ctDNA Liquid Biopsy Assay Validation

Reviewed November 28, 2026 by the Sound Lab Consultants laboratory team · Molecular genetics and genomics

Short answer

NGS-based circulating tumor DNA (ctDNA) assays are validated with the same core studies as other NGS tests, but at much lower variant allele fractions and with more reliance on contrived samples. BLOODPAC's generic protocols for the analytical validation of NGS-based ctDNA assays (Clinical Chemistry, 2020) give a standard starting point for limit of blank, limit of detection, accuracy, precision, input and interference studies, and BLOODPAC has since published protocols for tumor-informed MRD assays.

Main reference
Generic Protocols for the Analytical Validation of NGS-Based ctDNA Assays (BLOODPAC Analytical Variables Working Group)
Citation
Godsey JH et al. Clin Chem 2020;66(9):1156-1166
Also
BLOODPAC MRD analytical validation protocols (JCO Precision Oncology); CLSI MM09 liquid biopsy content
Specimen
Plasma cell-free DNA from blood in cfDNA-stabilizing or EDTA tubes

Why ctDNA validation is harder

Core studies

StudyWhat to show
Limit of blank (LoB)How often variants are called in true negative (wild-type) samples; sets the false-positive floor.
Limit of detection (LoD)The lowest allele fraction (at stated input) detected at the claimed rate, per variant type.
AccuracyAgreement with an orthogonal method (such as digital PCR) or tissue results.
PrecisionRepeatability and reproducibility near the LoD.
DNA input and preanalytical variablesEffect of blood tube type, time to processing, plasma volume and cfDNA yield.
Interference and cross-contaminationHemolysis, lipemia, genomic DNA contamination and sample-to-sample carryover.

Preanalytical controls labs often miss

Specify collection tube types and maximum time to plasma separation, the double-spin centrifugation protocol, plasma storage temperature, and acceptance criteria for cfDNA yield and fragment size. Validate each tube type you accept. These are where most ctDNA validations get questioned.

What CLIA requires for any NGS validation

RequirementWhat CLIA requires
Validation before patient testingMost clinical NGS tests are laboratory developed or modified, so the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals (42 CFR 493.1253(b)(2)).
PersonnelHigh complexity testing: a qualified laboratory director (42 CFR 493.1443) and technical supervisor (42 CFR 493.1449) who approve the validation and procedures.
FacilitiesAmplification steps outside closed systems need a unidirectional workflow (42 CFR 493.1101(a)(3)).
Quality controlControl procedures for each run, including controls that detect extraction errors and, for amplification, two control materials (42 CFR 493.1256(d)(3)(iv) and 42 CFR 493.1256(d)(3)(v)).
Proficiency testingMost NGS analytes have no CLIA-regulated PT, so verify accuracy at least twice a year through external surveys, split samples or other means (42 CFR 493.1236(c)).
RecordsKeep validation data, pipeline versions and QC records for at least two years after the procedure is discontinued (42 CFR 493.1105).

Professional guidelines describe how to meet these requirements; CAP inspectors and New York State reviewers expect your validation to follow them or document why it differs. For New York patients, see New York genetic testing requirements.

Validating an NGS assay?

Our laboratory consultants write NGS validation plans, select reference materials, review pipeline validation, build the validation report your director signs, and prepare labs for CAP and New York State review, in every state.

Book a free consultationAll NGS validation guidelines

Or call (602) 874-1311 · info@soundlabconsult.com

Frequently asked questions

What is the difference between LoB and LoD?

The limit of blank describes the highest apparent signal expected from samples with no variant; the limit of detection is the lowest true variant level detected reliably above the limit of blank.

Can contrived samples be used to validate a liquid biopsy assay?

Yes, and they are usually necessary for low allele fractions, but the lab should show they behave like patient cfDNA (fragment size, input) and include real patient samples where possible.

Related NGS validation guides

NGS oncology panel validation (AMP/CAP)The AMP/CAP joint consensus guidelines for NGS-based oncology panels (Jennings et al., J Mol Diagn 2017) are the standard framework for validating tar.NGS validation metricsA clinical NGS validation reports accuracy as positive percent agreement (PPA) and positive predictive value (PPV) for each variant type, precision as.NGS bioinformatics pipeline validationThe AMP/CAP Standards and Guidelines for Validating NGS Bioinformatics Pipelines (Roy et al., J Mol Diagn 2018) give 17 recommendations for designing,.

Also see NGS lab consulting, clinical NGS workflow compliance, NGS proficiency testing and the CAP checklist and how NGS tests work.