Sysmex XP-300 Hematology Analyzer Operator Course
Operator training for the Sysmex XP-300 compact 3-part differential hematology analyzer: specimens, QC, running CBCs, flags and maintenance.
Free45 min6 lessons10-question quiz
This course covers general operator principles for the Sysmex XP-300. Always follow the manufacturer's current instructions for use and your lab's written procedures. Product names are trademarks of their owners; Sound Lab Consultants is not affiliated with the manufacturer.
Lessons
Lesson 1: About the analyzer
The Sysmex XP-300 is a compact automated hematology analyzer that reports a complete blood count (CBC) with a 3-part white cell differential. It is common in physician office labs, clinics and small hospitals. CBC testing on this analyzer is non-waived, so CLIA personnel, QC, proficiency testing and competency requirements apply.
Lesson 2: Specimen requirements
- Use EDTA (lavender top) whole blood, filled to the correct volume and mixed by gentle inversion right after collection.
- Check for clots before running; a clotted specimen gives falsely low counts and must be recollected.
- Run within the stability time in the IFU, and mix again just before aspiration.
- Follow the IFU for pre-dilute mode on small samples.
Lesson 3: Startup, background check and running samples
- Check reagent levels and expiration dates, and record new reagent lots.
- Perform startup; the analyzer runs a background check. Do not test patients until background passes.
- Run controls (see the next lesson).
- Enter the sample ID, mix the tube, aspirate as shown in the IFU, and review results and flags.
Lesson 4: Quality control and calibration
Use the manufacturer's hematology control material at the levels and frequency your QC plan requires, and verify calibration at least every 6 months (calibration verification) or as the manufacturer and CLIA require.
Quality control (QC) proves the test system is working before you report patient results. Run QC at the frequency required by the manufacturer, CLIA (42 CFR 493.1256) and your lab's policy, and whenever you open a new reagent lot or shipment, after major maintenance, and after troubleshooting. Many labs use an Individualized Quality Control Plan (IQCP) for eligible non-waived tests; if yours does, follow the frequencies written in it.
- Record every QC result, the lot numbers and expiration dates, and who ran it.
- Do not report patient results when QC is out of range. Stop, troubleshoot, document corrective action, rerun QC, and review patient results since the last acceptable QC.
- Store controls and reagents at the temperatures on the package insert and never use them past their expiration date.
Lesson 5: Flags and when to make a smear
Analyzer flags (abnormal distributions, interference, counts outside the linear range) mean the result needs review. Your lab must have written criteria for when to repeat, dilute, review a peripheral smear or refer the sample. Never release a flagged result without following that procedure.
Lesson 6: Maintenance, reporting and competency
- Perform daily shutdown and cleaning and the periodic maintenance in the IFU, and log it.
- Dispose of waste as biohazard according to your safety plan.
- Identify the patient with at least two identifiers before collecting or testing.
- Report results with units and reference ranges, and follow your lab's critical value policy: notify the responsible provider promptly and document who was told, when, and read-back.
- Correct errors with a documented corrected report; never erase or overwrite original records.
- Keep test records, QC and maintenance logs for at least 2 years (longer where your state or accreditor requires).
Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy.
This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.
Quiz and certificate
Answer all 10 questions. Score 80% or higher to get your free certificate from Sound Lab Consultants - Technical Consultants.
Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy. This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.