

Partly. Inserts can cover many required items (§493.1251(c)), but the lab must add what's missing, such as specimen rejection, critical value reporting and downtime procedures. The director must approve the procedure.

Does a new lab director have to re-sign every SOP?

Procedures must be approved by the current director (§493.1251(d)). Most labs have a new director review and sign the manual promptly.

How often must SOPs be reviewed?

CLIA ties approval to the current director and to procedure changes. Accreditors may require periodic reviews, so follow your accreditor's standard.

How long do we keep old procedures?

At least two years after a procedure is discontinued, with dates of initial use and discontinuance (§493.1105(a)(2)).