Short answer: An Individualized Quality Control Plan (IQCP) is the CMS-approved alternative to default CLIA quality control for nonwaived testing. It has three required parts: a risk assessment, a quality control plan signed by the laboratory director, and ongoing quality assessment. It's voluntary, can't be less stringent than the manufacturer's instructions, and excludes pathology.
Last reviewed November 25, 2026 by Sound Lab Consultants. Primary sources: 42 CFR Part 493, CMS Clinical Laboratory Improvement Amendments (CLIA) program.

Key facts and official sources
Default QC is defined in 42 CFR 493.1256(d), which allows an equivalent procedure if CMS approves it in SOM Appendix C.
IQCP guidance, including Table 1 (Eligibility for IQCP), is in SOM Appendix C.
CMS fully implemented IQCP on January 1, 2016, replacing the earlier "equivalent quality control" options.
ASM's IQCP protocol templates for microbiology and molecular testing
IQCP guides
IQCP Risk Assessment: Five Components and How to Document Them
IQCP Quality Assessment and Review: Keeping the Plan Working
IQCP for Microbiology: Media, Identification Systems and AST
IQCP for Molecular Testing: Cartridge PCR, LDTs and Exclusions
When does IQCP make sense?
IQCP fits test systems with built-in or electronic controls, unit-use cartridges, or other designs where running external liquid controls every day adds cost without adding much error detection. For an analyzer that already runs two levels daily, default QC is usually simpler.
What goes in an IQCP risk assessment?
Identify possible errors across the pre-analytic, analytic and post-analytic phases for five components:
Specimen: collection, labeling, transport and acceptability
Test system: the instrument and its built-in controls, calibration and maintenance
Reagents: storage, shipping, lot changes and expiration
Environment: temperature, humidity, power and space
Testing personnel: training, competency and staffing
Base the assessment on your lab's own QC, PT, complaint, error and maintenance history, not only the manufacturer's claims.
What goes in the QC plan?
The QC plan sets out the controls for each risk you identified, including the type and frequency of QC. It must:
meet or exceed the manufacturer's instructions
be specific to the test, location and setting
be approved and signed by the laboratory director before use
Questions about CLIA, accreditation or laboratory operations?
Book a free consultation. Bring the certificate type and the question you want to sort out first.
Book a free consultationHow is quality assessment done?
Monitor QC and PT results, specimen rejections, complaints, corrected reports and turnaround problems. Reassess the plan after any error, and after any change such as a new reagent formulation, staff turnover or a new testing site.
What does an IQCP template include?
Test, test system, location, and the CLIA specialty and subspecialty
Eligibility check against SOM Appendix C, Table 1
A summary of the manufacturer's QC instructions
A risk assessment table: component, possible error, likelihood or impact, existing control, added control
The QC plan: control type, frequency, acceptance criteria and corrective action
The quality assessment plan: indicators, review frequency and reviewer
The director's approval signature and date, which must precede use
A review and revision history
How do IQCP and PT failures relate?
An unsuccessful PT event on a test covered by an IQCP is a quality assessment trigger. Investigate whether the risk assessment missed the cause, then document any change to the QC plan, such as more frequent external QC. Repeated PT failures can lead to CLIA sanctions regardless of the IQCP. See failed PT corrective action.
What can't IQCP be used for?
Waived tests, which follow the manufacturer's instructions only.
Pathology, histopathology, oral pathology and cytology, which are not eligible under Appendix C, Table 1.
Specialty-specific requirements that stay mandatory even in eligible specialties, such as immunohematology (§493.1271) and clinical cytogenetics (§493.1276).
Accredited labs should also check their accreditor's IQCP requirements.
What IQCP weaknesses do surveyors cite?
Generic templates with no lab data.
QC set below the manufacturer's requirements.
A missing or late director signature.
No ongoing quality assessment.
One plan covering different test systems or sites without assessing each.
Talk with Sound Lab Consultants
Book a free consultation. Tell us the laboratory type, the certificate or accreditor and what you want help with first.
Book a free consultationFrequently asked questions
Is IQCP required?
No. It's optional. Labs can follow the default QC requirements in §493.1256(d)(3) instead.
Can IQCP be used for waived tests?
No. It applies only to nonwaived testing.
Who approves an IQCP?
The laboratory director, who signs it before use.
Can an IQCP require less QC than the manufacturer?
No.
When should an IQCP be reassessed?
After an error or PT failure, and whenever reagents, staff, location or the test system change.
What does IQCP stand for?
Individualized Quality Control Plan, the CLIA alternative QC option described in CMS State Operations Manual Appendix C for 42 CFR 493.1250.
What are the steps of an IQCP?
Gather information, do a risk assessment, write the QC plan, get the laboratory director's signed approval, then monitor through quality assessment. See the IQCP template.
Can IQCP move QC to every 30 days?
It can if your in-house data spans the interval and supports it, and the manufacturer allows it. See IQCP 30-day QC.
Is there an IQCP certification?
CMS doesn't certify individual IQCPs or people for IQCP. CDC, CMS and accreditors publish free education, such as the CDC/CMS workbook.
Do CAP and COLA labs follow CLIA IQCP rules?
Accredited labs follow their accreditor's IQCP requirements, per CMS. See IQCP requirements under CLIA, CAP and COLA.
When did IQCP start?
CMS introduced IQCP in 2013 and it fully replaced EQC on January 1, 2016. See IQCP vs EQC.
Sources
- 42 CFR Part 493, Laboratory Requirements (CLIA)
- CMS Clinical Laboratory Improvement Amendments (CLIA) program
Related pages
Last reviewed November 25, 2026. Sound Lab Consultants is an independent laboratory consulting firm. This page is general information, not legal advice; confirm current requirements with CMS, your state agency and your accreditor. We are not affiliated with or endorsed by CMS, CDC, FDA, OSHA, CAP, COLA, The Joint Commission, any state agency or any instrument manufacturer. Illustrations are original works by Sound Lab Consultants. Sound Lab Consultants, 25 S Arizona Pl, Chandler, AZ 85225.