NGS Germline Variant Confirmation
The AMP and NSGC joint report on NGS germline variant confirmation (Crooks et al., J Mol Diagn 2023) recommends that every lab offering germline NGS have a written orthogonal confirmation policy, approved by a qualified professional with NGS experience. Labs can skip confirmation only for variant calls that meet quality criteria validated with their own data, and every report should summarize the confirmation policy.
- Guideline
- Recommendations for Next-Generation Sequencing Germline Variant Confirmation
- Published by
- Association for Molecular Pathology and National Society of Genetic Counselors
- Citation
- Crooks KR et al. J Mol Diagn 2023;25(7):411-427
- Companion
- AMP white paper: Determining and Validating Germline Variant Confirmation Criteria (2023)
- Out of scope
- Tumor sequencing, liquid biopsy and noninvasive prenatal screening
The recommendations in plain language
- Have a written policy on orthogonal confirmation of NGS results.
- Have the policy overseen and approved by a qualified, certified professional with training and experience in clinical NGS.
- Use confirmation methods that are truly orthogonal (for example Sanger sequencing, a different NGS chemistry, MLPA or array for copy number), and investigate discrepancies.
- If you stop confirming some calls, base that on validated criteria from your own data (call quality, depth, allele balance, strand bias, genomic context) rather than on NGS performance in general.
- Make the policy, criteria and methods available on request, summarize the policy on every report, and flag exceptions.
- Apply special care to difficult variant types and regions: large indels, copy number variants, low-complexity and homologous regions (pseudogenes), and mosaic variants.
Validating criteria to skip confirmation
The AMP companion white paper describes how to build the data set: collect many NGS calls that were also tested by an orthogonal method, include true positives and false positives across variant types, then find quality thresholds that separate them. A threshold is only defensible if it catches all known false positives in the training data and is then checked on an independent set. Re-check the criteria after any change to the assay or pipeline.
Related germline standards
| Document | What it covers |
|---|---|
| ACMG technical standards: NGS for constitutional variants (2021 revision) | Test design, validation, quality metrics and reporting for germline panels, exomes and genomes. |
| ACMG/AMP variant classification (Richards et al., 2015) | Pathogenic to benign classification framework for germline variants. |
| CLSI MM09 (3rd ed., 2023) | Design, validation and quality management of sequencing tests, including germline applications. |
What CLIA requires for any NGS validation
| Requirement | What CLIA requires |
|---|---|
| Validation before patient testing | Most clinical NGS tests are laboratory developed or modified, so the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals (42 CFR 493.1253(b)(2)). |
| Personnel | High complexity testing: a qualified laboratory director (42 CFR 493.1443) and technical supervisor (42 CFR 493.1449) who approve the validation and procedures. |
| Facilities | Amplification steps outside closed systems need a unidirectional workflow (42 CFR 493.1101(a)(3)). |
| Quality control | Control procedures for each run, including controls that detect extraction errors and, for amplification, two control materials (42 CFR 493.1256(d)(3)(iv) and 42 CFR 493.1256(d)(3)(v)). |
| Proficiency testing | Most NGS analytes have no CLIA-regulated PT, so verify accuracy at least twice a year through external surveys, split samples or other means (42 CFR 493.1236(c)). |
| Records | Keep validation data, pipeline versions and QC records for at least two years after the procedure is discontinued (42 CFR 493.1105). |
Professional guidelines describe how to meet these requirements; CAP inspectors and New York State reviewers expect your validation to follow them or document why it differs. For New York patients, see New York genetic testing requirements.
Validating an NGS assay?
Our laboratory consultants write NGS validation plans, select reference materials, review pipeline validation, build the validation report your director signs, and prepare labs for CAP and New York State review, in every state.
Book a free consultationAll NGS validation guidelinesOr call (602) 874-1311 · info@soundlabconsult.com
Frequently asked questions
Does every germline NGS variant need Sanger confirmation?
No. The AMP/NSGC recommendations allow labs to skip orthogonal confirmation for calls that meet quality criteria the lab has validated with its own data, as long as the policy is written, approved and disclosed on reports.
Do these recommendations apply to tumor NGS?
No. They apply to germline DNA sequencing only; tumor sequencing, liquid biopsy and NIPT are out of scope.
Related NGS validation guides
Also see NGS lab consulting, clinical NGS workflow compliance, NGS proficiency testing and the CAP checklist and how NGS tests work.
Sources
- Crooks KR et al., J Mol Diagn 2023 (PubMed 37207865)
- AMP white paper: Determining and Validating Germline Variant Confirmation Criteria
Independent summary for laboratories, reviewed November 28, 2026. Guideline names belong to their publishers (AMP, CAP, CLSI, NSGC, ACMG, BLOODPAC, FDA); Sound Lab Consultants is not affiliated with them. Read the original documents before finalizing a validation. Not legal advice.