Zika Virus Antibody, IgM: Labcorp test 163084
Labcorp test 163084, CPT 86794. Specimen: Serum. Turnaround: 2 - 5 days.
Use
Presumptive qualitative detection of Zika virus IgM antibodies in human sera collected from individuals meeting CDC Zika virus clinical criteria (eg, a history of clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (eg, history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiological criteria for which Zika virus testing may be indicated).
Methodology
Antibody Capture Chemiluminescence Immunoassay (CLIA)
Collection instructions
Blood collection aseptically by venipuncture. Samples should be allowed to clot and should be centrifuged and separated from the clot as soon as possible.
Stability
| Temperature | Period |
|---|---|
| Room temperature | 24 hours (stability provided by manufacturer or literature reference) |
| Refrigerated | 7 days (stability provided by manufacturer or literature reference) |
| Frozen | >7 days (stability provided by manufacturer or literature reference) |
| Freeze/thaw cycles | Stable x3 (stability provided by manufacturer or literature reference) |
Causes for rejection
Grossly hemolyzed or lipemic samples, as well as samples containing particulate or exhibiting obvious microbial contamination
Special instructions
Please refer to CDC guidance prior to submitting specimens for testing. This assay is intended for individuals meeting CDC Zika virus clinical criteria and/or CDC Zika virus epidemiological criteria. Specimens from symptomatic patients or returning travelers from endemic areas must not be collected earlier than day 8 after the onset of symptoms or risk of exposure. Zika virus serologic testing is not recommended for asymptomatic pregnant women.
Limitations
Interpretation of Zika IgM results must account for the possibility of false-negative and false-positive results. False-negative results can arise from: a) Specimen collection conducted before IgM has reached detectable levels (typically around 4 days post onset of symptoms); b) Specimen collection conducted after IgM levels have decreased below detectable levels (around 12 weeks post onset of symptoms); c) Improper collection, storage, or transport of specimens. False-positive results can arise from: a) Specific cross-reactivity with IgM for other flaviviruses (eg, Dengue virus); b) Non-specific cross-reactivity with other immunoglobulins; c) Cross-reactivity with Dengue IgM, Parvovirus…
LOINC map
| Result code | Result name | Units | Result LOINC |
|---|---|---|---|
| 163085 | Zika Virus Antibody, IgM | 80824-6 |
View this test on Labcorp's website → · Last checked 2026-11-29
Other tests billed with CPT 86794
- Zika Virus IgM Antibody Capture (MAC), by ELISA
- Zika Virus IgM Antibody Capture (MAC), by ELISA
- Zika Virus MAC-ELISA, IgM, Serum (Mayo)