cobas liat (Roche): CLIA status, QC and lab requirements
For clinics and labs using the cobas liat (Roche) for point-of-care PCR in urgent care, emergency and physician office settings.
CLIA status: Multiple cobas liat PCR tests are CLIA-waived, including SARS-CoV-2/influenza A/B (with or without RSV), strep A, and more recently a CT/NG panel and a Bordetella panel. Always confirm your exact test, cartridge and specimen type in the FDA CLIA database; categorizations can change.
If your tests are waived
- You need at least a CLIA Certificate of Waiver for the site ($248 for two years).
- Follow the package insert exactly, including the manufacturer's QC frequency and specimen requirements; deviating makes the test non-waived.
- Train each operator, keep QC, lot and temperature records, and check expiration dates.
- Anyone may perform waived testing under federal CLIA, but state law can limit it (who can perform waived tests).
If any test is moderate complexity
- You need a Certificate of Compliance or Accreditation (starting with a Registration certificate) and a qualified laboratory director.
- Verify accuracy, precision and reportable range before patient testing (42 CFR 493.1253); the free method verification calculator does the math.
- Run QC at least as CLIA requires (two levels each day of testing) or under an IQCP, enroll in proficiency testing where required, and assess staff competency.
- Testing personnel must meet 42 CFR 493.1423 (check qualifications).