CLSI MM09 (3rd Ed., 2023) & CAP NGS Worksheets

Reviewed November 28, 2026 by the Sound Lab Consultants laboratory team · Molecular genetics and genomics

Short answer

CLSI MM09, Human Genetic and Genomic Testing Using Traditional and High-Throughput Nucleic Acid Sequencing Methods (3rd edition, April 2023), is the main consensus standard for designing, validating, reporting and managing quality of NGS and Sanger sequencing tests across hereditary disease, tumor testing, HLA typing, NIPT and liquid biopsy. FDA recognizes it, and it incorporates the CAP NGS worksheets, which walk a lab through each stage of building a clinical NGS test.

Standard
CLSI MM09, 3rd edition
Published
April 20, 2023 (replaces MM09-A2, 2014)
Scope
Hereditary disorders, solid tumors and hematologic malignancies, HLA typing, NIPT, liquid biopsy, RNA-seq; Sanger and NGS
Status
FDA-recognized consensus standard
Companion
CAP NGS worksheets (seven worksheets developed with CLSI)

What MM09 covers

The CAP NGS worksheets

CAP developed the worksheets to turn the abstract steps of building an NGS test into practical forms, and expanded them to seven worksheets in partnership with CLSI, which incorporated them into MM09. They cover test familiarization and content design, assay design and optimization, validation, quality management and bioinformatics, with teaching examples rather than prescriptive cut-offs. They are a convenient backbone for a validation plan and the record an inspector will ask for.

How MM09 fits with CLIA and CAP

SourceRole
CLIA (42 CFR Part 493)Legal requirements: establish performance specifications, QC, PT or alternative assessment, personnel and records.
CAP Molecular Pathology checklistAccreditation requirements for NGS, including validation records, pipeline control and confirmatory testing.
CLSI MM09 and CAP worksheetsHow to do it: study design, examples and forms.
AMP/CAP, AMP/NSGC and ACMG guidelinesApplication-specific detail (somatic panels, pipelines, germline confirmation).

What CLIA requires for any NGS validation

RequirementWhat CLIA requires
Validation before patient testingMost clinical NGS tests are laboratory developed or modified, so the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals (42 CFR 493.1253(b)(2)).
PersonnelHigh complexity testing: a qualified laboratory director (42 CFR 493.1443) and technical supervisor (42 CFR 493.1449) who approve the validation and procedures.
FacilitiesAmplification steps outside closed systems need a unidirectional workflow (42 CFR 493.1101(a)(3)).
Quality controlControl procedures for each run, including controls that detect extraction errors and, for amplification, two control materials (42 CFR 493.1256(d)(3)(iv) and 42 CFR 493.1256(d)(3)(v)).
Proficiency testingMost NGS analytes have no CLIA-regulated PT, so verify accuracy at least twice a year through external surveys, split samples or other means (42 CFR 493.1236(c)).
RecordsKeep validation data, pipeline versions and QC records for at least two years after the procedure is discontinued (42 CFR 493.1105).

Professional guidelines describe how to meet these requirements; CAP inspectors and New York State reviewers expect your validation to follow them or document why it differs. For New York patients, see New York genetic testing requirements.

Validating an NGS assay?

Our laboratory consultants write NGS validation plans, select reference materials, review pipeline validation, build the validation report your director signs, and prepare labs for CAP and New York State review, in every state.

Book a free consultationAll NGS validation guidelines

Or call (602) 874-1311 · info@soundlabconsult.com

Frequently asked questions

Is CLSI MM09 free?

No. MM09 is purchased from CLSI. The CAP NGS worksheets are available from CAP's website.

Does MM09 cover microbial or metagenomic sequencing?

MM09 focuses on human genetic and genomic testing. Infectious disease and metagenomic NGS draw on other references, such as published mNGS validation frameworks.

Related NGS validation guides

NGS oncology panel validation (AMP/CAP)The AMP/CAP joint consensus guidelines for NGS-based oncology panels (Jennings et al., J Mol Diagn 2017) are the standard framework for validating tar.NGS germline variant confirmation (AMP/NSGC)The AMP and NSGC joint report on NGS germline variant confirmation (Crooks et al., J Mol Diagn 2023) recommends that every lab offering germline NGS h.NGS validation metricsA clinical NGS validation reports accuracy as positive percent agreement (PPA) and positive predictive value (PPV) for each variant type, precision as.Infectious disease and metagenomic NGS validationInfectious disease NGS tests, from whole-genome sequencing of bacterial isolates to metagenomic NGS (mNGS) that detects any pathogen in CSF or plasma,.

Also see NGS lab consulting, clinical NGS workflow compliance, NGS proficiency testing and the CAP checklist and how NGS tests work.