Copeptin: Labcorp test 010505
Labcorp test 010505, CPT 84588. Specimen: Plasma (EDTA). Turnaround: 3 - 5 days.
Use
Measurement of copeptin levels in plasma
Methodology
The ThermoFisher/BRAHMS KRYPTOR® assay employs Time-Resolved Amplified Cryptate Emission Cryptate Emission (TRACE) technology based on a non-radioactive energy transfer between a donor (europium cryptate) and an acceptor (XL665) in a sandwich immunofluorescent format using two mouse monoclonal antibodies.
Collection instructions
Separate plasma from cells and transfer to a plastic transport tube.
Stability
| Temperature | Period |
|---|---|
| Room temperature | 14 days |
| Refrigerated | 14 days |
| Frozen | 14 days |
| Freeze/thaw cycles | Stable x3 |
Causes for rejection
Gross lipemia; gross hemolysis; gross icterus
Reference range
0.0–5.9 pmol/L Reference interval derived from testing of non-water deprived, non-fasting adults.
Limitations
This test was developed and its performance characteristics determined by Labcorp. It has not been cleared or approved by the Food and Drug Administration. • Sepsis, severe sepsis, septic shock, lower respiratory tract infections, chronic obstructive pulmonary disease (COPD), and cardiovascular diseases, i.e., chronic heart failure may increase copeptin concentrations. • Arginine vasopressin (AVP) receptor antagonist therapies and other diseases in which AVP has been shown to play an important pathophysiologic role may also increase copeptin concentration. • Mixed forms of diabetes insipidus (DI) can exist, and both central and peripheral DI may be incomplete, complicating the…
LOINC map
| Result code | Result name | Units | Result LOINC |
|---|---|---|---|
| 010506 | Copeptin | pmol/L | 78987-5 |
View this test on Labcorp's website → · Last checked 2026-11-29
Other tests billed with CPT 84588
- Antidiuretic Hormone (ADH) Profile
- Antidiuretic Hormone (ADH), Plasma
- Copeptin proAVP, Plasma
- Copeptin proAVP, Plasma
Search the National Lab Test Directory · All Labcorp tests and CLIA locations
Sound Lab Consultants handles CLIA applications, state licensing, director coverage, validations and survey prep, remotely or on site. Prefer software? MyLabCompliance.io tracks QC, competency and personnel files.
Book a free consultation