Short answer: If your test will be performed only in your own US CLIA high complexity lab, it is a laboratory developed test (LDT). FDA's May 2024 LDT rule was vacated on March 31, 2025, and FDA reverted 21 CFR 809.3 on September 19, 2025, so LDTs are governed by CLIA and states such as New York. To sell a test kit to other US labs you need FDA marketing authorization: a 510(k) for tests with a predicate, De Novo for novel moderate-risk tests, or PMA for high-risk tests. Foreign establishments must register and list with FDA and name a US agent.
Last reviewed November 26, 2026. Primary sources: FDA, Federal Register, FDA.
Sound Lab Consultants is an independent, advisory clinical laboratory consulting firm. This page is general information about US laboratory requirements, not legal, tax, immigration or customs advice. For entity, tax, investment-review or visa questions, work with a US attorney, a CPA and, for visas, an immigration attorney; always confirm current rules with the agency named.

Routes compared
| Route | Use when | Key requirements |
|---|---|---|
| LDT in your CLIA lab | You perform the test as a service | CLIA high complexity, validation (42 CFR 493.1253), state approval such as New York |
| 510(k) | A legally marketed predicate exists | Substantial equivalence data, QMSR, registration and listing |
| De Novo | Novel, low-to-moderate risk, no predicate | Risk-based classification request with special controls |
| PMA | High-risk (Class III) tests, many companion diagnostics | Clinical evidence of safety and effectiveness, inspections |
| EUA (history) | Only during a declared emergency | Used widely for COVID-19 tests; not a routine route |
Extra steps for foreign manufacturers
- Establishment registration and device listing, including a US agent for foreign establishments. See FDA US agent and regulatory correspondent.
- QMSR compliance, which incorporates ISO 13485 (FDA QMSR). See QMSR and ISO 13485 design controls.
- A Pre-Submission (Q-Sub) to agree the study plan.
- US clinical samples or data, often needed to support US intended use.
- CLIA categorization of the cleared test, and CLIA waiver by application if aiming for point of care.
From CE-IVD / IVDR to FDA
A CE mark under the EU IVDR does not give US market access. Your performance evaluation, risk file and design history can support a US submission, but FDA decides on its own standards and may need US data. Related: 510(k) IVD consulting, De Novo IVD consulting, LDT vs IVD vs RUO.
LDT status in 2026
Read FDA LDT rule status and LDT regulatory oversight. New York still requires approval of LDTs offered to New York patients: NYS CLEP LDT approval.
Plan your US laboratory with us
Book a free consultation. Tell us your home country, test menu, target states and timing, and we will map the US licensing, director, validation and launch steps in order. Sound Lab Consultants is advisory: we work alongside your attorney and accountant.
Book a free consultationFrequently asked questions
Do we need FDA approval to run our own test in our US lab?
No, if it is an LDT run in your CLIA high complexity lab. The 2024 FDA LDT rule was vacated in 2025.
Can our US lab use a kit with only a CE mark?
Only as a modified or non-FDA test validated as an LDT under 42 CFR 493.1253. It cannot be marketed in the US as an IVD without FDA authorization.
Does a foreign IVD manufacturer need a US agent?
Yes. Foreign establishments registering with FDA must name a US agent.
Which pathway is fastest?
A 510(k) is usually the most direct when a suitable predicate exists. Novel tests often need De Novo.
Is EUA still available?
Only under a declared emergency. It was widely used for COVID-19 tests.
Does SLC prepare FDA submissions?
We advise on IVD strategy and laboratory-side evidence and work alongside your regulatory team.
Official sources
- FDA, Laboratory developed tests
- Federal Register, Medical Devices; Laboratory Developed Tests; Implementation of Vacatur (Sept. 19, 2025)
- FDA, Premarket notification 510(k)
- FDA, De Novo classification request
- FDA, Premarket approval (PMA)
- FDA, Emergency Use Authorization
- FDA, Device registration and listing
- FDA, Quality Management System Regulation (QMSR)
- FDA, Requests for feedback and meetings for medical device submissions (Q-Submission program)
Related pages
Sound Lab Consultants is an independent, advisory clinical laboratory consulting firm. This page is general information about US laboratory requirements, not legal, tax, immigration or customs advice. For entity, tax, investment-review or visa questions, work with a US attorney, a CPA and, for visas, an immigration attorney; always confirm current rules with the agency named. Sound Lab Consultants is not affiliated with or endorsed by CMS, FDA, CDC, OSHA, HHS, OIG, CAP, COLA, The Joint Commission, CLSI, ISO, ILAC, any national accreditation body or foreign regulator, any state agency, or any instrument or reagent manufacturer or distributor. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.