Labcorp test 123590, CPT 81599. Specimen: Plasma (preferred) or serum. Turnaround: 2 - 3 days.

Metabolic Vulnerability Index (MVX) by NMR is Labcorp test code 123590, billed with CPT 81599. Specimen: Plasma (preferred) or serum. Turnaround time: 2 - 3 days. Performing laboratory: Labcorp, CLIA 34D0655059 (Burlington, NC). Other labs in the directory with tests billed under the same CPT: Cleveland Clinic Laboratories, ARUP Laboratories.
LaboratoryLabcorp
CLIA number34D0655059
Test code123590
CPT81599
LOINC—
SpecimenPlasma (preferred) or serum
Volume1 mL
Minimum volume0.5 mL ( Note : This volume does not allow for repeat testing.)
ContainerLavender-top (EDTA-no gel) tube (preferred) or green-top (sodium heparin) plasma tubes; plain red-top tube and NMR LipoTube (black-and-yellow-top tube) also are acceptable
Turnaround time2 - 3 days
Storage / transportRefrigerate.

Test includes

Inflammation Vulnerability Index (IVX), Metabolic Malnutrition Index (MMX), Metabolic Vulnerability Index (MVX)

Use

The Metabolic Vulnerability Index (MVX) is intended for use in adult subjects for the quantitative determination of a mortality risk score in serum or plasma. The MVX score (1-100) may be used as a prognostic aid for assessing comparative survival vulnerability stemming from metabolic dysfunction.

Methodology

The MVX test is a nuclear magnetic resonance spectroscopy (NMR)-derived multi-analyte algorithm that calculates a score based on the results of GlycA, citrate, branched-chain amino acids and small HDL particles from 400 MHz proton NMR spectra of serum or plasma samples. These analytes are used to produce the IVX, MMX and MVX scores.

Collection instructions

Collect specimen in a lavender-top (EDTA-no gel) tube, which is the preferred specimen. Separate plasma from lavender-top (EDTA-no gel) tube immediately after collection. Plasma should then be transferred to a transport tube for storage at (2°C to 8°C) until time of shipment. Separate plasma from green-top (sodium heparin) tube immediately after collection. Plasma should then be transferred to a transport tube for storage at (2°C to 8°C) until time of shipment. For plain red-top tube, hold tube upright at room temperature for 45 minutes after drawing and allow to clot. Centrifuge specimen after clotting according to manufacturer's specifications. Transfer to a transport tube for storage at (2°C to 8°C) until shipped. For NMR LipoTube (black-and-yellow-top tube), keep upright at room temperature for 30 minutes and allow to clot. Centrifuge at 1600 to 1800 xg for 10 to 15 minutes immediately after clotting. If the sample cannot be centrifuged immediately, it must be refrigerated at (2°C to 8°C and centrifuged within 24 hours of collection. The NMR tube should then be stored at (2°C to 8°C) until shipped. All specimens should be shipped for testing within 48 hours of collection. Shipment should maintain (2°C to 8°C) storage conditions to maximize sample longevity. Frozen storage and/or fresh shipment of specimens are acceptable.

Stability

TemperaturePeriod
Room temperature24 hours (stability provided by manufacturer or literature reference)
Refrigerated14 days (stability provided by manufacturer or literature reference)
Frozen14 days (stability provided by manufacturer or literature reference)
Frozen (-80°C) :EDTA plasma: 9 years
NMR LipoTube: 6 yearsRed-top serum: 3 years
Heparin plasma: No long-term dataFreeze/thaw cycles

Patient preparation

Fasting for 10 to 12 hours is preferred since there is a postprandial effect on MVX parameters that is not consistent or uniform and that depends on food and individual response.

Causes for rejection

Unspun LipoTube or unseparated plain red-top, EDTA, or sodium heparin tube; serum or plasma specimens drawn in gel-barrier collection tubes other than the NMR LipoTubes

Limitations

MMX measurements from plasma specimens are on average 10% to 15% higher than from serum specimens. IVX measurements from plasma specimens are on average 5% to 10% lower than from serum specimens. Therefore, MVX measurements from plasma specimens are on average 5% higher than from serum specimens. MVX results are increased 10% to 15% at hemoglobin concentrations higher than 123 mg/dL. IVX results are increased 10% to 20% at hemoglobin concentrations higher than 123 mg/dL. This test was developed and its performance characteristics determined by Labcorp. It has not been cleared or approved by the Food and Drug Administration.

View this test on Labcorp's website → · Last checked 2026-11-29

Other tests billed with CPT 81599

Search the National Lab Test Directory · All Labcorp tests and CLIA locations

Run a CLIA-certified lab? Add your test menu free. Upload your tests with our CSV template and reach clinics, physicians and labs searching by test name and CPT code. Submit your lab's test menu →
Test information is summarized from each laboratory's publicly published test menu, or uploaded by the laboratory, and linked to the original source. Requirements change: always confirm specimen and ordering details with the performing laboratory before collection. Sound Lab Consultants is not affiliated with the laboratories listed. CPT is a registered trademark of the American Medical Association. LOINC is a registered trademark of Regenstrief Institute, Inc.
Setting up or running a lab? Get it right the first time.

Sound Lab Consultants handles CLIA applications, state licensing, director coverage, validations and survey prep, remotely or on site. Prefer software? MyLabCompliance.io tracks QC, competency and personnel files.

Book a free consultation