Clinical laboratory consulting in all 50 states, remote and on-site Call (602) 874-1311
Menu
Sound Lab Consultantsby Sound Lab Consultants
Sound Lab Consultants › National Lab Test Directory › Allina Health Laboratory (Abbott Northwestern) › Ammonia

Ammonia: Allina Health Laboratory (Abbott Northwestern) test 27

Allina Health Laboratory (Abbott Northwestern) test 27, CPT 82140. Specimen: EDTA plasma. Turnaround: 1 day.

Ammonia is Allina Health Laboratory (Abbott Northwestern) test code 27, billed with CPT 82140. Specimen: EDTA plasma. Turnaround time: 1 day. Performing laboratory: Allina Health Laboratory (Abbott Northwestern), CLIA 24D0401558 (Minneapolis, MN). Other labs in the directory with tests billed under the same CPT: OHSU Laboratory Services, Michigan Medicine Laboratories (MLabs), Cleveland Clinic Laboratories, ARUP Laboratories, Quest Diagnostics.
LaboratoryAllina Health Laboratory (Abbott Northwestern)
CLIA number24D0401558
Test code27
CPT82140
LOINC—
SpecimenEDTA plasma
Volume1.0 mL
Minimum volume0.2 mL
ContainerLavender (EDTA), 4mL
Turnaround time1 day

Also known as: NH3

Use

Ammonia, derived from the catabolism of amino acids and from the action of intestinal bacteria on dietary protein, is converted to urea in the liver hepatocytes and so rendered non-toxic. Studies have shown that excess ammonia can have a toxic effect on the central nervous system and clinical manifestations are typically neurological disturbances. Elevated levels of ammonia may be either due to "Inborn Errors of Metabolism", or secondary to other conditions. Inborn errors of metabolism are the major cause of elevated ammonia in infants and usually the result of urea cycle enzyme deficiencies. Inherited disorders affecting the metabolism of the dibasic amino acids (lysine and ornithine) and…

Methodology

Enzymatic

Collection instructions

Following collection, mix sample by inverting 8 - 10 times to prevent clotting Immediately place sample on ice

Stability

Frozen (preferred) - 3 days Refrigerated - 2 hours Ambient - 30 minutes

Causes for rejection

Hemolysis Erythrocytes contain approximately 3 times more ammonia than does plasma Specimens should not be repeatedly frozen and thawed Specimens should be free of particulate matter and red blood cells Improper labels (unlabeled or mislabeled) Improper anticoagulant or ratio Delay in transport Improper storage temperature affecting results Inappropriate timing of collection Improper container Leaking container resulting in compromised specimen Quantity not sufficient (QNS)

Reference range

Male 11 - 51 µmol/L Female 16 - 60 µmol/L Results that require action (critical): >110 µmol/L

Performed

Daily

Abbreviation

AMM

Order number

82140

Transport container

False bottom plasma/serum transport vial/tube (AHL)

Performing Laboratory

AHL - Chemistry: C

Processing instructions

Spin and transfer plasma to a False bottom plasma/serum transport vial/tube (AHL), labelled as EDTA plasma, within 30 minutes of collection . If separation is delayed, the results may be falsely elevated. If the specimen will not be tested within 1 hour of collection , the plasma must be frozen at <-20°C.

View this test on Allina Health Laboratory (Abbott Northwestern)'s website → · Last checked 2026-11-29

Other tests billed with CPT 82140

Search the National Lab Test Directory · All Allina Health Laboratory (Abbott Northwestern) tests and CLIA locations

Run a CLIA-certified lab? Add your test menu free. Upload your tests with our CSV template and reach clinics, physicians and labs searching by test name and CPT code. Submit your lab's test menu →
Test information is summarized from each laboratory's publicly published test menu, or uploaded by the laboratory, and linked to the original source. Requirements change: always confirm specimen and ordering details with the performing laboratory before collection. Sound Lab Consultants is not affiliated with the laboratories listed. CPT is a registered trademark of the American Medical Association. LOINC is a registered trademark of Regenstrief Institute, Inc.
Setting up or running a lab? Get it right the first time.

Sound Lab Consultants handles CLIA applications, state licensing, director coverage, validations and survey prep, remotely or on site. Prefer software? MyLabCompliance.io tracks QC, competency and personnel files.

Book a free consultation