Clinical laboratory consulting in all 50 states, remote and on-site Call (602) 874-1311
Menu
Sound Lab Consultantsby Sound Lab Consultants
Sound Lab Consultants › National Lab Test Directory › Labcorp › Neurofilament Light Chain, Serum

Neurofilament Light Chain, Serum: Labcorp test 140455

Labcorp test 140455, CPT 83884. Specimen: Serum. Turnaround: 1 - 3 days.

Neurofilament Light Chain, Serum is Labcorp test code 140455, billed with CPT 83884. Specimen: Serum. Turnaround time: 1 - 3 days. Performing laboratory: Labcorp, CLIA 34D0655059 (Burlington, NC). Other labs in the directory with tests billed under the same CPT: Cleveland Clinic Laboratories, Michigan Medicine Laboratories (MLabs), ARUP Laboratories.
LaboratoryLabcorp
CLIA number34D0655059
Test code140455
CPT83884
LOINC101281-4
SpecimenSerum
Volume1 mL
Minimum volume0.7 mL ( Note: This volume does not allow for repeat testing.)
ContainerRed-top tube, gel-barrier tube
Turnaround time1 - 3 days
Storage / transportRoom temperature

Use

This test is used for the measurement of the level of Neurofilament Light Chain in serum.

Methodology

Roche Diagnostics Electrochemiluminescence Immunoassay (ECLIA)

Collection instructions

If a red-top tube is used, transfer separated serum to a plastic transport tube.

Stability

TemperaturePeriod
Room temperature14 days
Refrigerated14 days
Frozen14 days
Freeze/thaw cyclesStable x3

Reference range

See table. Age (y) (pg/mL) 0 to 4 y 0.00–1.56 5 to 9 y 0.00–1.30 10 to 14 y 0.00–1.13 15 to 19 y 0.00–1.26 20 to 29 y 0.00–1.30 30 to 39 y 0.00–1.49 40 to 49 y 0.00–1.69 50 to 59 y 0.00–2.99 60 to 69 y 0.00–3.65 70 to 79 y 0.00–6.04 >79 y 0.00–9.13

Special instructions

This test may exhibit interference when sample is collected from a person who is consuming a supplement with a high dose of biotin (also termed as vitamin B7 or B8, vitamin H, or coenzyme R). It is recommended to ask all patients who may be indicated for this test about biotin supplementation. Patients should be cautioned to stop biotin consumption at least 72 hours prior to the collection of a sample.

Limitations

This test was developed and its performance characteristics determined by Labcorp. It has not been cleared or approved by the Food and Drug Administration. There are significant variations in measured serum NfL levels among different methods and labs. Care must be taken when interpreting results obtained in different studies. Direct comparisons of absolute values can only be done on the same source fluid (plasma or serum). A rise in NfL is not specific for a specific disease factor and may be caused by both neurodegenerative diseases or a head impact during sports. Results should only be used in conjunction with other clinical information when evaluating patients with neurodegeneration. Due…

LOINC map

Result codeResult nameUnitsResult LOINC
140456Neurofilament Light Chainpg/mL101281-4

View this test on Labcorp's website → · Last checked 2026-11-29

Other tests billed with CPT 83884

Search the National Lab Test Directory · All Labcorp tests and CLIA locations

Run a CLIA-certified lab? Add your test menu free. Upload your tests with our CSV template and reach clinics, physicians and labs searching by test name and CPT code. Submit your lab's test menu →
Test information is summarized from each laboratory's publicly published test menu, or uploaded by the laboratory, and linked to the original source. Requirements change: always confirm specimen and ordering details with the performing laboratory before collection. Sound Lab Consultants is not affiliated with the laboratories listed. CPT is a registered trademark of the American Medical Association. LOINC is a registered trademark of Regenstrief Institute, Inc.
Setting up or running a lab? Get it right the first time.

Sound Lab Consultants handles CLIA applications, state licensing, director coverage, validations and survey prep, remotely or on site. Prefer software? MyLabCompliance.io tracks QC, competency and personnel files.

Book a free consultation