Transcription-Mediated Amplification (TMA)
Transcription-mediated amplification (TMA) is an isothermal nucleic acid amplification method that copies an RNA target using reverse transcriptase and RNA polymerase at a single temperature, producing billions of RNA copies in under an hour. It is widely used for chlamydia, gonorrhea, trichomonas, Mycoplasma genitalium and HPV mRNA testing.
- Specimen
- Vaginal, endocervical, urethral, throat and rectal swabs and urine in the assay's own collection kit.
- CLIA complexity
- High complexity
- CLIA specialty
- Molecular diagnostics
- Common CPT codes
- 87491 / 87591, 87661, 87563
How it works
- Target capture: Magnetic particles with capture probes pull the target RNA out of the lysed specimen.
- Reverse transcription: A primer carrying an RNA polymerase promoter binds the RNA, and reverse transcriptase makes a DNA copy.
- Transcription: RNA polymerase transcribes 100 to 1,000 RNA copies from each DNA template, and each new RNA re-enters the cycle.
- Detection: Chemiluminescent or fluorescent probes (hybridization protection or real-time molecular torches) report the target.
TMA compared with PCR
| Feature | TMA | PCR |
|---|---|---|
| Temperature | One constant temperature (isothermal) | Repeated cycling |
| Main target | RNA (often ribosomal RNA, present in many copies) | DNA (RNA after RT) |
| Product | RNA amplicons, which degrade more easily | DNA amplicons, very stable |
| Typical platform | Hologic Panther and Panther Fusion (Aptima assays) | Many open and closed platforms |
Specimen collection
- Vaginal, endocervical, urethral, throat and rectal swabs and urine in the assay's own collection kit.
- ThinPrep liquid Pap specimens for some HPV and STI assays.
Accuracy, results and interpretation
- Targeting high-copy ribosomal RNA gives TMA very high analytical sensitivity.
- Results are qualitative (positive, negative or equivocal and repeat).
Methods, platforms and CLIA complexity
| Approach | Examples | Typical CLIA category |
|---|---|---|
| Aptima assays on Panther | Combo 2 (CT/NG), Trichomonas vaginalis, Mycoplasma genitalium, HPV, BV and CV/TV | Typically high complexity |
CLIA complexity is assigned to each specific test system (instrument, reagent and specimen type), not to a method in general. Confirm yours in the FDA CLIA database or our weekly list of new FDA CLIA categorizations.
Billing codes
| Code | Description (summarized) |
|---|---|
| 87491 / 87591 | Chlamydia / gonorrhea, amplified probe |
| 87661 | Trichomonas vaginalis, amplified probe |
| 87563 | Mycoplasma genitalium, amplified probe |
| 87624 / 87625 | HPV high-risk types, pooled / genotyping 16, 18 (45) |
CPT is a registered trademark of the American Medical Association. Descriptions are summarized; check the current CPT code set, your MAC's coverage articles and each payer's policy before billing.
Offering this test in your lab
- Panther-class instruments need floor space, dedicated power and a service contract; they suit labs above a few hundred STI specimens a week.
- Even closed, automated systems need the unidirectional workflow rules if any step happens outside the closed system.
CLIA requirements checklist
| Requirement | What CLIA requires |
|---|---|
| CLIA certificate | A Certificate of Waiver covers only test systems FDA has categorized as waived (42 CFR 493.15). Moderate and high complexity testing needs a Certificate of Compliance or a Certificate of Accreditation (CAP, COLA, The Joint Commission and others) after a Registration certificate (42 CFR 493.43). Add the molecular diagnostics specialty to your certificate before reporting results. |
| Personnel | High complexity: laboratory director (42 CFR 493.1443), technical supervisor (42 CFR 493.1449), general supervisor (42 CFR 493.1461) and qualified testing personnel (42 CFR 493.1489). Moderate complexity: director (42 CFR 493.1405), technical consultant (42 CFR 493.1411) and testing personnel (42 CFR 493.1423). |
| Facility and workflow | Amplification not contained in a closed system needs a unidirectional workflow with separate areas for specimen preparation, amplification and detection, and reagent preparation (42 CFR 493.1101(a)(3)). |
| Before patient testing | Verify accuracy, precision, reportable range and reference intervals for an unmodified FDA-cleared test; fully validate a modified test or LDT (42 CFR 493.1253). |
| Quality control | Control procedures for every run under 42 CFR 493.1256, including two controls for each molecular amplification procedure and an inhibition control where inhibition causes false negatives (42 CFR 493.1256(d)(3)(v)). |
| Proficiency testing | Where no approved proficiency testing exists for a test, verify its accuracy at least twice a year, for example by split-sample comparison or an available external survey (42 CFR 493.1236(c)). |
| Competency | Assess each person's competency semiannually in the first year and annually after that, using the six required methods (42 CFR 493.1235 with the director and supervisor duties in 42 CFR 493.1451). |
| Records | Keep QC, PT, maintenance and test records for at least 2 years, and longer for some specialties and states (42 CFR 493.1105). |
For specialty-level requirements and common survey findings, see Molecular diagnostics lab consulting and how to start a molecular PCR lab.
How Sound Lab Consultants helps labs offering TMA assays
Bringing TMA assays in-house?
Our laboratory consultants scope the test menu, pick the platform, write the validation plan and procedures, and provide the director and technical oversight CLIA requires, for clinics, urgent cares, hospitals and independent labs in every state.
Book a free consultationAll lab test guidesOr call (602) 874-1311 · info@soundlabconsult.com
Frequently asked questions
Is TMA more sensitive than PCR?
Both are highly sensitive. TMA often targets ribosomal RNA, which exists in thousands of copies per cell, which can give it very low detection limits for organisms like chlamydia and trichomonas.
Is TMA a NAAT?
Yes. TMA is a nucleic acid amplification test (NAAT), as are PCR, SDA, LAMP and NEAR.
Related lab test guides