NGS Oncology Panel Validation
The AMP/CAP joint consensus guidelines for NGS-based oncology panels (Jennings et al., J Mol Diagn 2017) are the standard framework for validating targeted tumor panels in a CLIA lab. They cover panel content selection, an optimization and familiarization phase before validation, use of reference cell lines and reference materials, positive percentage agreement (PPA) and positive predictive value (PPV) for each variant type, minimum depth of coverage, and an error-based approach led by the laboratory director.
- Guideline
- Guidelines for Validation of Next-Generation Sequencing-Based Oncology Panels
- Published by
- Association for Molecular Pathology and College of American Pathologists
- Citation
- Jennings LJ et al. J Mol Diagn 2017;19(3):341-365
- Scope
- Targeted DNA panels for somatic variants in solid tumors and hematologic malignancies
The validation phases
- Test design: define intended use, tumor types, specimen types (FFPE, fresh tissue, bone marrow, cytology), genes and regions, and the variant types you will report: SNVs, small indels, larger indels, copy number changes and fusions.
- Optimization and familiarization: run the assay before formal validation to learn its behavior, set provisional thresholds and fix the protocol. Data from this phase is not the validation.
- Validation: run a predefined sample set against locked procedures and pipeline, and calculate performance for each variant type separately.
- Quality management after go-live: run controls, monitor QC metrics and proficiency testing, and revalidate when anything material changes.
Performance characteristics to establish
| Characteristic | How it is usually established |
|---|---|
| Positive percentage agreement (PPA, sensitivity) | Known positives from reference materials, cell lines and previously tested patient samples, per variant type and across the allele fraction range you claim. |
| Positive predictive value (PPV) | Proportion of reported variants that are true, confirmed by an orthogonal method or a reference result. |
| Limit of detection | Dilution series of known variants into wild-type DNA to find the lowest variant allele fraction detected reliably at your minimum depth. |
| Precision (repeatability and reproducibility) | Replicates within run, between runs, operators, lots and instruments. |
| Depth of coverage | Minimum depth set so the claimed LoD is achievable; region-level failures reported or flagged. |
| Reportable range | All targeted regions that pass coverage and quality thresholds; regions that fail are disclosed. |
How many samples?
The guideline recommends a statistically justified minimum number of samples per variant type rather than a fixed number. A common way to express it uses confidence and reliability: to show with 95% confidence that performance is at least 95%, with no failures, you need 59 positive results (n = ln(1 − confidence) / ln(reliability)). Fewer samples support only weaker claims, so labs combine reference materials and cell lines with real patient specimens, especially for hard variant types such as long indels and fusions.
The error-based approach
AMP/CAP put the laboratory director at the center: identify what can go wrong at each step (specimen tumor content, extraction, library complexity, capture, sequencing, alignment, variant calling, interpretation), then control each risk through test design, validation studies or ongoing QC. Your validation report should show which risks were addressed and how.
What CLIA requires for any NGS validation
| Requirement | What CLIA requires |
|---|---|
| Validation before patient testing | Most clinical NGS tests are laboratory developed or modified, so the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals (42 CFR 493.1253(b)(2)). |
| Personnel | High complexity testing: a qualified laboratory director (42 CFR 493.1443) and technical supervisor (42 CFR 493.1449) who approve the validation and procedures. |
| Facilities | Amplification steps outside closed systems need a unidirectional workflow (42 CFR 493.1101(a)(3)). |
| Quality control | Control procedures for each run, including controls that detect extraction errors and, for amplification, two control materials (42 CFR 493.1256(d)(3)(iv) and 42 CFR 493.1256(d)(3)(v)). |
| Proficiency testing | Most NGS analytes have no CLIA-regulated PT, so verify accuracy at least twice a year through external surveys, split samples or other means (42 CFR 493.1236(c)). |
| Records | Keep validation data, pipeline versions and QC records for at least two years after the procedure is discontinued (42 CFR 493.1105). |
Professional guidelines describe how to meet these requirements; CAP inspectors and New York State reviewers expect your validation to follow them or document why it differs. For New York patients, see New York genetic testing requirements.
Validating an NGS assay?
Our laboratory consultants write NGS validation plans, select reference materials, review pipeline validation, build the validation report your director signs, and prepare labs for CAP and New York State review, in every state.
Book a free consultationAll NGS validation guidelinesOr call (602) 874-1311 · info@soundlabconsult.com
Frequently asked questions
Are the AMP/CAP NGS oncology guidelines required by CLIA?
No. CLIA requires the lab to establish performance specifications under 42 CFR 493.1253(b)(2). The AMP/CAP guidelines are professional best practice, and CAP inspectors and New York State reviewers expect validations to follow them or justify differences.
Do I need to validate each variant type separately?
Yes. SNVs, indels, copy number changes and fusions behave differently, so PPA, PPV and LoD should be established for each type you report.
Can I use cell lines instead of patient samples?
Reference materials and cell lines are recommended for known variants and allele fractions, but the guideline also expects real clinical specimens of the types you will test, such as FFPE tissue.
Related NGS validation guides
Also see NGS lab consulting, clinical NGS workflow compliance, NGS proficiency testing and the CAP checklist and how NGS tests work.
Sources