Abaxis Piccolo Xpress Chemistry Analyzer Operator Course
Operator training for the Piccolo Xpress point-of-care chemistry analyzer: reagent discs, specimens, built-in checks, QC and reporting.
Free40 min6 lessons10-question quiz
This course covers general operator principles for the Piccolo Xpress. Always follow the manufacturer's current instructions for use and your lab's written procedures. Product names are trademarks of their owners; Sound Lab Consultants is not affiliated with the manufacturer.
Lessons
Lesson 1: About the analyzer
The Piccolo Xpress is a portable chemistry analyzer that runs single-use reagent discs, each producing a panel of results (for example metabolic, lipid or liver panels) from a small sample. Some panels are CLIA-waived and others are moderate complexity, so check the FDA CLIA database for each disc your lab uses and follow the requirements for that category.
Lesson 2: Discs and storage
- Store reagent discs as the package insert directs and check expiration dates.
- Bring a disc to the required temperature before use and use it within the time the insert allows once the pouch is opened.
- Handle the disc by its edges; never touch the optical surfaces.
Lesson 3: Specimens
- Use only the specimen types listed for the disc (whole blood, plasma or serum, with the anticoagulant specified).
- Fill the disc with the exact volume indicated; underfilled or overfilled discs give errors or wrong results.
- Note hemolysis, lipemia and icterus; the analyzer may suppress affected results as described in the IFU.
Lesson 4: Running a disc and built-in checks
- Enter the operator and patient IDs.
- Load the sample into the disc as shown in the IFU and insert the disc.
- The analyzer performs internal quality checks on every run; if a check fails, results are suppressed and an error is shown.
- Review results and any suppressed analytes before reporting.
Lesson 5: Quality control
Built-in checks do not replace external QC. Run external controls at the frequency required by the manufacturer, your QC plan and CLIA (for non-waived panels), and with each new lot or shipment of discs.
Quality control (QC) proves the test system is working before you report patient results. Run QC at the frequency required by the manufacturer, CLIA (42 CFR 493.1256) and your lab's policy, and whenever you open a new reagent lot or shipment, after major maintenance, and after troubleshooting. Many labs use an Individualized Quality Control Plan (IQCP) for eligible non-waived tests; if yours does, follow the frequencies written in it.
- Record every QC result, the lot numbers and expiration dates, and who ran it.
- Do not report patient results when QC is out of range. Stop, troubleshoot, document corrective action, rerun QC, and review patient results since the last acceptable QC.
- Store controls and reagents at the temperatures on the package insert and never use them past their expiration date.
Lesson 6: Reporting and competency
- Identify the patient with at least two identifiers before collecting or testing.
- Report results with units and reference ranges, and follow your lab's critical value policy: notify the responsible provider promptly and document who was told, when, and read-back.
- Correct errors with a documented corrected report; never erase or overwrite original records.
- Keep test records, QC and maintenance logs for at least 2 years (longer where your state or accreditor requires).
Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy.
This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.
Quiz and certificate
Answer all 10 questions. Score 80% or higher to get your free certificate from Sound Lab Consultants - Technical Consultants.
Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy. This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.