Clinical laboratory consulting in all 50 states, remote and on-site Call (602) 874-1311
Menu
Services
Start, license and accredit
New lab setup and startup programClinical laboratory consultingLab design and layoutLab Startup Concierge (vendor offers)Manufacturer lab startup programsHow to start a clinical laboratoryLab setup checklistCMS-116 CLIA applicationCLIA certificate changesCLIA certificate renewalMulti-state lab licensingState lab licenses (all 50 states)New York CLEP permits and LDT approvalCAP and COLA accreditation
Lab leadership and personnel
CLIA laboratory directorLab consultantCLIA consultantLaboratory director servicesInterim laboratory directorCLIA technical consultantTechnical supervisorClinical consultantGeneral supervisorRemote lab managementTesting personnel qualificationsTechnical consultant scope of workCompetency assessment
Inspections and remediation
Mock CLIA inspectionCAP inspection preparationCOLA survey preparationACHC survey preparationISO 15189 accreditationFailed CLIA, CAP or COLA surveyPlan of correctionEvidence of correctionCorrective action (CAPA)Unsuccessful proficiency testingProficiency testing and QCCease-testing lettersSanctions and civil money penaltiesCLIA expert witness
Quality, operations and finance
QMS, IQCP and instrument validationMethod verificationSOP writingLDT development and validationLab instrument expertsLab management consultingWorkflow optimizationReference lab consultingHospital laboratory managementSend-out vs in-house analysisClinical lab billingLab revenue cycle managementLab outreach programsLab pro forma calculatorSkilled nursing facility labs
Lab software and informatics
LIS verification and validationLIS and LIMS QC module setupQC software selection and implementationEMR and result interface verificationBio-Rad Unity QC consultingUnity Real Time Westgard rules setupNGS bioinformatics pipeline validationAI diagnostics complianceLab billing systems we work withMyLabCompliance.io compliance softwareAll lab consulting services →Get help with your lab →
Specialties
Resources
New lab setup Failed survey help Guides Free courses Lab billing Get help State licenses Locations Careers Contact Book a consultation
Sound Lab Consultantsby Sound Lab Consultants
Sound Lab Consultants › Lab training courses › Sysmex XP-300 Hematology Analyzer Operator Course

Sysmex XP-300 Hematology Analyzer Operator Course

Operator training for the Sysmex XP-300 compact 3-part differential hematology analyzer: specimens, QC, running CBCs, flags and maintenance.

Free45 min6 lessons10-question quiz

This course covers general operator principles for the Sysmex XP-300. Always follow the manufacturer's current instructions for use and your lab's written procedures. Product names are trademarks of their owners; Sound Lab Consultants is not affiliated with the manufacturer.

Lessons

Lesson 1: About the analyzer

The Sysmex XP-300 is a compact automated hematology analyzer that reports a complete blood count (CBC) with a 3-part white cell differential. It is common in physician office labs, clinics and small hospitals. CBC testing on this analyzer is non-waived, so CLIA personnel, QC, proficiency testing and competency requirements apply.

Lesson 2: Specimen requirements
Lesson 3: Startup, background check and running samples
  1. Check reagent levels and expiration dates, and record new reagent lots.
  2. Perform startup; the analyzer runs a background check. Do not test patients until background passes.
  3. Run controls (see the next lesson).
  4. Enter the sample ID, mix the tube, aspirate as shown in the IFU, and review results and flags.
Lesson 4: Quality control and calibration

Use the manufacturer's hematology control material at the levels and frequency your QC plan requires, and verify calibration at least every 6 months (calibration verification) or as the manufacturer and CLIA require.

Quality control (QC) proves the test system is working before you report patient results. Run QC at the frequency required by the manufacturer, CLIA (42 CFR 493.1256) and your lab's policy, and whenever you open a new reagent lot or shipment, after major maintenance, and after troubleshooting. Many labs use an Individualized Quality Control Plan (IQCP) for eligible non-waived tests; if yours does, follow the frequencies written in it.

Lesson 5: Flags and when to make a smear

Analyzer flags (abnormal distributions, interference, counts outside the linear range) mean the result needs review. Your lab must have written criteria for when to repeat, dilute, review a peripheral smear or refer the sample. Never release a flagged result without following that procedure.

Lesson 6: Maintenance, reporting and competency

Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy.

This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.

Quiz and certificate

Answer all 10 questions. Score 80% or higher to get your free certificate from Sound Lab Consultants - Technical Consultants.

We use your details to issue and email your certificate and keep a record of completion. See how we help labs at /lab-help.

1. Which tube is used for a CBC on the XP-300?

2. A clotted EDTA sample should be:

3. Before testing patients each day the analyzer must pass:

4. CBC testing on the XP-300 is:

5. Calibration verification is required at least every:

6. A result has an analyzer flag. You should:

7. Just before aspiration the EDTA tube should be:

8. New reagent lots should be:

9. Test records and QC logs must be kept for at least:

10. Who performs your ongoing CLIA competency assessment?

Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy. This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.