Blood tube fill volume, inversion and clotting standards
Fill, mix and process tubes correctly so results are accurate the first time.
Tube fill and mixing standards
- Evacuated tubes draw a set volume. Let each fill until blood stops flowing. Short-draw (partial-volume) tubes exist for pediatric and difficult draws.
- Light blue citrate tubes must be filled to the fill line, at least 90% of the stated volume, to keep the 9:1 blood-to-citrate ratio. Underfilled citrate tubes falsely prolong PT and aPTT and are rejected.
- For patients with a hematocrit above 55%, the citrate volume must be adjusted (CLSI H21). Ask the laboratory for a special tube.
- Underfilled EDTA tubes have excess anticoagulant, which shrinks cells and can falsely lower the hematocrit and change cell counts.
- Invert gently, end over end, the number of times the manufacturer specifies. Never shake; it causes hemolysis.
| Tube (BD) | Inversions | Clotting time before centrifuge |
|---|---|---|
| Light blue citrate | 3 to 4 | Not applicable (plasma) |
| Red plastic serum | 5 | 60 minutes |
| Gold SST | 5 | 30 minutes |
| Orange RST | 5 to 6 | 5 minutes |
| Green heparin and PST | 8 to 10 | Not applicable (plasma) |
| Lavender, pink, pearl EDTA | 8 to 10 | Not applicable (whole blood or plasma) |
| Gray fluoride | 8 to 10 | Not applicable (plasma) |
Inversion counts and clotting times are BD Vacutainer values. Follow the instructions for the brand of tube you use.
Handling, processing and transport
- Keep tubes upright, closed and at the temperature the test requires.
- Let serum tubes clot fully before centrifuging (see clotting times above); centrifuge at the speed and time in your procedure.
- Separate serum or plasma from cells within the time limit for each test.
- Ship specimens to reference labs as UN3373 Biological Substance, Category B, in triple packaging, by trained staff. See our free UN3373 shipping course.
Common rejection reasons
- Unlabeled or mislabeled specimen
- Wrong tube type or expired tube
- Underfilled citrate tube (quantity not sufficient)
- Clotted anticoagulated specimen
- Hemolyzed specimen for affected tests
- Specimen too old or stored at the wrong temperature
Preanalytical errors and how they change results
| Error | Effect on results |
|---|---|
| Hemolysis (small needle, shaking, forcing blood through a syringe, milking) | Falsely high potassium, LD, AST, magnesium and phosphorus |
| EDTA carryover (wrong order of draw) | Falsely high potassium and falsely low calcium, magnesium and alkaline phosphatase |
| Tourniquet longer than 1 minute or fist pumping | Hemoconcentration: higher proteins, calcium, cholesterol; fist pumping raises potassium |
| Drawing above an IV | Dilution and contamination: glucose, electrolytes or drug levels match the IV fluid |
| Underfilled light blue tube | Falsely prolonged PT and aPTT |
| Clotted EDTA tube (poor mixing) | Falsely low platelet count and unreliable CBC |
| Delayed separation of serum or plasma | Falling glucose and rising potassium as cells keep metabolizing; CLSI guidance is to separate within 2 hours unless stability data allow longer |
| Wrong tube or wrong patient label | Results reported on the wrong patient or rejected specimens; the most dangerous error |
More phlebotomy resources
- Complete preanalytical phlebotomy guide
- Order of draw chart and infographic
- Blood collection tube colors and additives
- Tube fill volume, inversion and clotting standards
- Free phlebotomy quiz
- Free phlebotomy standards course with certificate
- Specimen processing and transport course
- Phlebotomist jobs
Frequently asked questions
How full does a light blue tube need to be?
Fill it to the fill line, at least 90% of the stated draw volume, to keep the 9:1 blood-to-citrate ratio. Underfilled tubes are rejected because they falsely prolong clotting times.
How many times should you invert blood tubes?
For BD Vacutainer tubes: light blue 3 to 4, red plastic and gold SST 5, orange RST 5 to 6, and green, lavender, pink and gray 8 to 10. Glass red tubes need no inversion. Always follow your tube manufacturer's instructions; never shake.
What happens if an EDTA tube is underfilled?
Excess EDTA shrinks red cells and can falsely lower the hematocrit and change other CBC values. Follow your lab's minimum fill limits.
How long should serum tubes clot before centrifuging?
For BD tubes: gold SST about 30 minutes, red plastic about 60 minutes and orange RST about 5 minutes. Follow your tube manufacturer's instructions.
What if the patient's hematocrit is above 55%?
CLSI H21 calls for adjusting the citrate volume for coagulation testing when the hematocrit is above 55%. Contact the laboratory for an adjusted tube.