Tosoh AIA-360 Immunoassay Analyzer Operator Course

Operator training for the Tosoh AIA-360 compact immunoassay analyzer: test cups, calibration, QC, maintenance and result review.

Free45 min6 lessons10-question quiz

This course covers general operator principles for the Tosoh AIA-360. Always follow the manufacturer's current instructions for use and your lab's written procedures. Product names are trademarks of their owners; Sound Lab Consultants is not affiliated with the manufacturer.

Lessons

Lesson 1: About the analyzer

The Tosoh AIA-360 is a compact automated immunoassay analyzer that uses single-use test cups containing dried reagents and magnetic beads with fluorescent detection. Labs use it for hormones, tumor markers, cardiac and other immunoassays. These assays are non-waived (moderate or high complexity depending on the assay), so full CLIA personnel, QC, calibration, proficiency testing and competency requirements apply.

Lesson 2: Reagents and specimens
Lesson 3: Calibration

Each assay must be calibrated as the manufacturer requires, including with each new reagent lot and at the interval stated in the insert, and recalibrated after major maintenance or when QC indicates a shift. Review and accept calibration curves only when they meet the manufacturer's criteria, and perform CLIA calibration verification at least every 6 months where required.

Lesson 4: Running samples and reviewing results
  1. Complete daily startup checks and maintenance.
  2. Run QC and confirm it is acceptable.
  3. Load samples and test cups, program the tests and start the run.
  4. Review results, flags and any values above the measuring range; dilute and rerun as the insert allows or report as greater than the upper limit.
Lesson 5: Quality control and maintenance

Quality control (QC) proves the test system is working before you report patient results. Run QC at the frequency required by the manufacturer, CLIA (42 CFR 493.1256) and your lab's policy, and whenever you open a new reagent lot or shipment, after major maintenance, and after troubleshooting. Many labs use an Individualized Quality Control Plan (IQCP) for eligible non-waived tests; if yours does, follow the frequencies written in it.

Lesson 6: Reporting and competency

Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy.

This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.

Quiz and certificate

Answer all 10 questions. Score 80% or higher to get your free certificate from Sound Lab Consultants - Technical Consultants.

We use your details to issue and email your certificate and keep a record of completion. See how we help labs at /lab-help.

1. AIA-360 assays use:

2. Assays on the AIA-360 are:

3. Calibration is required with:

4. Calibration verification is generally required at least every:

5. A result above the measuring range should be:

6. Specimens should be free of:

7. Before patient testing each day you should:

8. If QC shows a shift you may need to:

9. Maintenance should be:

10. Competency for non-waived testing is assessed using:

Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy. This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.