Short answer: A POCT program SOP defines oversight for bedside and clinic testing: approved tests, operator certification, QC, connectivity, result documentation and comparison with the main lab. This free template downloads as a fillable PDF and an editable Word file.
Last reviewed November 26, 2026. Governing requirement: 42 CFR 493.1256, 493.1281 and 493.15.
These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use.
What this Point-of-Care Testing (POCT) Program SOP must contain
Purpose: To manage point-of-care testing across sites so it meets CLIA and facility requirements.
Scope: All POCT devices and operators under [CLIA number(s)].
The template includes these sections:
- Header block: laboratory name, CLIA number, document number, version, effective date, next review, author, approval by the laboratory director with signature and date
- Purpose and scope
- Responsibility by role
- Definitions
- Materials, equipment and records
- Step-by-step procedure (7 steps)
- Quality control
- Interpretation, reporting and documentation
- Limitations and cautions
- References to the governing regulation and guidance
- Lab-specific notes
- Revision history and annual review log (fillable)
Why labs need it
This procedure supports 42 CFR 493.1256, 493.1281 and 493.15. Under 42 CFR 493.1251, a nonwaived laboratory must have written procedures that staff follow, and the laboratory director must approve, sign and date each procedure and every change before use. Waived labs must follow the manufacturer's instructions exactly, and written procedures help show they do. Surveyors and accreditation inspectors usually ask to see the procedure, the related records and evidence that staff were trained on it.
Procedure outline
The highlighted items are fill-in fields in the PDF and Word versions.
- Approve new POCT tests through the lab director; verify categorization.
- Train and certify operators; lock out uncertified operators in devices where possible.
- Ensure QC per manufacturer (waived) or per CLIA/IQCP (nonwaived).
- Ensure results are documented in the patient record.
- Compare POCT results with the main lab at least twice a year when the same analyte is tested (493.1281).
- Review QC lockouts, operator activity and errors monthly.
- Recertify operators [Recertification interval].
Quality control and documentation
- Follow the laboratory's QC policy and the manufacturer's instructions; record QC before reporting patient results.
- Stop patient testing and take corrective action when QC or acceptance criteria are not met.
- Record the activity, date, result and initials on the form or log, and have the supervisor review records [Review frequency].
- Keep records per the record retention SOP (at least 2 years under 42 CFR 493.1105 unless a longer period applies).
Limitations
- This is a generic template. Adapt it to your instruments, test systems, manufacturer instructions, state law and accreditation standards before use.
Key definitions
- POCT coordinator: Person who manages devices, operators, QC and records [Coordinator name].
Fields you fill in
Besides the header block, revision history and annual review log, this template has 4 lab-specific fields:
- CLIA number(s)
- Recertification interval
- Review frequency
- Coordinator name
How to adopt this template
- Download the PDF or Word version above.
- Fill in the header block and every highlighted field. Where a field says "per manufacturer", copy the value from your current package insert or operator manual.
- Delete steps that do not apply and add any steps your test system or state requires.
- Have the technical consultant or technical supervisor review it, then have the laboratory director sign and date it before use.
- Train staff, document their review, and add the SOP to your document control master list.
Need procedures written for your lab?
Sound Lab Consultants writes and reviews complete procedure manuals that match your instruments, test menu and certificate type. Book a free consultation to talk through what your lab needs.
Book a free consultationFrequently asked questions
Is this Point-of-Care Testing (POCT) Program SOP template free?
Yes. The Point-of-Care Testing (POCT) Program SOP template is a free download as a fillable PDF and an editable Word (.docx) file. There is no email gate or sign-up.
What regulation does a point-of-care testing (poct) program procedure address?
This template is built around 42 CFR 493.1256, 493.1281 and 493.15. Read the current text at the official source linked on this page, and check your state and accreditation organization for added requirements.
Can our lab use this SOP as-is?
No. Use it as a starting point. Fill in the highlighted fields, adapt the steps to your instruments and the manufacturer's instructions, and have the laboratory director approve, sign and date it before use (42 CFR 493.1251(d)).
Which labs is this template for?
It is written for waived, moderate and high complexity laboratories, as marked in the template's 'Applies to' boxes. Uncheck any that do not apply to your lab.
How do I fill in the PDF?
Open the PDF in a PDF reader that supports forms, click the shaded fields and type. Save a copy. In the Word version, click each highlighted field (a content control) and type over it.
Official sources
- eCFR, 42 CFR 493.1256, 493.1281 and 493.15
- CMS, State Operations Manual Appendix C: Survey procedures and interpretive guidelines for laboratories
- CDC, Ready? Set? Test! Patient testing is important (waived testing booklet)
- eCFR, 42 CFR 493.15
- eCFR, 42 CFR 493.1256
- eCFR, 42 CFR 493.1281
Related SOP templates
These templates are free, general starting points, not legal or regulatory advice. Every template must be reviewed and approved, signed and dated by your laboratory director, and adapted to your lab's instruments, test systems and the manufacturer's instructions, your state rules and your accreditation organization's standards before use. Sound Lab Consultants is not affiliated with or endorsed by CMS, CDC, FDA, OSHA, HHS, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer. CLSI and other standards are referenced by name only. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.
