Quidel Sofia 2 Fluorescent Immunoassay Analyzer Operator Course

Operator training for the Sofia 2 rapid fluorescent immunoassay analyzer: specimen collection, Walk Away and Read Now modes, calibration checks and external QC.

Free30 min6 lessons10-question quiz

This course covers general operator principles for the Sofia 2. Always follow the manufacturer's current instructions for use and your lab's written procedures. Product names are trademarks of their owners; Sound Lab Consultants is not affiliated with the manufacturer.

Lessons

Lesson 1: About the analyzer

Sofia 2 is a benchtop analyzer that reads lateral-flow fluorescent immunoassay cassettes for rapid tests such as influenza A/B, SARS-CoV-2 antigen, strep A and RSV. Many of these assays are CLIA-waived for specific specimen types; using a specimen type or procedure not in the waived package insert makes the test non-waived. Always follow the current package insert for the kit you are using.

Lesson 2: Specimen collection
Lesson 3: Running a test: Walk Away and Read Now

In Walk Away mode the cassette is inserted right away and the analyzer times development and reads it. In Read Now mode the operator times development outside the analyzer and inserts the cassette when the development time is complete. Timing errors cause false results, so follow the insert exactly for the mode you use.

Lesson 4: Calibration check and QC
Lesson 5: Results and reporting
Lesson 6: Waived testing responsibilities

Waived labs must hold a CLIA Certificate of Waiver (or higher), follow the manufacturer's instructions and allow inspections. Good practice, recommended by CDC, includes training records, QC logs and periodic competency checks for every operator.

Quiz and certificate

Answer all 10 questions. Score 80% or higher to get your free certificate from Sound Lab Consultants - Technical Consultants.

We use your details to issue and email your certificate and keep a record of completion. See how we help labs at /lab-help.

1. What does Sofia 2 read?

2. Using a specimen type not in the waived insert makes the test:

3. In Walk Away mode:

4. In Read Now mode:

5. External controls should be run with:

6. Do built-in procedural controls replace external QC?

7. An invalid result should be:

8. Timing errors can cause:

9. The calibration check uses:

10. Waived labs must:

Under CLIA, everyone who performs non-waived testing must have documented training before reporting patient results, and documented competency assessment at 6 months and 12 months in the first year and annually after that. The six required methods are direct observation of routine testing, monitoring of recording and reporting of results, review of worksheets, QC and maintenance records, direct observation of maintenance and function checks, testing of previously analyzed or external samples, and assessment of problem-solving skills (42 CFR 493.1451 and 493.1413). Waived testing follows the manufacturer's instructions and your lab's own training and competency policy. This course documents training. It does not replace the competency assessment your laboratory director or technical consultant must perform at your site.