Short answer: This template is a test procedure shell that walks through every element 42 CFR 493.1251(b) requires in a CLIA procedure manual entry, from specimen requirements to inoperable-system steps. This free template downloads as a fillable PDF and an editable Word file.
Last reviewed November 26, 2026. Governing requirement: 42 CFR 493.1251.
These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use.
What this Procedure Manual (42 CFR 493.1251 Format) SOP must contain
Purpose: To provide a test procedure in the format required for the CLIA procedure manual.
Scope: Test: [Test name] on [Instrument / test system] in [Department].
The template includes these sections:
- Header block: laboratory name, CLIA number, document number, version, effective date, next review, author, approval by the laboratory director with signature and date
- Purpose and scope
- Responsibility by role
- Definitions
- Materials, equipment and records
- Step-by-step procedure (15 steps)
- Quality control
- Interpretation, reporting and documentation
- Limitations and cautions
- References to the governing regulation and guidance
- Lab-specific notes
- Revision history and annual review log (fillable)
Why labs need it
This procedure supports 42 CFR 493.1251. Under 42 CFR 493.1251, a nonwaived laboratory must have written procedures that staff follow, and the laboratory director must approve, sign and date each procedure and every change before use. Waived labs must follow the manufacturer's instructions exactly, and written procedures help show they do. Surveyors and accreditation inspectors usually ask to see the procedure, the related records and evidence that staff were trained on it.
Procedure outline
The highlighted items are fill-in fields in the PDF and Word versions.
- Principle: [Test principle in one paragraph].
- Patient preparation: [Patient preparation].
- Specimen: type [Specimen type], container [Container], minimum volume [Minimum volume], stability [Stability], rejection criteria [Rejection criteria].
- Reagent preparation and storage: [Reagent preparation steps].
- Calibration: frequency [Calibration frequency per manufacturer] and acceptance [Calibration acceptance criteria].
- Calibration verification: at least every 6 months and after major maintenance or reagent changes per 493.1255.
- Quality control: [QC levels and frequency]; acceptable ranges per lab-established limits.
- Step-by-step procedure: [Test steps].
- Calculations: [Calculations].
- Reportable range: [Reportable range]; results outside are diluted [Dilution protocol] or reported as < or >.
- Reference intervals: [Reference intervals].
- Critical values: [Critical values]; report per the critical value SOP.
- Interferences and limitations: [Interfering substances].
- Result entry and reporting: [Reporting steps].
- If the test system is inoperable: [Backup method or referral laboratory] and notify clients of delays.
Quality control and documentation
- Corrective action when QC or calibration fails: stop patient testing, troubleshoot, repeat, and evaluate patient results since the last acceptable QC.
- Report units: [Units].
Limitations
- The manufacturer's package insert may be referenced for items it covers; add the rest.
- This is a generic template. Adapt it to your instruments, test systems, manufacturer instructions, state law and accreditation standards before use.
Key definitions
- Reportable range: The span of results the lab verified it can report without dilution or concentration.
Fields you fill in
Besides the header block, revision history and annual review log, this template has 28 lab-specific fields:
- Test name
- Instrument / test system
- Department
- Instrument and model
- Reagent names and catalog numbers
- Calibrator material
- Control material and levels
- Test principle in one paragraph
- Patient preparation
- Specimen type
- Container
- Minimum volume
- Stability
- Rejection criteria
- Reagent preparation steps
- Calibration frequency per manufacturer
- Calibration acceptance criteria
- QC levels and frequency
- Test steps
- Calculations
- Reportable range
- Dilution protocol
- Reference intervals
- Critical values
- Interfering substances
- Reporting steps
- Backup method or referral laboratory
- Units
How to adopt this template
- Download the PDF or Word version above.
- Fill in the header block and every highlighted field. Where a field says "per manufacturer", copy the value from your current package insert or operator manual.
- Delete steps that do not apply and add any steps your test system or state requires.
- Have the technical consultant or technical supervisor review it, then have the laboratory director sign and date it before use.
- Train staff, document their review, and add the SOP to your document control master list.
Need procedures written for your lab?
Sound Lab Consultants writes and reviews complete procedure manuals that match your instruments, test menu and certificate type. Book a free consultation to talk through what your lab needs.
Book a free consultationFrequently asked questions
Is this Procedure Manual (42 CFR 493.1251 Format) SOP template free?
Yes. The Procedure Manual (42 CFR 493.1251 Format) SOP template is a free download as a fillable PDF and an editable Word (.docx) file. There is no email gate or sign-up.
What regulation does a procedure manual (42 cfr 493.1251 format) procedure address?
This template is built around 42 CFR 493.1251. Read the current text at the official source linked on this page, and check your state and accreditation organization for added requirements.
Can our lab use this SOP as-is?
No. Use it as a starting point. Fill in the highlighted fields, adapt the steps to your instruments and the manufacturer's instructions, and have the laboratory director approve, sign and date it before use (42 CFR 493.1251(d)).
Which labs is this template for?
It is written for moderate and high complexity laboratories, as marked in the template's 'Applies to' boxes. Uncheck any that do not apply to your lab.
How do I fill in the PDF?
Open the PDF in a PDF reader that supports forms, click the shaded fields and type. Save a copy. In the Word version, click each highlighted field (a content control) and type over it.
Official sources
- eCFR, 42 CFR 493.1251
- eCFR, 42 CFR Part 493 Laboratory Requirements (CLIA)
- CMS, State Operations Manual Appendix C: Survey procedures and interpretive guidelines for laboratories
- Manufacturer's package insert, operator manual and safety data sheets for each test system in use
- eCFR, 42 CFR 493.1253
- eCFR, 42 CFR 493.1255
- eCFR, 42 CFR 493.1256
Related SOP templates
These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use. Sound Lab Consultants is not affiliated with or endorsed by CMS, CDC, FDA, OSHA, HHS, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer. CLSI and other standards are referenced by name only. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.