Short answer: For each molecular amplification procedure, CLIA requires two control materials each day patient specimens are tested, plus an inhibition control if inhibition is a significant cause of false negatives (42 CFR 493.1256(d)(3)(v)), and an extraction control for systems with an extraction phase. Labs must also enroll in proficiency testing or verify accuracy at least twice a year for tests without required PT (493.1236(c)).
Last reviewed November 25, 2026. Primary sources: 42 CFR 493.1256, 42 CFR 493.1236, 42 CFR 493.801.
Daily QC Minimums
| Situation | CLIA minimum | Citation |
|---|---|---|
| Each molecular amplification procedure | Two control materials each day of testing; inhibition control if inhibition is a significant source of false negatives | 493.1256(d)(3)(v) |
| Test systems with an extraction phase | Two control materials, one able to detect extraction errors | 493.1256(d)(3)(iv) |
| Qualitative procedures generally | Negative and positive control | 493.1256(d)(3)(ii) |
| After complete reagent change, major maintenance or critical part replacement | Run controls before resuming patient testing | 493.1256(d)(6) |
| Every run (CMS guideline) | A blank/no-template control to detect cross-contamination | Appendix C, D3005 |
Controls rotate among operators, are tested like patient specimens, and must be acceptable before results are reported (42 CFR 493.1256(d)(7)-(8), (f)).
IQCP for Cartridge Systems
Some labs use an Individualized Quality Control Plan, a CMS-approved alternative described in Appendix C, to justify a different external control frequency for systems with built-in controls. See IQCP for molecular testing for what qualifies and what is excluded.
Planning a PCR or molecular lab?
Talk through your test menu, space, instruments and compliance plan with Sound Lab Consultants.
Book a free consultationProficiency Testing and Alternative Assessment
Labs enroll in an HHS-approved PT program for regulated analytes and specialties they test (42 CFR 493.801). For tests without required PT, the lab must verify accuracy at least twice a year (42 CFR 493.1236(c)(1)), for example through ungraded PT, split samples with another lab, or reference materials. PT samples are tested with patient workload by staff who routinely test, and must never be referred to another lab.
Review and Trending
Document QC review, Ct trends for controls where available, NTC history and PT results, and act on shifts. These feed the analytic systems quality assessment (42 CFR 493.1289). Contamination signals lead to a contamination investigation.
How Sound Lab Consultants Helps
Design of QC, IQCP and proficiency testing or alternative assessment programs for PCR and molecular labs under 42 CFR 493.1256 and 493.1236.
- Independent of instrument and reagent manufacturers
- Written plans, reports and SOPs your director can sign
- Remote and on-site support in all 50 states
- Coordination with your accreditor, billing company and counsel
Get your PCR or molecular lab right the first time
Sound Lab Consultants helps practices, laboratories and diagnostic developers plan, validate and run PCR and molecular testing under CLIA, remotely or on site, in all 50 states.
Book a free consultationFrequently asked questions
How many controls does a PCR test need?
At least two control materials each day patient specimens are tested for each molecular amplification procedure, plus an inhibition control when inhibition is a significant source of false negatives (42 CFR 493.1256(d)(3)(v)). Manufacturer instructions may require more.
Is a no-template control required for PCR?
CMS interpretive guidelines say labs performing molecular amplification should have a way to detect cross-contamination, such as a blank control with each run. Most labs run a no-template control on every run.
Do molecular tests require proficiency testing?
Labs must enroll in HHS-approved PT for regulated analytes and specialties. For tests without required PT, 42 CFR 493.1236(c)(1) requires verifying accuracy at least twice a year.
Can IQCP reduce PCR QC?
For eligible test systems, an IQCP can support a different external control frequency based on a documented risk assessment. Not every test qualifies.
Can Sound Lab Consultants build our PCR QC program?
Yes. Sound Lab Consultants designs QC, IQCP and PT or alternative assessment programs and the review and trending records.
Sources
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- 42 CFR 493.801, Condition: Enrollment and testing of samples (proficiency testing)
- 42 CFR 493.1289, Analytic systems quality assessment
- CMS State Operations Manual Appendix C, Subpart J interpretive guidelines (D3005-D3007)
- CDC MMWR, Good Laboratory Practices for Molecular Genetic Testing for Heritable Diseases and Conditions (2009)
Related PCR and molecular lab guides
Sound Lab Consultants is an independent laboratory consulting firm. We are not affiliated with or endorsed by CMS, FDA, CDC, CAP, COLA, The Joint Commission, any state agency or any instrument manufacturer. This page is general information, not legal or billing advice; confirm requirements with the agency, accreditor or payer. Sound Lab Consultants, 25 Arizona Pl, Chandler, AZ 85225.