SOPs and the CLIA procedure manual for a new lab

Your procedure manual is the first thing a surveyor reads. Here is what CLIA requires in it and how to keep it controlled.

Short answer: CLIA requires a written procedure manual for every test, available to testing staff, approved, signed and dated by the laboratory director before use and whenever it changes. Each procedure must cover specimen collection, processing and rejection criteria; step-by-step performance including calculations; preparation of reagents, calibrators and controls; calibration and calibration verification; reportable range; quality control; remedial actions; limitations; reference intervals; critical values; references; specimen storage; how results are reported; and what to do when the test system is down. Retired procedures are kept for at least two years after use ends.

Last reviewed November 26, 2026. Primary sources: eCFR, eCFR, eCFR.

Sound Lab Consultants is an independent, vendor-neutral consulting firm. We do not sell instruments or reagents and are not affiliated with any manufacturer, distributor or supplier startup program.

SOPs and the CLIA procedure manual for a new lab overview graphic from Sound Lab Consultants

What each procedure must include

  • Patient preparation, specimen collection, labeling, storage, preservation, transport, processing, referral and rejection criteria.
  • Microscopic examination steps where applicable.
  • Step-by-step performance, including calculations and interpretation.
  • Preparation of slides, solutions, calibrators, controls, reagents and other materials.
  • Calibration and calibration verification procedures.
  • Reportable range for patient results.
  • Control procedures and corrective action when calibration or controls fail.
  • Limitations in methods, including interfering substances.
  • Reference intervals and imminently life-threatening (critical) results.
  • Literature references.
  • Specimen storage and preservation to ensure integrity until testing is done.
  • Reporting results, including critical value reporting.
  • The course of action if the test system becomes inoperable.

Policies a new lab needs beyond test procedures

  • Quality assessment plan covering pre-analytic, analytic and post-analytic phases.
  • Proficiency testing handling (no inter-lab communication or referral of PT samples).
  • Personnel training and competency assessment.
  • Safety: bloodborne pathogens exposure control plan and chemical hygiene plan.
  • Specimen referral, result reporting, corrected reports and record retention.

Document control

  • Director approval, signature and date before first use and after every change.
  • Version history and a list of who was trained on each change.
  • Manufacturer package inserts can be used as part of a procedure, but lab-specific steps still need to be written.
  • Keep retired procedures at least two years after use ends (42 CFR 493.1105).

How we help

We write lab-specific procedures from your instruments and validation data, build the quality system and train staff. See CLIA compliance checklist and the lab startup program.

Start your lab with a plan

Book a free consultation. Tell us your test menu, location and target date, and we will lay out the licensing, validation and staffing steps in order.

Book a free consultation

Frequently asked questions

Does CLIA require a written procedure manual?

Yes. 42 CFR 493.1251 requires a written procedure manual for all tests, approved, signed and dated by the laboratory director.

Can I use the manufacturer's package insert as my SOP?

Package inserts can be used as part of the procedure, but the lab must add lab-specific information such as QC, reporting, critical values and actions when the system is down.

How long must old procedures be kept?

At least two years after a procedure is discontinued, under 42 CFR 493.1105.

Who approves lab procedures?

The laboratory director must approve, sign and date each procedure and each change before use.

Can you write our SOPs?

Yes. We write procedure manuals and quality systems for new labs. Book a free consultation.

Official sources

Sound Lab Consultants is an independent, vendor-neutral consulting firm. We do not sell instruments or reagents and are not affiliated with any manufacturer, distributor or supplier startup program. Sound Lab Consultants is not affiliated with or endorsed by CMS, FDA, CDC, OSHA, HHS, OIG, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer or distributor. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.