Clinical laboratory consulting in all 50 states, remote and on-site Call (602) 874-1311
Menu
Sound Lab Consultantsby Sound Lab Consultants
Sound Lab Consultants › National Lab Test Directory › Labcorp › Fibrinogen Activity

Fibrinogen Activity (Women's Health): Labcorp test 001610

Labcorp test 001610, CPT 85384. Specimen: Whole blood or plasma. Turnaround: Within 1 day.

Fibrinogen Activity is Labcorp test code 001610, billed with CPT 85384. Specimen: Whole blood or plasma. Turnaround time: Within 1 day. Performing laboratory: Labcorp, CLIA 34D0655059 (Burlington, NC). Other labs in the directory with tests billed under the same CPT: Cleveland Clinic Laboratories, ARUP Laboratories.
LaboratoryLabcorp
CLIA number34D0655059
Test code001610 (Women's Health)
CPT85384
LOINC3255-7
SpecimenWhole blood or plasma
Volume4.5 mL, 2.7 mL, 1.8 mL
Minimum volume90% of full draw ( Note: This volume does not allow for repeat testing.)
ContainerBlue-top (sodium citrate) tube; do not open tube.
Turnaround timeWithin 1 day
Storage / transportSpecimens are stable at room temperature for 24 hours. If testing cannot be completed within 24 hours, specimens should be centrifuged for at least 10 minutes at 1500xg. Plasma should then be transferred to a Labcorp PP transpak frozen purple tube with screw cap (Labcorp No. 49482). Freeze immediately and maintain frozen until tested. Refer to Coagulation Collection Procedures for directions.

Also known as: Clottable Fibrinogen; Factor I Activity

Use

This test is used for diagnosis of homozygous and heterozygous fibrinogen deficiency as well as dysfibrinogenemia, and diagnosis of disseminated intravascular coagulation. Fibrinogen levels can be used to assess the effectiveness of thrombolytic therapy.

Methodology

The measurement of fibrinogen activity is based on determination of fibrin polymerization function by the Clauss method. This method measures the rate of clot formation after adding a high concentration of thrombin to citrated plasma. The fibrinogen activity is then derived from a standard curve relating the clotting time to plasma standards of known fibrinogen activity.

Collection instructions

Blood should be collected in a blue-top tube containing 3.2% buffered sodium citrate. Evacuated collection tubes must be filled to completion to ensure a proper blood to anticoagulant ratio. The sample should be mixed immediately by gentle inversion at least six times to ensure adequate mixing of the anticoagulant with the blood. A discard tube is not required prior to collection of coagulation samples. When noncitrate tubes are collected for other tests, collect sterile and nonadditive (red-top) tubes prior to citrate (blue-top) tubes. Any tube containing an alternate anticoagulant should be collected after the blue-top tube. Gel-barrier tubes and serum tubes with clot initiators should also be collected after the citrate tubes. If testing cannot be completed within 24 hours, specimens should be centrifuged for at least 10 minutes at 1500xg. Plasma should then be transferred to a Labcorp PP transpak frozen purple tube with screw cap (Labcorp No. 49482). Freeze immediately and maintained frozen until tested. Refer to Coagulation Collection Procedures for directions. Please print and use the Volume Guide for Coagulation Testing to ensure proper draw volume.

Stability

TemperaturePeriod
Room temperature1 day
FrozenPlasma: 14 days

Causes for rejection

Clotted specimen; gross lipemia or hemolysis; tubes <90% full; improper labeling; specimen collected in tube other than 3.2% citrate

Reference range

Age Reference interval (mg/dL) 1 d 3 d 1 to 11 m 1 to 5 y 6 to 10 y 11 to 16 y >16 y

Special instructions

If the patient's hematocrit exceeds 55%, the volume of citrate in the collection tube must be adjusted. Refer to Coagulation Collection Procedures for directions.

Limitations

Fibrinogen is an acute-phase reactant and can often become significantly increased in conditions involving tissue damage, infection, or inflammation. Increased levels may be seen in smokers, during pregnancy, and in women taking oral contraceptives. Fibrinogen levels can be diminished in advanced liver disease. Very high levels of heparin, presence of direct thrombin inhibitors, or fibrin breakdown products may falsely reduce fibrinogen levels because they interfere with the rate of clot formation. Lipemia or hemolysis may interfere with this assay.

LOINC map

Result codeResult nameUnitsResult LOINC
001610Fibrinogen Activitymg/dL3255-7

View this test on Labcorp's website → · Last checked 2026-11-29

Other tests billed with CPT 85384

Search the National Lab Test Directory · All Labcorp tests and CLIA locations

Run a CLIA-certified lab? Add your test menu free. Upload your tests with our CSV template and reach clinics, physicians and labs searching by test name and CPT code. Submit your lab's test menu →
Test information is summarized from each laboratory's publicly published test menu, or uploaded by the laboratory, and linked to the original source. Requirements change: always confirm specimen and ordering details with the performing laboratory before collection. Sound Lab Consultants is not affiliated with the laboratories listed. CPT is a registered trademark of the American Medical Association. LOINC is a registered trademark of Regenstrief Institute, Inc.
Setting up or running a lab? Get it right the first time.

Sound Lab Consultants handles CLIA applications, state licensing, director coverage, validations and survey prep, remotely or on site. Prefer software? MyLabCompliance.io tracks QC, competency and personnel files.

Book a free consultation