Method Verification for FDA-Cleared or Approved Tests SOP Template

Free, fillable method verification for FDA-Cleared or approved tests standard operating procedure for CLIA moderate and high complexity labs. Download it as a PDF form or a Word document. No email required.

Short answer: A method verification SOP shows how the lab verifies accuracy, precision, reportable range and reference intervals for unmodified FDA-cleared tests before reporting patient results. This free template downloads as a fillable PDF and an editable Word file.

Last reviewed November 26, 2026. Governing requirement: 42 CFR 493.1253(b)(1).

Waived (n/a)Moderate complexityHigh complexity

These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use.

First page of the free Method Verification for FDA-Cleared or Approved Tests SOP template, a fillable PDF and Word form for CLIA moderate and high complexity laboratories
Page 1 of the Method Verification for FDA-Cleared or Approved Tests SOP template, with fillable header fields, applicability boxes and the director approval block.

What this Method Verification for FDA-Cleared or Approved Tests SOP must contain

Purpose: To verify the manufacturer's performance specifications for an unmodified FDA-cleared or approved nonwaived test before patient testing.

Scope: New tests and new instruments for unmodified moderate or high complexity FDA-cleared/approved test systems.

The template includes these sections:

  • Header block: laboratory name, CLIA number, document number, version, effective date, next review, author, approval by the laboratory director with signature and date
  • Purpose and scope
  • Responsibility by role
  • Definitions
  • Materials, equipment and records
  • Step-by-step procedure (8 steps)
  • Quality control
  • Interpretation, reporting and documentation
  • Limitations and cautions
  • References to the governing regulation and guidance
  • Fillable form: Verification summary (6 columns)
  • Lab-specific notes
  • Revision history and annual review log (fillable)

Why labs need it

This procedure supports 42 CFR 493.1253(b)(1). Under 42 CFR 493.1251, a nonwaived laboratory must have written procedures that staff follow, and the laboratory director must approve, sign and date each procedure and every change before use. Waived labs must follow the manufacturer's instructions exactly, and written procedures help show they do. Surveyors and accreditation inspectors usually ask to see the procedure, the related records and evidence that staff were trained on it.

Procedure outline

The highlighted items are fill-in fields in the PDF and Word versions.

  1. Write a verification plan with acceptance criteria approved by the director before starting [Acceptance criteria].
  2. Accuracy: test [Number] specimens spanning the reportable range by the new and comparative method; evaluate bias and agreement.
  3. Precision: run [Number] replicates of at least two levels within run and across [Number of days] days.
  4. Reportable range: test at least low, mid and high materials near the limits.
  5. Reference intervals: verify the manufacturer's or published intervals are appropriate for the lab's population (for example with 20 healthy-donor samples).
  6. Qualitative tests: test known positives and negatives, including near the cutoff.
  7. Summarize data, compare to criteria and document conclusions.
  8. Laboratory director signs approval before reporting patient results.

Quality control and documentation

  • Follow the laboratory's QC policy and the manufacturer's instructions; record QC before reporting patient results.
  • Stop patient testing and take corrective action when QC or acceptance criteria are not met.
  • Keep verification records as long as the test is in use, at least 2 years (493.1105(a)(3)(i)).

Limitations

  • Any modification to an FDA-cleared test makes it a validation under 493.1253(b)(2).
  • This is a generic template. Adapt it to your instruments, test systems, manufacturer instructions, state law and accreditation standards before use.

Key definitions

  • Accuracy: Agreement with a comparative method or known values.
  • Precision: Agreement of repeated measurements.
  • Reportable range: The range of results verified as reliable.

Fields you fill in

Besides the header block, revision history and annual review log, this template has 5 lab-specific fields:

  • Comparative method
  • Software
  • Acceptance criteria
  • Number
  • Number of days

How to adopt this template

  1. Download the PDF or Word version above.
  2. Fill in the header block and every highlighted field. Where a field says "per manufacturer", copy the value from your current package insert or operator manual.
  3. Delete steps that do not apply and add any steps your test system or state requires.
  4. Have the technical consultant or technical supervisor review it, then have the laboratory director sign and date it before use.
  5. Train staff, document their review, and add the SOP to your document control master list.

Need procedures written for your lab?

Sound Lab Consultants writes and reviews complete procedure manuals that match your instruments, test menu and certificate type. Book a free consultation to talk through what your lab needs.

Book a free consultation

Frequently asked questions

Is this Method Verification for FDA-Cleared or Approved Tests SOP template free?

Yes. The Method Verification for FDA-Cleared or Approved Tests SOP template is a free download as a fillable PDF and an editable Word (.docx) file. There is no email gate or sign-up.

What regulation does a method verification for fda-cleared or approved tests procedure address?

This template is built around 42 CFR 493.1253(b)(1). Read the current text at the official source linked on this page, and check your state and accreditation organization for added requirements.

Can our lab use this SOP as-is?

No. Use it as a starting point. Fill in the highlighted fields, adapt the steps to your instruments and the manufacturer's instructions, and have the laboratory director approve, sign and date it before use (42 CFR 493.1251(d)).

Which labs is this template for?

It is written for moderate and high complexity laboratories, as marked in the template's 'Applies to' boxes. Uncheck any that do not apply to your lab.

How do I fill in the PDF?

Open the PDF in a PDF reader that supports forms, click the shaded fields and type. Save a copy. In the Word version, click each highlighted field (a content control) and type over it.

Official sources

These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use. Sound Lab Consultants is not affiliated with or endorsed by CMS, CDC, FDA, OSHA, HHS, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer. CLSI and other standards are referenced by name only. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.