Free lab SOP template · Analytic: QC, calibration, verification and equipment

Individualized Quality Control Plan (IQCP) SOP Template

Free, fillable individualized quality control plan (IQCP) standard operating procedure for CLIA moderate and high complexity labs. Download it as a PDF form or a Word document. No email required.

Short answer: An IQCP SOP covers the risk assessment, quality control plan and quality assessment needed to use alternative QC frequencies for a test system. This free template downloads as a fillable PDF and an editable Word file.

Last reviewed November 26, 2026. Governing requirement: CMS IQCP (42 CFR 493.1256(d) alternative QC).

Waived (n/a)Moderate complexityHigh complexity

These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use.

First page of the free Individualized Quality Control Plan (IQCP) SOP template, a fillable PDF and Word form for CLIA moderate and high complexity laboratories
Page 1 of the Individualized Quality Control Plan (IQCP) SOP template, with fillable header fields, applicability boxes and the director approval block.

What this Individualized Quality Control Plan (IQCP) SOP must contain

Purpose: To document a risk-based QC plan for a test system as allowed by CMS.

Scope: Nonwaived tests where the lab elects IQCP (not available for waived tests or pathology specialties excluded by CMS).

The template includes these sections:

  • Header block: laboratory name, CLIA number, document number, version, effective date, next review, author, approval by the laboratory director with signature and date
  • Purpose and scope
  • Responsibility by role
  • Definitions
  • Materials, equipment and records
  • Step-by-step procedure (6 steps)
  • Quality control
  • Interpretation, reporting and documentation
  • Limitations and cautions
  • References to the governing regulation and guidance
  • Lab-specific notes
  • Revision history and annual review log (fillable)

Why labs need it

This procedure supports CMS IQCP (42 CFR 493.1256(d) alternative QC). Under 42 CFR 493.1251, a nonwaived laboratory must have written procedures that staff follow, and the laboratory director must approve, sign and date each procedure and every change before use. Waived labs must follow the manufacturer's instructions exactly, and written procedures help show they do. Surveyors and accreditation inspectors usually ask to see the procedure, the related records and evidence that staff were trained on it.

Procedure outline

The highlighted items are fill-in fields in the PDF and Word versions.

  1. Define the test system, specimen, environment, reagents and testing personnel components.
  2. Perform a risk assessment of all five components across all phases.
  3. Write the quality control plan: control types, frequency [QC frequency in IQCP], and other measures (built-in controls, training).
  4. Ensure QC is never less than the manufacturer requires.
  5. Laboratory director signs and dates the IQCP before implementation.
  6. Quality assessment: monitor errors, complaints and QC trends; review the IQCP at least [Review frequency] and after failures.

Quality control and documentation

  • Follow the laboratory's QC policy and the manufacturer's instructions; record QC before reporting patient results.
  • Stop patient testing and take corrective action when QC or acceptance criteria are not met.
  • Record the activity, date, result and initials on the form or log, and have the supervisor review records [Review frequency].
  • Keep records per the record retention SOP (at least 2 years under 42 CFR 493.1105 unless a longer period applies).

Limitations

  • CMS excludes some specialties and situations; check current CMS guidance.
  • This is a generic template. Adapt it to your instruments, test systems, manufacturer instructions, state law and accreditation standards before use.

Key definitions

  • IQCP: Risk assessment + quality control plan + quality assessment, approved by the laboratory director.

Fields you fill in

Besides the header block, revision history and annual review log, this template has 2 lab-specific fields:

  • QC frequency in IQCP
  • Review frequency

How to adopt this template

  1. Download the PDF or Word version above.
  2. Fill in the header block and every highlighted field. Where a field says "per manufacturer", copy the value from your current package insert or operator manual.
  3. Delete steps that do not apply and add any steps your test system or state requires.
  4. Have the technical consultant or technical supervisor review it, then have the laboratory director sign and date it before use.
  5. Train staff, document their review, and add the SOP to your document control master list.

Need procedures written for your lab?

Sound Lab Consultants writes and reviews complete procedure manuals that match your instruments, test menu and certificate type. Book a free consultation to talk through what your lab needs.

Book a free consultation

Frequently asked questions

Is this Individualized Quality Control Plan (IQCP) SOP template free?

Yes. The Individualized Quality Control Plan (IQCP) SOP template is a free download as a fillable PDF and an editable Word (.docx) file. There is no email gate or sign-up.

What regulation does a individualized quality control plan (iqcp) procedure address?

This template is built around CMS IQCP (42 CFR 493.1256(d) alternative QC). Read the current text at the official source linked on this page, and check your state and accreditation organization for added requirements.

Can our lab use this SOP as-is?

No. Use it as a starting point. Fill in the highlighted fields, adapt the steps to your instruments and the manufacturer's instructions, and have the laboratory director approve, sign and date it before use (42 CFR 493.1251(d)).

Which labs is this template for?

It is written for moderate and high complexity laboratories, as marked in the template's 'Applies to' boxes. Uncheck any that do not apply to your lab.

How do I fill in the PDF?

Open the PDF in a PDF reader that supports forms, click the shaded fields and type. Save a copy. In the Word version, click each highlighted field (a content control) and type over it.

Official sources

These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use. Sound Lab Consultants is not affiliated with or endorsed by CMS, CDC, FDA, OSHA, HHS, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer. CLSI and other standards are referenced by name only. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.