Short answer: This SOP explains how to set control limits, plot Levey-Jennings charts and apply Westgard multirules to accept or reject runs. This free template downloads as a fillable PDF and an editable Word file.
Last reviewed November 26, 2026. Governing requirement: 42 CFR 493.1256(d)(8).
These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use.
What this QC Review with Levey-Jennings Charts and Westgard Rules SOP must contain
Purpose: To evaluate quantitative QC consistently and detect shifts and trends.
Scope: Quantitative nonwaived tests.
The template includes these sections:
- Header block: laboratory name, CLIA number, document number, version, effective date, next review, author, approval by the laboratory director with signature and date
- Purpose and scope
- Responsibility by role
- Definitions
- Materials, equipment and records
- Step-by-step procedure (7 steps)
- Quality control
- Interpretation, reporting and documentation
- Limitations and cautions
- References to the governing regulation and guidance
- Lab-specific notes
- Revision history and annual review log (fillable)
Why labs need it
This procedure supports 42 CFR 493.1256(d)(8). Under 42 CFR 493.1251, a nonwaived laboratory must have written procedures that staff follow, and the laboratory director must approve, sign and date each procedure and every change before use. Waived labs must follow the manufacturer's instructions exactly, and written procedures help show they do. Surveyors and accreditation inspectors usually ask to see the procedure, the related records and evidence that staff were trained on it.
Procedure outline
The highlighted items are fill-in fields in the PDF and Word versions.
- Establish the mean and SD for each lot of each control.
- Plot each result on the Levey-Jennings chart.
- Apply rules: reject on 1-3s, 2-2s, R-4s, 4-1s, 10x [Rules in use].
- When a run is rejected, do not report patient results; troubleshoot (reagent, calibrator, instrument, control material).
- Document action and repeat results.
- Review charts weekly for shifts and trends.
- Update cumulative statistics monthly.
Quality control and documentation
- Follow the laboratory's QC policy and the manufacturer's instructions; record QC before reporting patient results.
- Stop patient testing and take corrective action when QC or acceptance criteria are not met.
- Record the activity, date, result and initials on the form or log, and have the supervisor review records [Review frequency].
- Keep records per the record retention SOP (at least 2 years under 42 CFR 493.1105 unless a longer period applies).
Limitations
- Manufacturer ranges for assayed controls are often too wide for lab use; establish lab limits.
- This is a generic template. Adapt it to your instruments, test systems, manufacturer instructions, state law and accreditation standards before use.
Key definitions
- Mean and SD: Calculated from at least [Number of data points, e.g. 20] control results over [Days].
- 1-2s warning: One control beyond 2 SD triggers inspection of other rules.
- 1-3s, 2-2s, R-4s, 4-1s, 10x: Common rejection rules.
Fields you fill in
Besides the header block, revision history and annual review log, this template has 4 lab-specific fields:
- Rules in use
- Review frequency
- Number of data points, e.g. 20
- Days
How to adopt this template
- Download the PDF or Word version above.
- Fill in the header block and every highlighted field. Where a field says "per manufacturer", copy the value from your current package insert or operator manual.
- Delete steps that do not apply and add any steps your test system or state requires.
- Have the technical consultant or technical supervisor review it, then have the laboratory director sign and date it before use.
- Train staff, document their review, and add the SOP to your document control master list.
Need procedures written for your lab?
Sound Lab Consultants writes and reviews complete procedure manuals that match your instruments, test menu and certificate type. Book a free consultation to talk through what your lab needs.
Book a free consultationFrequently asked questions
Is this QC Review with Levey-Jennings Charts and Westgard Rules SOP template free?
Yes. The QC Review with Levey-Jennings Charts and Westgard Rules SOP template is a free download as a fillable PDF and an editable Word (.docx) file. There is no email gate or sign-up.
What regulation does a qc review with levey-jennings charts and westgard rules procedure address?
This template is built around 42 CFR 493.1256(d)(8). Read the current text at the official source linked on this page, and check your state and accreditation organization for added requirements.
Can our lab use this SOP as-is?
No. Use it as a starting point. Fill in the highlighted fields, adapt the steps to your instruments and the manufacturer's instructions, and have the laboratory director approve, sign and date it before use (42 CFR 493.1251(d)).
Which labs is this template for?
It is written for moderate and high complexity laboratories, as marked in the template's 'Applies to' boxes. Uncheck any that do not apply to your lab.
How do I fill in the PDF?
Open the PDF in a PDF reader that supports forms, click the shaded fields and type. Save a copy. In the Word version, click each highlighted field (a content control) and type over it.
Official sources
- eCFR, 42 CFR 493.1256(d)(8)
- eCFR, 42 CFR Part 493 Laboratory Requirements (CLIA)
- CMS, State Operations Manual Appendix C: Survey procedures and interpretive guidelines for laboratories
- CLSI guidelines relevant to this procedure (referenced by name only; obtain current editions from CLSI)
- eCFR, 42 CFR 493.1256
Related SOP templates
These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use. Sound Lab Consultants is not affiliated with or endorsed by CMS, CDC, FDA, OSHA, HHS, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer. CLSI and other standards are referenced by name only. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.