Short answer: For an unmodified FDA-cleared or approved non-waived test, CLIA (42 CFR 493.1253(b)(1)) requires the lab to verify accuracy, precision and reportable range, and verify that the manufacturer's reference intervals fit its patient population. For a modified test, a test not subject to FDA clearance, or a laboratory developed test, the lab must establish accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals and any other needed characteristic. In both cases the lab sets calibration and control procedures from the results. Waived tests follow the manufacturer's instructions.
Last reviewed November 26, 2026. Primary sources: eCFR, eCFR, eCFR.
Sound Lab Consultants is an independent, vendor-neutral consulting firm. We do not sell instruments or reagents and are not affiliated with any manufacturer, distributor or supplier startup program.

Verification vs validation
| Characteristic | Unmodified FDA-cleared test (verify) | Modified test or LDT (establish) |
|---|---|---|
| Accuracy | Yes | Yes |
| Precision | Yes | Yes |
| Reportable range | Yes | Yes |
| Reference intervals | Verify manufacturer's intervals | Establish |
| Analytical sensitivity | No | Yes |
| Analytical specificity and interferences | No | Yes |
| Other characteristics | No | As needed for the test |
What a validation plan contains
- Purpose, test system, specimen types and intended use.
- Study design, sample numbers and sources for each characteristic.
- Acceptance criteria set before the study, based on intended use and recognized guidance.
- Statistical methods and how failures are handled.
- Calibration and control procedures that come out of the data (42 CFR 493.1256).
- Director review, approval and sign-off before patient testing.
Common validation mistakes
- Acceptance criteria written after the data are in.
- Too few samples across the reportable range.
- Reference intervals copied without checking the population.
- No interference or carryover work on modified tests or LDTs.
- Records that are not retained; CLIA retention rules in 42 CFR 493.1105 apply.
How we help
We write the plan, run or supervise the studies, analyze the data and prepare the director's summary. See QMS, IQCP and instrument validation and LDT development and validation.
Start your lab with a plan
Book a free consultation. Tell us your test menu, location and target date, and we will lay out the licensing, validation and staffing steps in order.
Book a free consultationFrequently asked questions
What is the difference between verification and validation in a clinical lab?
Verification confirms an unmodified FDA-cleared or approved test performs to the manufacturer's specifications in your lab. Validation establishes performance specifications for modified tests and laboratory developed tests.
What does CLIA require for an FDA-cleared test?
Accuracy, precision and reportable range verified, and the manufacturer's reference intervals verified for your patient population, before reporting patient results.
What extra studies does an LDT need?
Analytical sensitivity, analytical specificity including interfering substances, established reference intervals and any other characteristic needed for test performance.
Do waived tests need validation under CLIA?
CLIA requires waived tests to be performed following the manufacturer's instructions; it does not require 493.1253 verification for waived tests. Many labs still document a basic check.
Can you write our validation plans?
Yes. Book a free consultation.
Official sources
- eCFR, 42 CFR 493.1253 Establishment and verification of performance specifications
- eCFR, 42 CFR 493.1256 Control procedures
- eCFR, 42 CFR 493.1105 Retention requirements
- eCFR, 42 CFR Part 493 Laboratory Requirements
- FDA, CLIA test categorization database
Related pages
Sound Lab Consultants is an independent, vendor-neutral consulting firm. We do not sell instruments or reagents and are not affiliated with any manufacturer, distributor or supplier startup program. Sound Lab Consultants is not affiliated with or endorsed by CMS, FDA, CDC, OSHA, HHS, OIG, CAP, COLA, The Joint Commission, CLSI, any state agency, or any instrument or reagent manufacturer or distributor. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.