NGS Validation Metrics
A clinical NGS validation reports accuracy as positive percent agreement (PPA) and positive predictive value (PPV) for each variant type, precision as within-run and between-run reproducibility, analytical sensitivity as the limit of detection (and limit of blank), analytical specificity including interference, and the quality thresholds the test will use every run: minimum read depth, coverage uniformity, base quality and variant-calling filters.
- Accuracy
- PPA = TP / (TP + FN); PPV = TP / (TP + FP)
- Precision
- Concordance of calls across replicates, runs, operators, lots and instruments
- Sensitivity
- Limit of detection by variant allele fraction and input
- Run QC
- Depth, uniformity, base quality, on-target rate, duplication rate, contamination
Accuracy: PPA and PPV
Compare each NGS call with a reference result (reference material truth set, orthogonal method, or prior validated test). Count true positives (TP), false positives (FP) and false negatives (FN) separately for each variant type and, for somatic tests, by allele fraction range. Report point estimates with 95% confidence intervals; a 100% PPA from 10 variants has a lower confidence bound near 69%, which is why sample size matters.
| Metric | Formula | Meaning |
|---|---|---|
| Positive percent agreement (PPA) | TP / (TP + FN) | How many true variants the test finds |
| Positive predictive value (PPV) | TP / (TP + FP) | How many reported variants are real |
| Negative percent agreement (NPA) | TN / (TN + FP) | Used per position or per region; very large denominators make it less informative |
Limit of detection and limit of blank
Make dilution series of known variants into wild-type DNA (or use quantified reference materials) at several allele fractions and inputs, run replicates (20 per level is common), and define the LoD as the lowest level detected at your target rate, typically 95%. Establish the limit of blank from true negatives so the calling threshold sits above background error. For somatic tests, state LoD as variant allele fraction at a minimum depth and DNA input.
Coverage and quality thresholds
| Metric | Why it matters |
|---|---|
| Minimum read depth per position | Determines whether the claimed LoD is achievable; below it, the region is reported as not assessed. |
| Coverage uniformity (for example % of bases at ≥0.2× mean) | Low uniformity leaves regions under-covered even when mean depth looks fine. |
| Base and mapping quality | Filters low-confidence reads and calls. |
| On-target rate, duplication rate, library complexity | Detect capture or library problems early. |
| Contamination and sample identity checks | Catch sample swaps and mixtures (for example SNP fingerprinting). |
Set each threshold from validation data, document it in the procedure, and check it on every run.
Precision
Run the same samples within a run, across runs and days, and across operators, reagent lots and instruments. Report variant-level concordance and the variability of allele fractions or copy number values near decision points.
What CLIA requires for any NGS validation
| Requirement | What CLIA requires |
|---|---|
| Validation before patient testing | Most clinical NGS tests are laboratory developed or modified, so the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals (42 CFR 493.1253(b)(2)). |
| Personnel | High complexity testing: a qualified laboratory director (42 CFR 493.1443) and technical supervisor (42 CFR 493.1449) who approve the validation and procedures. |
| Facilities | Amplification steps outside closed systems need a unidirectional workflow (42 CFR 493.1101(a)(3)). |
| Quality control | Control procedures for each run, including controls that detect extraction errors and, for amplification, two control materials (42 CFR 493.1256(d)(3)(iv) and 42 CFR 493.1256(d)(3)(v)). |
| Proficiency testing | Most NGS analytes have no CLIA-regulated PT, so verify accuracy at least twice a year through external surveys, split samples or other means (42 CFR 493.1236(c)). |
| Records | Keep validation data, pipeline versions and QC records for at least two years after the procedure is discontinued (42 CFR 493.1105). |
Professional guidelines describe how to meet these requirements; CAP inspectors and New York State reviewers expect your validation to follow them or document why it differs. For New York patients, see New York genetic testing requirements.
Validating an NGS assay?
Our laboratory consultants write NGS validation plans, select reference materials, review pipeline validation, build the validation report your director signs, and prepare labs for CAP and New York State review, in every state.
Book a free consultationAll NGS validation guidelinesOr call (602) 874-1311 · info@soundlabconsult.com
Frequently asked questions
What read depth is required for clinical NGS?
No single number is required. Depth is set during validation so the claimed limit of detection is met; somatic panels detecting low allele fractions typically need several hundred unique reads per position, while germline tests need less.
Should I report sensitivity and specificity or PPA and PPV?
For NGS, PPA and PPV per variant type are preferred because true negatives across millions of positions inflate specificity.