Sound Lab Consultants › NGS validation guidelines › NGS germline variant confirmation (AMP/NSGC)

NGS Germline Variant Confirmation

Reviewed November 28, 2026 by the Sound Lab Consultants laboratory team · Molecular genetics and genomics

Short answer

The AMP and NSGC joint report on NGS germline variant confirmation (Crooks et al., J Mol Diagn 2023) recommends that every lab offering germline NGS have a written orthogonal confirmation policy, approved by a qualified professional with NGS experience. Labs can skip confirmation only for variant calls that meet quality criteria validated with their own data, and every report should summarize the confirmation policy.

Guideline
Recommendations for Next-Generation Sequencing Germline Variant Confirmation
Published by
Association for Molecular Pathology and National Society of Genetic Counselors
Citation
Crooks KR et al. J Mol Diagn 2023;25(7):411-427
Companion
AMP white paper: Determining and Validating Germline Variant Confirmation Criteria (2023)
Out of scope
Tumor sequencing, liquid biopsy and noninvasive prenatal screening

The recommendations in plain language

Validating criteria to skip confirmation

The AMP companion white paper describes how to build the data set: collect many NGS calls that were also tested by an orthogonal method, include true positives and false positives across variant types, then find quality thresholds that separate them. A threshold is only defensible if it catches all known false positives in the training data and is then checked on an independent set. Re-check the criteria after any change to the assay or pipeline.

Related germline standards

DocumentWhat it covers
ACMG technical standards: NGS for constitutional variants (2021 revision)Test design, validation, quality metrics and reporting for germline panels, exomes and genomes.
ACMG/AMP variant classification (Richards et al., 2015)Pathogenic to benign classification framework for germline variants.
CLSI MM09 (3rd ed., 2023)Design, validation and quality management of sequencing tests, including germline applications.

What CLIA requires for any NGS validation

RequirementWhat CLIA requires
Validation before patient testingMost clinical NGS tests are laboratory developed or modified, so the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals (42 CFR 493.1253(b)(2)).
PersonnelHigh complexity testing: a qualified laboratory director (42 CFR 493.1443) and technical supervisor (42 CFR 493.1449) who approve the validation and procedures.
FacilitiesAmplification steps outside closed systems need a unidirectional workflow (42 CFR 493.1101(a)(3)).
Quality controlControl procedures for each run, including controls that detect extraction errors and, for amplification, two control materials (42 CFR 493.1256(d)(3)(iv) and 42 CFR 493.1256(d)(3)(v)).
Proficiency testingMost NGS analytes have no CLIA-regulated PT, so verify accuracy at least twice a year through external surveys, split samples or other means (42 CFR 493.1236(c)).
RecordsKeep validation data, pipeline versions and QC records for at least two years after the procedure is discontinued (42 CFR 493.1105).

Professional guidelines describe how to meet these requirements; CAP inspectors and New York State reviewers expect your validation to follow them or document why it differs. For New York patients, see New York genetic testing requirements.

Validating an NGS assay?

Our laboratory consultants write NGS validation plans, select reference materials, review pipeline validation, build the validation report your director signs, and prepare labs for CAP and New York State review, in every state.

Book a free consultationAll NGS validation guidelines

Or call (602) 874-1311 · info@soundlabconsult.com

Frequently asked questions

Does every germline NGS variant need Sanger confirmation?

No. The AMP/NSGC recommendations allow labs to skip orthogonal confirmation for calls that meet quality criteria the lab has validated with its own data, as long as the policy is written, approved and disclosed on reports.

Do these recommendations apply to tumor NGS?

No. They apply to germline DNA sequencing only; tumor sequencing, liquid biopsy and NIPT are out of scope.

Related NGS validation guides

CLSI MM09 and the CAP NGS worksheetsCLSI MM09, Human Genetic and Genomic Testing Using Traditional and High-Throughput Nucleic Acid Sequencing Methods (3rd edition, April 2023), is the m.NGS validation metricsA clinical NGS validation reports accuracy as positive percent agreement (PPA) and positive predictive value (PPV) for each variant type, precision as.NGS bioinformatics pipeline validationThe AMP/CAP Standards and Guidelines for Validating NGS Bioinformatics Pipelines (Roy et al., J Mol Diagn 2018) give 17 recommendations for designing,.

Also see NGS lab consulting, clinical NGS workflow compliance, NGS proficiency testing and the CAP checklist and how NGS tests work.

Sources

Independent summary for laboratories, reviewed November 28, 2026. Guideline names belong to their publishers (AMP, CAP, CLSI, NSGC, ACMG, BLOODPAC, FDA); Sound Lab Consultants is not affiliated with them. Read the original documents before finalizing a validation. Not legal advice.